AGFA DRYSTAR AXYS Service Manual Download Page 1

H

EALTH

C

ARE

 

Imaging Services

 

Service Manual

 
Document No: DD+DIS033.07E 

 
 

CONFIDENTIALITY NOTE: 

Use, dissemination, distribution or reproduction of this document by unauthorized personnel is not permitted and may be unlawful. 

DOCUMENT CONTROL NOTE: 

The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. 

10-2007   printed in Germany 

 

DRYSTAR AXYS

Type 5367/100

1

st

 edition

 

 

 

 

 

 

 

Agfa Company Confidential 

Document Node ID: 14350040 

eq_00_about manual_e_template_v02 

Copyright © 2006 Agfa HealthCare N.V. 

 

Summary of Contents for DRYSTAR AXYS

Page 1: ...zed personnel is not permitted and may be unlawful DOCUMENT CONTROL NOTE The controlled version of this document resides on MedNet Any printed copy of this document is uncontrolled 10 2007 printed in Germany DRYSTAR AXYS Type 5367 100 1st edition Agfa Company Confidential Document Node ID 14350040 eq_00_about manual_e_template_v02 Copyright 2006 Agfa HealthCare N V ...

Page 2: ...r operation or service activities may cause damage or injuries INSTRUCTION 1 Read the Generic Safety Directions document see MEDNET GSO General Info Agfa HealthCare Publications Service Manual prior to attempting any operation repair or maintenance task on the equipment 2 Strictly observe all safety directions within the Generic Safety Directions and on the product Edition 1 Revision 0 DRYSTAR AXY...

Page 3: ...opy of this document is uncontrolled Edition 1 Revision 0 DRYSTAR AXYS Chapter 0 3 10 2007 Type 5367 100 Agfa Company Confidential Purpose of this document This document provides information on the structure and contents of the Service Manual Document History Edition Revision Release Date Changes 1 0 10 2007 Initial Revision ...

Page 4: ...ools Software Tools and Auxiliary Equipment 3 3 Troubleshooting 3 4 Electrical and Mechanical Codes Fuses LEDs 3 5 Replacement of Parts 3 6 Adjustments and Calibrations 3 7 Software Menus and Setting 3 8 Software Releases Patches 3 9 FAQ Frequently Asked Questions 4 Reference and Circuit Diagrams 5 Spare Parts List 6 Accessories 7 Field Modifications 8 Manufacturing Standard Modifications 9 Mainte...

Page 5: ...the consequence Not safety relevant notes Icon Name Type of Information INSTRUCTION Indicates an instruction where it is important to follow literally the described actions IMPORTANT Highlights very important actions which have to be carried out to prevent malfunction NOTE Indicates advice to facilitate the following step or action without having a direct influence on the step or action Highlights...

Page 6: ...aphy application printing on DT2 Mammo blue base films It is designed for high throughput and for use as a central or de central printer 536711JK CDR Figure 1 System Overview Device Comment Modality The modality has to support the DICOM print protocol DRYSTAR AXYS Dry printer for 14x17 14x14 14x11 10x12 8x10 DT2 blue base or clear base films 14x11 10x12 8x10 DT2 Mammo blue base films This manual c...

Page 7: ......

Page 8: ...cument is uncontrolled Agfa Company Confidential Document No DD DIS033 07E Copyright 2007 Agfa HealthCare N V All rights reserved Technical modifications reserved Published by Agfa HealthCare N V Tegernseer Landstraße 161 D 81539 München Germany AGFA and the Agfa Rhombus are trademarks of Agfa HealthCare N V ...

Page 9: ...ing Products and part of each Service Documentation as well as Installation Planning document The latest version is available via MedNet GSO Library path General Info Agfa HealthCare Publications Service Manual Document History Edition Revision Release Date Changes compared to previous Revision 1 2 1 3 07 2009 Updated table with laser classification to latest changes of the corresponding standard ...

Page 10: ... Germany Copyright 2009 Agfa HealthCare N V All rights reserved Technical modifications reserved AGFA and the Agfa Rhombus are trademarks of Agfa HealthCare N V WARNING Improper operation or service activities may cause damage or injuries INSTRUCTION 1 Read the Generic Safety Directions document see MEDNET GSO General Info Agfa HealthCare Publications Service Manual prior to attempting any operati...

Page 11: ...abels concerning Laser Radiation 8 4 PRODUCT COMPLAINTS 10 5 REFERENCES 10 6 INTENDED USE 11 7 INTENDED USER 11 8 QUALIFICATIONS FOR OPERATION AND SERVICE TASKS 11 9 ENVIRONMENTAL AND OCCUPATIONAL SAFETY INSTRUCTIONS 12 10 CONNECTIONS TO OTHER EQUIPMENT 13 11 ACCESSORIES AND SPARE PARTS 14 12 COMPLIANCE 14 13 SAFETY DIRECTIONS FOR OPERATION 17 14 RADIATION PROTECTION 18 15 SAFETY DIRECTIONS FOR CL...

