The System Menu
The
key provides access to the system menu. This menu leads to additional menus
which control various aspects of the analyzer system. The following
are located
within the system menu:
provides access to the
menu.
. . . . . . . .
... .. ...
.. . . . .
Guide).
The Limits Menu
You can have limit lines drawn on the display to represent upper and lower limits or device
specifications with which to compare the test device. Limits are
in segments, where
each segment is a portion of the stimulus span. Each limit segment has an upper and a lower
starting limit value. Three types of segments are available: flat line, sloping line, and single
point.
Limits can be defined independently for the two channels, up to 22 segments for each channel.
These can be in any combination of the three limit types.
Limit testing compares the measured data with the
limits, and provides pass or fail
information for each measured data point. An out-of-limit test condition’ is indicated
ways:
with
a
FAIL
message on the screen, with a beep, by changing the color of the failing
portions of a trace, with an asterisk in tabular listings of data, and with a bit in the HP-IB
event status register B. (The analyzer also has a BNC rear panel output that includes this status,
but is only
for a single channel measurement.)
Note
The limit test output has three selectable modes For more information, refer to
Chapter 2, “Making Measurements.
Limit lines and limit testing can be used simultaneously or independently. If limit lines are
on and limit testing is off, the limit lines are shown on the display for visual comparison and
adjustment of the measurement trace. However, no
information is provided. If limit
testing is on and limit lines are off, the specified limits are still valid and the pass/fail status is
indicated even though the limit lines are not shown on the display.
Limits are entered in tabular form. Limit lines and limit testing can be either on or off while
limits are defined. As new limits are entered, the tabular columns
on
the display are updated,
and the limit lines (if on) are modified to the new definitions. The complete limit set can be
offset in either stimulus or amplitude value.
Limits are checked only at the actual measured data points It is possible for a device to be out
of specification without a limit test failure indication if the point density is
Be sure
to specify a high enough number of measurement points in the stimulus menu.
Application and Operation Concepts
Summary of Contents for 8719D
Page 153: ...M a k i n g M i x e r M e a s u r e m e n t s O p t i o n 089 Only1 3 19 ...
Page 189: ... 4 18 Printing Plotting and Saving Measurement Results ...
Page 338: ... Figure 6 41 Open Circuit Rmnination 6 64 Application and Operation Concepts ...
Page 467: ......
Page 468: ...8 2 Menu M a p s ...
Page 469: ......
Page 470: ...84 Menu Maps ...
Page 473: ...Menu Maps 8 7 ...
Page 475: ...Menu Maps a 9 ...
Page 476: ...I 9 10 MenuYaps ...
Page 477: ...Menu Maps 9 11 ...
Page 478: ...MlFl EF FEFEFEII TE 9 l 2 Menu Maps ...
Page 479: ......
Page 480: ...8 14 Menu Maps ...
Page 481: ...r c 1 pb6le Menu Maps 8 l 6 ...
Page 537: ...6 66 Key Definitions ...
Page 556: ... Ihble 9 2 Softkey Locations continued Softkey Front Panel Access Key Key Definitions 9 75 ...
Page 557: ...9 76 Key Definitions ...
Page 558: ...lhble 9 2 Softkey Locations continued Softkey Front Panel Access Key Key Definitions D 77 ...
Page 559: ...Ihble 9 2 Softkey Locations continued Softkey Key Definitions ...
Page 560: ... Jhble 9 2 Softkey Locations continued Key Definitions a 78 ...
Page 561: ...able 9 2 Softkey Locations continued 8 80 Key Definitions ...
Page 563: ... ble 9 2 Softkey Locations continued 842 Definitions ...
Page 565: ...able 9 2 Softkey Locations continued Softkey Front Panel Access Key 8 64 Key Definitions ...
Page 566: ... Ihble 9 2 Softkey Locations continued Key Definitions 8 85 ...
Page 567: ...Kay Definitions ...
Page 568: ... lbble 9 2 Softkey Locations continued Key Definitions 8 87 ...
Page 569: ...lhble 9 2 Softkey Locations continued Softkey Front Panel Access Key 8 88 KeyDefinitions ...
Page 570: ... able 9 2 Softkey Locations continued Softkey Front Panel Access Key Key Definitions 8 88 ...
Page 571: ...840 Key Definitions Ihble 9 2 Softkey Locations continued Softkey Front Panel Access Key ...
Page 572: ...able 9 2 Softkey Locations continued Softkey Front Panel Key Definitions 8 81 ...
Page 573: ... Ihble 9 2 Softkey Locations continued Softkey Front Panel Access Key 842 Key Definitions ...