Making Accurate Measurements of Electrically Long Devices
A device with a long electrical delay, such as a long length of cable or a SAW
presents
some unusual measurement problems to a network analyzer operating in swept sweep mode.
Often
the measured response is dependent on the analyzer’s sweep time, and incorrect data
may be obtained. At faster sweep rates, the magnitude of the response may seem to drop and
look distorted, while at slower sweep rates it looks correct. The results may indicate that a
cable has more loss than it truly does, or that a
has some unusual ripple in the
which isn’t
there.
This section describes the cause of this behavior, and how to accurately measure these
electrically long devices.
When using a vector network analyzer to measure a device that has a long electrical delay
(AT), the device’s time delay causes a frequency shift between its input and output signals. The
frequency shift, AF, equals the product of the sweep rate and the time delay:
AF=
AT
Since frequency is changing with time as the analyzer sweeps, the time delay of the DUT
causes a frequency offset between its input and output. In the analyzer receiver, the test and
reference input signals will differ in frequency by Because the test signal frequency is
slightly different than the receiver frequency, the analyzer will err in measuring its magnitude
or phase. The faster the analyzer’s sweep rate, the larger becomes, and the larger the error
in the test channel.
The HP
network analyzers do not sweep at a constant rate. The frequency
range is covered in several bands, and the sweep rate may be different in each band. So if an
operator sets up a broadband sweep with the minimum sweep time, the error in measuring a
long device will be different in each band, and the data will be discontinuous at each band
edge. This can produce confusing results which make it difficult to determine the true response
of the device.
lb reduce the error in these measurements, the frequency shift, AF, must be reduced. can
be reduced by using the following three methods:
decreasing the sweep rate
n
decreasing the time delay (AT)
n
using stepped sweep mode
Decreasing the Sweep Rate
The sweep rate can be decreased by increasing
sweep time. lb increase the
. . . . . . . . . . . . . . .
step keys, or the front panel keypad enter in the appropriate sweep time.
Selection of the appropriate sweep time depends on the device being measured; the longer the
electrical delay of the device under test, the slower the sweep rate must be. A good way to
tell when the sweep rate is slow enough is to put the vector network analyzer into a stepped
frequency mode of sweeping, and compare the data. In this mode, the vector network analyzer
does not sweep the frequency, but steps to each frequency point, stops, makes a measurement,
then goes on to the next point. Because errors do not occur in the stepped-frequency mode, it
Optimizing Measurement Results
Summary of Contents for 8719D
Page 153: ...M a k i n g M i x e r M e a s u r e m e n t s O p t i o n 089 Only1 3 19 ...
Page 189: ... 4 18 Printing Plotting and Saving Measurement Results ...
Page 338: ... Figure 6 41 Open Circuit Rmnination 6 64 Application and Operation Concepts ...
Page 467: ......
Page 468: ...8 2 Menu M a p s ...
Page 469: ......
Page 470: ...84 Menu Maps ...
Page 473: ...Menu Maps 8 7 ...
Page 475: ...Menu Maps a 9 ...
Page 476: ...I 9 10 MenuYaps ...
Page 477: ...Menu Maps 9 11 ...
Page 478: ...MlFl EF FEFEFEII TE 9 l 2 Menu Maps ...
Page 479: ......
Page 480: ...8 14 Menu Maps ...
Page 481: ...r c 1 pb6le Menu Maps 8 l 6 ...
Page 537: ...6 66 Key Definitions ...
Page 556: ... Ihble 9 2 Softkey Locations continued Softkey Front Panel Access Key Key Definitions 9 75 ...
Page 557: ...9 76 Key Definitions ...
Page 558: ...lhble 9 2 Softkey Locations continued Softkey Front Panel Access Key Key Definitions D 77 ...
Page 559: ...Ihble 9 2 Softkey Locations continued Softkey Key Definitions ...
Page 560: ... Jhble 9 2 Softkey Locations continued Key Definitions a 78 ...
Page 561: ...able 9 2 Softkey Locations continued 8 80 Key Definitions ...
Page 563: ... ble 9 2 Softkey Locations continued 842 Definitions ...
Page 565: ...able 9 2 Softkey Locations continued Softkey Front Panel Access Key 8 64 Key Definitions ...
Page 566: ... Ihble 9 2 Softkey Locations continued Key Definitions 8 85 ...
Page 567: ...Kay Definitions ...
Page 568: ... lbble 9 2 Softkey Locations continued Key Definitions 8 87 ...
Page 569: ...lhble 9 2 Softkey Locations continued Softkey Front Panel Access Key 8 88 KeyDefinitions ...
Page 570: ... able 9 2 Softkey Locations continued Softkey Front Panel Access Key Key Definitions 8 88 ...
Page 571: ...840 Key Definitions Ihble 9 2 Softkey Locations continued Softkey Front Panel Access Key ...
Page 572: ...able 9 2 Softkey Locations continued Softkey Front Panel Key Definitions 8 81 ...
Page 573: ... Ihble 9 2 Softkey Locations continued Softkey Front Panel Access Key 842 Key Definitions ...