Page 12: ...on 3 Generic Safety Directions for HealthCare Imaging Products Page 4 of 28 07 2009 Agfa Company Confidential PARTS ACCESSORIES AND DEVICES 27 22 21 SAFETY DIRECTIONS FOR TRANSPORT AND SHIPMENT OF SPARE SAFETY DIRECTIONS CONCERNING MODIFICATIONS 27 23 SAFETY DIRECTIONS CONCERNING HAZARDOUS MATERIALS 27 24 RECYCLING 27 25 WASTE DISPOSAL 28 26 ERASING PROTECTED HEALTH INFORMATION PHI 28 ...

Page 13: ... reproduced copied adapted or transmitted in any form or by any means without the written permission of Agfa HealthCare Agfa HealthCare makes no warranties or representation expressed or implied with respect to the accuracy completeness or usefulness of the information contained in this document and specifically disclaims warranties of suitability for any particular purpose Agfa HealthCare shall u...

Page 14: ...le mistreatment of patients can be the consequence DANGER Direct immediate danger Death or heavy injuries can be the consequence INSTRUCTION If used in combination with the warning or caution sign Indicates a specific instruction which if followed exactly avoids the subject of the warning or caution If used without warning or caution sign Indicates an instruction where it is important to follow li...

Page 15: ...CE Conformity becomes invalid if the product is changed without explicit consent of the manufacturer This applies to all parts not only to safety elements 3 2 System Labels All system labels and software version number locations are referred to within this service document in the appropriate section Enclosed an overview of common labels according to ISO 3864 This list is not complete Magnetic Fiel...

Page 16: ...angerous to the human eye even when using optical instruments Can nevertheless produce irritating effects especially with low ambient light conditions CLASS 1 LASER PRODUCT Class 1 M Not dangerous to the human eye if no optical instruments magnifying glass or binocular are used Can nevertheless produce irritating effects especially with low ambient light conditions CLASS 1M LASER PRODUCT DO NOT VI...

Page 17: ...IRECTLY WITH OPTICAL INSTRUMENTS Class 3R Possibly dangerous to the human eye for direct view into the beam Risks of an eye injury is increasing with duration of exposure Can produce dazzling and irritating effects especially with low ambient light conditions CLASS 3R LASER PRODUCT LASER RADIATION AVOID DIRECT EYE EXPOSURE Class 3B Normally dangerous to the human eye for direct view into the beam ...

Page 18: ...Agfa HealthCare complaint procedure If the product malfunctions and may have caused or contributed to a serious injury of a patient or an accident or if there are any hazards which may cause an accident Agfa HealthCare must be notified immediately by telephone fax or written correspondence to the following address Agfa Service Support local support addresses and phone numbers listed on www agfa co...

Page 19: ... conditions 7 Intended User This manual is written for Agfa trained Field Service Engineers and Clinical Application Specialists trained users of Agfa HealthCare products and trained diagnostic X Ray clinical personnel who have received proper training Users are considered as the persons who handle the equipment as well as the persons having authority over the equipment 8 Qualifications for Operat...

Page 20: ...angerous occurrences to their line manager and co operate fully in any investigation Shall co operate with company management on matters relating to health safety and environment and where appropriate discuss with and or assist their manager in resolving matters relating to health safety and environment Shall ensure that any company equipment issued to them or for which they are responsible is cor...

Page 21: ...e jurisdiction of the hospital The Agfa HealthCare products are designed to communicate with other devices in the hospital network using DICOM protocols Connections to other equipment Warning Accessory equipment not complying with the safety requirements of this product may lead to a safety hazard INSTRUCTION Consult the Technical Documentation before making any connections to other equipment Cons...

Page 22: ...ible accessories contact your local Agfa HealthCare organization or www agfa com 12 Compliance Directive for HealthCare Imaging Products Council Directive 93 42 EEC of 14 June 1993 concerning medical devices OJ No L 169 1 of 1993 07 12 ANNEX I ESSENTIAL REQUIREMENTS GENERAL REQUIREMENTS The products are designed and manufactured in such a way that when used under the conditions and for the purpose...

Page 23: ... and provide its ESSENTIAL PERFORMANCE without disturbing other equipment Additional standards for documentation IEC 62079 Ed 1 Preparation of instructions Structuring content and presentation Harmonization Global Harmonization Task Force GHTF www ghtf org This document has been prepared to comply with Study Group 1 guidance document of the Global Harmonization Task Force GHTF www ghtf org to assi...

Page 24: ...ot installed and used in accordance with the service manual may cause interference to radio communication Note This product has been tested and found to comply with the limits for a Class A computing device pursuant to Subpart B of Part 15 of FCC Rules which are designed to provide reasonable protection against such interference when operated in a commercial environment Operation of this equipment...

Page 25: ...thCare service organization Details on the contents of messages can be found in this Technical Documentation All images created using any image technology can show artifacts which could be confused with diagnostic information If there is any doubt that the diagnostic information could be corrupted additional investigations must be performed to get clear diagnostic information Ventilation openings ...

Page 26: ...nt in the jurisdiction in which the X Ray equipment is being used and authorized means those authorized by the authority controlling the use of the X Ray equipment Full use must be made of all radiation protection features devices systems procedures and accessories Ionizing radiation can lead to radiation injuries if handled incorrectly When radiation is applied the required protective measures mu...

