Mixer
or frequency converters, by
exhibit the characteristic of having different
input and output frequencies. Mixer tests can be performed using the frequency offset
operation of the analyzer (with an external LO source) or using the tuned receiver operation of
the analyzer (with an external RF and LO source). The most common and convenient method
used is frequency offset.
Frequency Offset (Option 089 Only)
For a single-sideband mixer measurement, the RF source can be offset in frequency from the
input receiver frequency, allowing for a swept RF stimulus over one frequency range and
measurement of the IF response over another (in this case the output IF).
use the frequency offset guided setup for configuring a mixer measurement:
1. Enter the IF and LO frequencies.
2. Set the LO source to the entered LO frequencies.
3. Specify up conversion or down conversion.
4. Select an RF that is higher or lower in frequency than the LO. (The RF frequencies needed
are calculated by the analyzer.)
Tuned Receiver
The analyzer’s tuned receiver mode allows you to tune its receiver to an arbitrary frequency
and measure signal power. This is only possible if the signal you want to analyze is at an exact
known frequency. Therefore, the RF and LO sources must be synthesized and synchronized
with the analyzer time base. Since the analyzer is not phaselocking in this configuration,
you can use it to measure conversion loss of a microwave mixer with an RF frequency range
output.
Note
You must take care to
the output of the mixer because some of the
intermodulation and leakage products may be very close in frequency to the
desired IF. If these components are not
off, the analyzer may have
difficulty selecting the correct signal to measure.
Tuned receiver mode also increases dynamic range. Broadband techniques like diode detection
have a high noise floor, while narrowband techniques like tuned receivers are much less
susceptible to noise.
Application and Operation Concepts
Summary of Contents for 8719D
Page 153: ...M a k i n g M i x e r M e a s u r e m e n t s O p t i o n 089 Only1 3 19 ...
Page 189: ... 4 18 Printing Plotting and Saving Measurement Results ...
Page 338: ... Figure 6 41 Open Circuit Rmnination 6 64 Application and Operation Concepts ...
Page 467: ......
Page 468: ...8 2 Menu M a p s ...
Page 469: ......
Page 470: ...84 Menu Maps ...
Page 473: ...Menu Maps 8 7 ...
Page 475: ...Menu Maps a 9 ...
Page 476: ...I 9 10 MenuYaps ...
Page 477: ...Menu Maps 9 11 ...
Page 478: ...MlFl EF FEFEFEII TE 9 l 2 Menu Maps ...
Page 479: ......
Page 480: ...8 14 Menu Maps ...
Page 481: ...r c 1 pb6le Menu Maps 8 l 6 ...
Page 537: ...6 66 Key Definitions ...
Page 556: ... Ihble 9 2 Softkey Locations continued Softkey Front Panel Access Key Key Definitions 9 75 ...
Page 557: ...9 76 Key Definitions ...
Page 558: ...lhble 9 2 Softkey Locations continued Softkey Front Panel Access Key Key Definitions D 77 ...
Page 559: ...Ihble 9 2 Softkey Locations continued Softkey Key Definitions ...
Page 560: ... Jhble 9 2 Softkey Locations continued Key Definitions a 78 ...
Page 561: ...able 9 2 Softkey Locations continued 8 80 Key Definitions ...
Page 563: ... ble 9 2 Softkey Locations continued 842 Definitions ...
Page 565: ...able 9 2 Softkey Locations continued Softkey Front Panel Access Key 8 64 Key Definitions ...
Page 566: ... Ihble 9 2 Softkey Locations continued Key Definitions 8 85 ...
Page 567: ...Kay Definitions ...
Page 568: ... lbble 9 2 Softkey Locations continued Key Definitions 8 87 ...
Page 569: ...lhble 9 2 Softkey Locations continued Softkey Front Panel Access Key 8 88 KeyDefinitions ...
Page 570: ... able 9 2 Softkey Locations continued Softkey Front Panel Access Key Key Definitions 8 88 ...
Page 571: ...840 Key Definitions Ihble 9 2 Softkey Locations continued Softkey Front Panel Access Key ...
Page 572: ...able 9 2 Softkey Locations continued Softkey Front Panel Key Definitions 8 81 ...
Page 573: ... Ihble 9 2 Softkey Locations continued Softkey Front Panel Access Key 842 Key Definitions ...