Page 27: ...llen screws used must be hardness rated 8 8 All screws must be secured in accordance with the corresponding data If Loctite has to be used to secure screws this is stated in the text Any Agfa service PC or tool which is to be connected via RS232 RJ45 USB or other interface to an Agfa device must not be connected to the mains but must be operated on its internal battery or indirect supply low volta...

Page 28: ... at the installation place will not exceed 1 mSv Protecting Film Screen Systems against scattered X Rays Warning Film is sensitive for X rays Poor image quality possible The film screen system shall be protected against X ray radiation this way that the annual dose equivalent at the installation place will not exceed 1 mSv Accessibility of the power disconnection device Warning Electrical device S...

Page 29: ...security and anti virus protection No ongoing computer virus protection or network security for this medical device is provided e g a computer firewall Network security and anti virus provisions are the ongoing responsibility of the user or institution Fixing equipment at the wall or floor Warning Unknown composition of wall or floor structure Risk of injury or damage Hospital management is respon...

Page 30: ...on fitted near the product and that it is easily accessible Defective covers sharp edges or protruding parts of equipment can cause injuries if accidentally knocked into Route cables and position equipment safely This device should be installed behind the institution firewall for network security and anti virus protection No ongoing computer virus protection or network security for this medical de...

Page 31: ...r for the PATIENT 19 Safety Directions for Maintenance and Repair Activities This Technical Documentation identifies the parts on which preventive inspection and maintenance shall be performed by Agfa HealthCare service personnel including the periods to be applied In general the device has to be switched off during service activities Exception If the device is switched on to perform tests pay par...

Page 32: ... according to national regulations WARNING Improper ground connections inside the device or too high leakage current may lead to electric shocks After any work at the power supply or at any component connected to mains voltage inform the responsible organization1 about the necessity of the electrical test according to national regulations If specific national regulations do not exist It is recomme...

Page 33: ...s emitting laser light For the meaning of the labels refer to section 3 3 in this document Do not look into the laser beam Do not open modules containing a laser Only open modules containing a laser if explicitly instructed to do so Do not keep tools in the laser beam unless explicitly instructed to do so Make yourself familiar with the path of the laser light and the conditions when the laser bea...

Page 34: ... Electrical shock and damage to the equipment possible Only open the PC or workstation if explicitly stated in the service manual Unplug before opening Observe anti static safety regulations Replacing fuses Warning Replacing fuses by wrong type may lead to fire hazard Use only fuses of the exact value and characteristics stated in the service manual or on the device 20 Safety Directions for remote...

Page 35: ...ical safety or radiation protection properties of a product e g changing of safety distances removal of locks instructions etc 23 Safety Directions concerning Hazardous Materials Hazardous materials is the designation for substances which can ignite or explode or which are toxic injurious to health corrosive or irritating The Hazardous Material instructions must be read and the required protective...

Page 36: ...ackaging indicates that this product shall not be treated as household waste For more detailed information about take back and recycling of this product please contact your local Agfa service organization By ensuring this product is disposed of correctly you will help prevent potential negative consequences for the environment and human health which could otherwise be caused by inappropriate waste...

Page 37: ...e_v06 Copyright 2008 Agfa HealthCare N V DRYSTAR AXYS Type 5367 100 Purpose of this document This document contains all important routines to be carried out prior to putting the machine in operation location and function of the controls and connectors of the machine all routines necessary to put an unpacked machine in operation Document History Edition Revision Release Date Changes compared to pre...

Page 38: ...V All rights reserved Technical modifications reserved AGFA and the Agfa Rhombus are trademarks of Agfa HealthCare N V WARNING Improper operation or service activities may cause damage or injuries INSTRUCTION 1 Read the Generic Safety Directions document see MEDNET GSO General Info Agfa HealthCare Publications Service Manual prior to attempting any operation repair or maintenance task on the equip...

Page 39: ...g the Film View 11 3 4 Setting Beeps on empty Tray 12 3 5 Setting Behavior for unreadable RF Tags 13 3 6 Activating the Mammo License 14 3 7 Entering Site data 15 4 FILM CALIBRATION 16 5 PRINTER MODALITY CONNECTION 18 5 1 Modality set up 18 5 2 Logical printers Set up 19 5 3 Printing of diagnostic and or test images 23 5 4 Image adjustments 23 5 4 1 Mammography specific Image Adjustments 24 5 4 2 ...

Page 40: ...further explained NOTE The DRYSTAR AXYS Plug Play Installation Manual Intranet Link Extranet Link is part of delivery in English Language and can be downloaded on MEDNET HealthCare library Step Task Code refer to Okay Installation Prerequisites 1 Check installation site prerequisites INS P P 5 2 Check shipment completeness INS Packing Slip Intranet Link Extranet Link 3 Prepare required tools INS P...

Page 41: ...ection 3 4 17 Adjust behavior for unreadable RF Tag optional APPL section 3 5 18 Enter Mammo license key optional APPL section 3 6 19 Enter site data if not yet done with Printer Installation Wizzard APPL section 3 7 Film calibration 20 Perform a film calibration for both film trays INS section 4 Printer Modality connection 21 Let the modality be set up SIC section 5 1 22 Set up logical printers o...

Page 42: ...beneath shows how to enter the Installation Wizard lists the parameters which have to be entered A more detailed description of the printer Installation Wizard is available in the DRYSTAR AXYS Reference Manual in section Using the installation wizard The Reference Manual Intranet Link Extranet Link is part of delivery and available on MEDNET HealthCare Library It is available in English Language o...

Page 43: ...tion of the printer installation wizard Proceeding for preparation of CF card with site data 1 Download the Site Data Tool from MEDNET GSO Library Hardcopy Dry Imagers DRYSTAR AXYS Freeware 2 Install it on the service PC 3 Start the Site Data Tool 4 Enter the Site Data 5 Select Save A zip file is created with the site data 6 Copy the ZIP file to CF card 7 Enter the CF card when prompted during exe...

Page 44: ...ervice menu 3 1 Default densitometer optional Key operator and Service menu 3 2 Film view optional Key operator and Service menu 3 3 Beep on empty tray optional Key operator and Service menu 3 4 RF Tag optional Service menu Browser only 3 5 Mammo license optional Service menu Browser only 3 6 Site data mandatory Service menu X3 73 7 if created offline via the Site Data Tool it is possible to enter...

Page 45: ... the setting A Define whether automatic calibration shall be executed at insertion of a new film pack B If set on ON define the number of filmpacks after which the film calibration shall be executed Default value ON every 5 filmpacks Example browser screen Default film calibration Default Densitometer Select frequency Every packs 5 Key Operator Change Image quality settings Key Operator Change Ima...

Page 46: ...ser Accessible via menu Key operator menu 2 Change settings 4 Image Quality 1 General Purpose of the setting Define a default densitometer Density values displayed during calibration or the QC procedure will be displayed re calculated for this default densitometer Default value X Rite 310 Example browser screen Default film calibration Default Densitometer Key Operator Change Image quality setting...

Page 47: ...cal keypad and browser Accessible via menu 2 Change settings 2 Input tray Purpose of the setting Change the film view in case the film curl is desired to be towards the light box Default value Normal Example browser screen Film view Tag film format Tag film type Film view Tag film format Tag film type Key Operator Change input tray settings Upper input tray Lower input tray Normal Normal Back Back...

Page 48: ...shall be informed via audible beeps about an empty tray Default value Audible alarm on beep 5 times NOTE The beep on empty tray menu is accessible via key operator menu on the keypad and key operator menu via browser Example browser screen 536701fq cdr Date Time Language Keypad contrast 60 80 Typical 68 Keypad backlight 0 100 Site name Site address City Country Country code tel Telephone Remote Co...

Page 49: ...menu 2 Change settings 2 Input Tray Purpose of the setting Define the behavior of the printer in case the RF tag is not readable or available Default value Printing not possible Example browser screen Film view Tag film format Tag film type Film view Tag film format Tag film type If RF tag is missing or malfunctioning Service Change input tray settings Upper input tray Lower input tray Normal Norm...

Page 50: ...s step Accessible via menu Browser Pages Service Engineer Tools Licenses Purpose of the setting Activation of the Mammo License Default value Mammo not activated Example browser screen HOME HELP Licenses Product options MAMMO NOT ACTIVATED Changes require reboot Activate 5364enc10 cdr HOME HELP Licenses OK Cancel Please enter the product key to activate MAMMO Figure 8 NOTE When the Mammo license i...

Page 51: ...have an entry in the infocounter about the installation site It is mandatory to fill in site data As long as the site data are not yet filled in after each switch on reboot of the printer a hint is displayed to enter the site data Default value None Example browser screen 536701fz cdr Date Time Language Keypad contrast 60 80 Typical 68 Keypad backlight 0 100 Site name Site address City Country Cou...

Page 52: ...n is to show the tasks that are required for film calibration 1 Load the films the customer will use in the upper and lower film tray Refer to note below 2 Select in key operator menu 6 Calibration 1 Film Calibration 3 Select Start Calibration The calibration film as shown in XFigure 9X is printed and the measured Dmax is shown See also note next page INPUT TRAY FILM CALIBRATION OK Dmax 3 20 XR310...

Page 53: ...on the default densitometer which is configured in Key operator menu 2 Change Settings 4 Image Quality 1 General Refer to table below Default Densitometer Target Dmax DT2B Target Dmax DT2C Target Dmax DT2 Mammo X Rite 310 3 20 O D 3 05 O D 3 80 O D X Rite 331 3 09 O D 2 93 O D 3 68 O D X Rite 341 3 05 O D 2 89 O D 3 65 O D X Rite 301 3 10 O D 2 88 O D 3 63 O D MacBeth 3 10 O D 2 91 O D 3 68 O D Gr...

Page 54: ...required for connection of the printer to the modality The following table lists these tasks Task Importance Accessible via Details see Let the modality be set up mandatory X5 15 1 Set up logical printers optional Browser pages X5 25 2 Print diagnostic and or test images mandatory X5 35 3X Adjust image settings optional Key operator menu and Browser pages X5 45 4 5 1 Modality set up 1 Let the moda...

Page 55: ...up more than one AE title for the printer See flowchart below Okay Logical Printer Setup not required Okay Logical Printer Setup not required Perform adjustments at modality Create two logical printers Proceeding see this section No Yes 536401mm cdr No Yes Adjustment at modality possible Logical Printer Setup Decision Flowchart Check whether modality supports more than one application e g Mammo CR...

Page 56: ...nter Internal name DEFAULT_SCP Called AE title DSAXYS Example browser screen Setup one or more Dicom SCP servers on this device One Dicom Print SCP server is capable of handling multiple Dicom print associations Dicom Print SCP Internal Name Called AE title Listen at Port Max allowed associations Secure channel SSL Comments DEFAULT_SCP DSAXYS 104 10 OK Create new Dicom Print SCP Cancel Changes req...

Page 57: ...Problem The PACS workstation has one AE title only The DRYSTAR AXYS does not recognize for which type of modality CT or Mammo the user wants to print Solution see XFigure 14 Example for logical printer setupX next page Create one logical printer DSAXYSMammo and one DSAXYSCT Create two different host profiles for the two logical printers with the required settings In the PACS workstation setup both...

Page 58: ...Print SCP Setup DEFAULT SCP MAMMO Printer custom custom No Yes Nick Name Nick Name Use this profile only for incoming Calling AE title Use this profile only for incoming Calling AE title Use this profile only for local SCP server Use this profile only for local SCP server Kernel Kernel cubicBell 1 1 cubicBell 0 9 Mammo modality Mammo modality CT PACS Mammo PACS PACS PACS Edit Custom Edit Custom OK...

Page 59: ...urther image adjustments required Perform image adjustments at the printer Refer to 5 4 4 No No No No No Yes Yes Yes Yes 536701mp cdr Yes Connectivity Release Document available Customer is satisfied with image quality Required image adjustments can be made at modality Image Adjustment Flowchart Check whether on MEDNET HealthCare Library a Connectivity Release Document for the Printer Modality com...

Page 60: ...ality yes has to be created This 2nd step is described here In case the printer has no Mammo modality as input skip this step 1 In the browser pages select Service Engineer Tools DICOM profile editor New 2 Enter Nickname A 3 Enter modality AE title B 4 Select Mammo Modality yes D 5 If a small clear border is required See Figure 17 next page adjust image orientation to r 180 C switch on Mammo CWR R...

Page 61: ...r Mammo CWR Rotation yes Image Orientation r180 Rotate 180 before Film tint depends on print direction Film 1 has slightly different color tint if viewed from the side Film overlap approx 2 to 4 mm only hanging at the lightbox R L Lightbox Print Direction 1 2 L R Film 1 Film 2 DICOM Host Profile Parameter Mammo CWR Rotation no default Image Orientation r0 default Same film tint for both films Film...

Page 62: ...described in the flowchart see Figure 15X MEDNET Path NOTE Connectivity Release Documents describe all parameters and required adjustments to guarantee Communication between modality and the printer Image adjustment at the modality and or the printer Connectivity Release Documents are available in MEDNET HealthCare Library General Info Connectivity Application 5 4 3 Image Adjustments at the Modali...

Page 63: ...the image See XFigure 18X 3 When finished print test images from the modality 4 Check the image quality together with the customer 5 In case the image quality is not satisfying continue as described in the flowchart See Figure 15 536401XK CDR sharp too sharp less sharp 299 199 101 200 148 143 245 235 251 138 Cubic High Res Cubic Bell Interpolation Smooth Factors DICOM values 299 200 251 245 235 Li...

Page 64: ...M Host Profile with the settings as defined in the printer modality database Corresponding printer parameter of the Installation wizard or Add modality function Brand UImportant remark U Changing parameter via key operator menu does not automatically ensure that the value is applied Two situations can be distinguished o UIn case the modality does not send a parameter U e g no Dmax value sent the v...

Page 65: ...to Figure 19 Dmax 3 20 Example browser screen Daily used name Modality AE title Look Up Table Interpolation Maximum density Key Operator Change Image quality settings Kanamori Kanamorilike Linear CubicHi res CubicBell 280 285 290 295 Linear Replication CT2 CT2_ARC Accept changes Discard changes Reset Values Welcome to this DryStar Axys printer Select abosulte smoothing factor advised 0 7 Select ka...

Page 66: ...ri 200 100 75 Kanamori like value Kanamori curve and Kanamori_like 100 Kanamori Kanamori_like 200 linear Kanamori_like 220 logarithmic Possible Look up table settings Dmax 220 0 Dmin Figure 18 536601GL CDR 5 9 0 1 1 1 0 9 0 7 Limits Legend Default Reasonable values sharp too sharp less sharp 5 9 4 9 4 9 0 1 0 2 0 7 0 9 0 7 1 1 1 2 Cubic High Res Cubic Bell Interpolation Smooth Factors Kernel value...

Page 67: ...local SCP server Host Profile type GENERIC GENERIC Number of Copies Print Priority Polarity Film Orientation Trim Film Size ID Medium Type Print even if Film Size ID Medium type not supported Kernel Magnification Pixel Size Perception LUT Decimate Crop behaviour Kanamori Like value Custom value Illumination cd m2 Reflected Ambient Light cd m2 Border Density OD x100 Empty Image Density OD x100 Mini...

Page 68: ... Check according to criteria defined by National Acceptance Test Procedure Drystar Axys Figure 23 1 Perform the acceptance test according to national regulations Example for DRYSTAR AXYS Mammo in France Acceptance test as defined by Afssaps EUREF For preparations refer to DRYSTAR AXYS Technical Documentation DD DIS033 07E Chapter 3 6 Adjustments and Calibrations Example for Germany DIN6868 Part 56...

Page 69: ...atory Key operator menu X7 27 2 7 1 Customer Instruction 1 Instruct the customer in following activities Activity Refer to User or Reference Manual Switching on the printer Warm up time 7 min User manual Switching on the DRYSTAR AXYS Switching off the printer It can take up to 20 secs until the printer switches off User Manual Switching off the DRYSTAR AXYS Loading films RF tag and its function Us...

Page 70: ... is uncontrolled Edition 1 Revision 1 DRYSTAR AXYS Chapter 1 34 06 2008 Type 5367 100 Agfa Company Confidential 7 2 Backup of Printer Settings 1 Open the upper film tray to take out the CF card for storage of the machine specific data 2 Select Key operator menu 4 Save Configuration 3 Follow the further instructions on the LCD display 5366_Reg01_017_e cdr 초 2 Figure 24 ...

Page 71: ...100 Purpose of this document This document explains the functional principle including the functions of the individual assemblies Document History Edition Revision Release Date Changes compared to previous Version 1 0 1 1 06 2009 Updated technical data in table where different Drystar printers are compared See section 2 1 Updated table with detailed comparison of Drystar 5302 with Drystar Axys Add...

Page 72: ...a HealthCare N V WARNING Improper operation or service activities may cause damage or injuries INSTRUCTION 1 Read the Generic Safety Directions document see MEDNET GSO General Info Agfa HealthCare Publications Service Manual prior to attempting any operation repair or maintenance task on the equipment 2 Strictly observe all safety directions within the Generic Safety Directions and on the product ...

Page 73: ...l LIST OF CONTENTS 1 INTRODUCTION 4 2 COMPARISON TO OTHER AGFA B W PRINTERS 5 2 1 Technical Data compared to Drystar 5300 and Drystar 5503 5 2 2 Detailed Comparison Drystar 5302 Drystar Axys 6 3 FILM RUN 7 3 1 Backside Printing Recognition 13 3 2 Left Right Shift Print Algorithm 14 3 3 Mammo Printing Functionality 16 4 CALIBRATIONS 18 5 MODULES 19 5 1 Printer Modules 19 5 2 Hardware 21 5 3 Modules...

Page 74: ...k is disabled Service Concept Customer installable No service PC needed for basic installation maintenance and service but is convenient for entering detailed host profile settings Strict modular design Production assemblies service assemblies All switches sensors and motors are monitored via current measurement on main engine board Devil CPU If online diagnosis does not detect the exact reason fo...

Page 75: ...minutes printing 4 minutes queuing 9 minutes printing Access time 65 s for 1st film 14x17 77 s for 1st film 14x17 76 s for 1st film 14x17 Film formats inch 8x10 10x12 11x14 14x14 14x17 11x14 14x17 8x10 10x12 11x14 14x14 14x17 Number trays 2 1 3 Applications All medical applications All medical applications especially CT MR except mammography All medical applications Dmax depends on densitometer wh...

Page 76: ...ar 5302 SW 2 0 0_C2 Drystar 5302 SW 2 0 0_C2 Drystar Axys Details see 1 Support of Mammo printing Section 3 6 of the Drystar Axys service manual 2 Entering Site Data via data from Site Data Tool Chapter 1 Controls Connections and Setup Procedure 3 Import Export of DICOM host profile Online HELP of the printer 4 True size deviation host profile parameter Online HELP of the printer 5 Cropping host p...

Page 77: ...hrough the printer 536602ag cdr TH 1 Edition 1 Revision 1 DRYSTAR AXYS Chapter 2 7 06 2009 Type 5367 100 Agfa Company Confidential 2 3 Filmrun lower tray Pickup Unit Rollers driven by gearbox module 1 to 5 Drum Cam drive Output Rollers Motor Suspension drive CDM Vacuum Pump Filmrun upper tray 4 5 Figure 2 Filmrun through the printer Description of one film cycle 1 The vacuum pump starts Figure 3 P...

Page 78: ... unit move down to the film Figure 4 Film Pickup 3 The gear box module starts driving the rollers In addition it opens the feeding rollers at the pickup unit 4 The pickup unit transports the film between the feeding rollers 5 The gearbox module transports the film further towards TH Figure 5 Gear Box Module Edition 1 Revision 1 DRYSTAR AXYS Chapter 2 8 06 2009 Type 5367 100 Agfa Company Confidenti...

Page 79: ...nt is uncontrolled 6 In case the film is transported upwards from the lower tray the input sensors monitor the film run Figure 6 Input Sensors Module 7 The cam drive moves the TH up to let the film edge be transported between TH and drum Figure 7 Cam drive Edition 1 Revision 1 DRYSTAR AXYS Chapter 2 9 06 2009 Type 5367 100 Agfa Company Confidential ...

Page 80: ...Depending on the previous film width 14 or 10 the suspension drive changes position or stays in the current position TH pressure adjustment 9 The cam drive closes the gap between TH and drum Figure 8 Suspension drive 10 The drum motor moves the film forward Figure 9 Drum Drive Edition 1 Revision 1 DRYSTAR AXYS Chapter 2 10 06 2009 Type 5367 100 Agfa Company Confidentia ...

Page 81: ... MedNet Any printed copy of this document is uncontrolled 11 The heat transfer from TH to film takes place Figure 10 Thermal Head 12 The print drum driven by the drum motor moves the film forward Figure 11 Print Drum Edition 1 Revision 1 DRYSTAR AXYS Chapter 2 11 06 2009 Type 5367 100 Agfa Company Confidential ...

Page 82: ... the film edge reaches the output module the CDM checks backside printing For details refer to section 3 1 Figure 12 CDM 14 When printing is finished the cam drive opens the gap between TH and drum 15 The output rollers transport the film out of the printer Figure 13 Output Module Edition 1 Revision 1 DRYSTAR AXYS Chapter 2 12 06 2009 Type 5367 100 Agfa Company Confidentia ...

Page 83: ...d the printer prints a small rectangle on top of the film and checks when the film arrives at the CDM whether this small rectangle has Dmax In case the density is below a certain threshold the film must have been inserted wrong The printer stops printing informs the user of back side printing to check the correct positioning of the film and to remove the film jam Thus the thermal head is only cont...

Page 84: ...TH looses its characteristic relatively fast which is determined by the Micro profile calibration in production As a consequence the TH produces after many prints small print stripes at this position in 14 films where the black border of the 10 films is located To prevent that the TH has to print the black border always at the same position a left right shift algorithm is built in For each film pa...

Page 85: ...ge Info Image Info Image Info Image Info Image Info Image Info Image Info Image Info Image Info Image Info Image Info Shift of complete Image Info 0 3 mm for each film pack Exaggerated print stripes without left right shift algorithm Left right shift algorithm Image Info Image Info Image Info Image Info Figure 16 Left right shift print algorithm Edition 1 Revision 1 DRYSTAR AXYS Chapter 2 15 06 20...

Page 86: ... for the left one for the right breast For diagnosis they are hanging symmetrically at the lightbox Figure 17 Special Mammo Adjustments No DICOM host profile required to print via DICOM at the printer from non Mammo modality A dedicated host profile is required for each Mammo modality A host profile for Mammo is defined via following DICOM host profile parameters o Mandatory Parameter Mammo Modali...

Page 87: ...or tint if viewed from the side only 8x10 and 11x14 Film overlap approx 1 mm only hanging at the lightbox L R R L Lightbox Print Direction 1 2 Film 1 Film 2 DICOM Host Profile Parameter Mammo CWR Rotation no default Same film tint for both films Film overlap approx 4 mm Fog of blank frame of film 1 adds in film 2 536401gq cdr L R Lightbox Without Chest Wall Rotation Print Direction R L Film 1 Film...

Page 88: ...rint width Make film calibration this accurate that it passes the acceptance criteria as defined by Afssaps EUREF The default film calibration procedure is not necessarily accurate enough To be done in following cases Whenever the internal densitometer has to be adapted to an external densitometer for image quality control If density reproduction check is not ok Unacceptable line artifacts visible...

Page 89: ...er Modules This section lists the different printer modules spare parts 13 10 11 15 DOCUMENT CONTROL NOTE The controlled version of this document resides on MedNet Any printed copy of this document is uncontrolled Edition 1 Revision 1 DRYSTAR AXYS Chapter 2 19 06 2009 Type 5367 100 Agfa Company Confidential 12 2 4 5 8 1 6 7 9 3 14 16 17 18 21 22 19 20 Figure 19 Printer Modules ...

Page 90: ...r Drystar 5300 Drystar 5302 13 PUMP VALVES ASSY Vacuum for suction cups of pickup module Drystar 5500 Drystar 5503 Drystar 5302 17 14 Input Sensors Module Filmrun check before print drum Drystar 5302 11 15 DRUM DRIVE MODULE Drum drive Drystar 5302 07 16 OSB SWITCH BOARD TH power Drystar 5300 Drystar 5302 17 PRINT DRUM MODULE Print drum Transport during printing Drystar 5302 18 THERMAL HEAD Yes Ima...

Page 91: ... Thermal Head IDE RS232 Data PCI Power Figure 20 Hardware Components Overview Hardware Component Description Thanos Board CPU Keeps the operating system and application software in RAM 512 MB during operation Keeps a flash EPROM for Bootup Performs image processing Devil CPU Controls all sensors motors valves and pump Controls the power supply Sends the image data to the Thermal Head Provides for ...

Page 92: ... buttons and LEDs required for operation Hard disk Keeps the operating system all application software and diagnostic data CF drive To save or install data via CF card type II Thermal head Transfers the diagnostic data pixels via heating elements to the heat sensitive film OSB board Switch board for the different TH heating elements Provides the power for heating up the resistor elements Power sup...

Page 93: ...mple for the readout of an E Label is given in Figure 22 In case the E label is defective or not connected the printer issues a warning message Printing is possible however also without E label The decision whether a module has an E label or not depends on whether the corresponding module Can be repaired in the repair center or Requires module specific data The table in section 5 1 lists which mod...

Page 94: ...ttp www agfa com module Output nrOfCycles NrOfCycles Value LastRepairDate Uri http www agfa com module Output lastRepairDate LastRepairDate Value RepairSiteName Uri http www agfa com module Output repairSiteName RepairSiteName Value NrOfRepairs Uri http www agfa com module Output nrOfRepairs NrOfRepairs Value LastRepairCode Uri http www agfa com module Output lastRepairCode LastRepairCode Value La...

Page 95: ...ently monitoring the correct feedback signal and timing of the valves and the pump motors and sensors which are connected to the Devil module In case of a mismatch an error is issued M Operating System Printer Software Application Software Online Diagnosis Devil CPU 536602aa cdr MOD04 Please replace module Example message Figure 23 Edition 1 Revision 1 DRYSTAR AXYS Chapter 2 25 06 2009 Type 5367 1...

Page 96: ...repaired by service personnel Warnings and error messages are written into the infocounter Filmrun Step 1 user clearable Errors Classification and treatment 536602af cdr key operator clearable service clearable Infocounter Online Diagnosis event Message in plain text E g close upper cover warning Message in plain text E g CDM not calibrated defect Error message HC10907 Suspension motor M3 not conn...

Page 97: ...l TMML Thermal Model Multi Layer R D name of the product Æ official name A Sharp describes this interaction very accurate leading to a very precise reproduction of the image With the thermal model TMML the image looks very sharp and has the image quality of a Laser Imager Pixel Stream Thermal Model Mathematical Description of Interaction Film TH Thermal Model Heat of neighbor pixel Heat of previou...

Page 98: ......

Page 99: ...ncontrolled 09 2007 printed in Germany Agfa Company Confidential Document Node ID 14350213 eq_03 1_safety repair_e_template_v02 Copyright 2006 Agfa HealthCare N V DRYSTAR AXYS Type 5367 Purpose of this document This document lists the printer specific safety notes in addition to the generic safety notes refer to chapter 0 Document History Edition Revision Release Date Changes 1 0 09 2007 Initial V...

Page 100: ...S Chapter 3 1 2 09 2007 Type 5367 Agfa Company Confidential WARNING Improper operation or service activities may cause damage or injuries INSTRUCTION 1 Read the Generic Safety Directions document see MEDNET GSO General Info Agfa HealthCare Publications Service Manual prior to attempting any operation repair or maintenance task on the equipment 2 Strictly observe all safety directions within the Ge...

Page 101: ...d Edition 1 Revision 0 DRYSTAR AXYS Chapter 3 1 3 09 2007 Type 5367 Agfa Company Confidential LIST OF CONTENTS 1 SAFETY NOTES FOR SERVICE ACTVITIES 4 1 1 Safety Notes for Installation Planning and Installation 4 1 2 Safety Notes for preventive and corrective Maintenance Work 4 1 3 General Safety Note for all Service Actvities 5 1 3 1 Safety switch 5 1 3 2 Electrical Check after Power Supply Replac...

Page 102: ...tion when lifting the device The table might break and cause injuries or damage to the device The device weight may be too heavy for the table Communicate to the customer that the table must be able to carry more than 90 kg 198 lb 1 2 Safety Notes for preventive and corrective Maintenance Work WARNING Thermal Print Head is hot up to 70 C Risk of burns Avoid contact with the Thermal Print Head Obse...

Page 103: ... cover is opened the safety switch is activated It cuts off the voltage to all motors and the print head S 8 536631aa cdr Figure 1 WARNING If the safety switch is overridden with a service key there are risks of injuries Keep your hands and garments away from the device Remove service key before the device is handed over to the customer Before the printer is closed the service key must be removed ...

Page 104: ... maintenance check the protective earth resistance according to the national regulations The electrical check is split up in two steps Step Task Details see 1 Perform the basic protective earth test 1 3 2 1 2 Perform the electrical check according to National Regulations 1 3 2 2 1 3 2 1 Basic Protective Earth Test 1 Use a multimeter that is capable to measure a resistance 1 Ohm 2 Put one probe at ...

Page 105: ... at a few metallic parts the customer could touch The electrical resistance in this case has to be 0 3 Ohm Refer to Figure 3 7 check the printer for loose grounding connections if the electrical resistance of one measurement exceeds the limits 0 2 Ohm without mains cable 0 3 Ohm with mains cable Figure 3 1 3 2 2 Electrical Check according to National Regulations 1 Perform the electrical check acco...

Page 106: ......

Page 107: ...n Germany Agfa Company Confidential Document Node ID 18789433 eq_03 2_tools aux_e_template_v02 Copyright 2007 Agfa HealthCare N V DRYSTAR AXYS Type 5367 100 Purpose of this document This document describes all tools software hardware which are required for servicing of the printer incl general instructions for software installation Document History Edition Revision Release Date Changes 1 0 10 2007...

Page 108: ...ties may cause damage or injuries INSTRUCTION 1 Read the Generic Safety Directions document see MEDNET GSO General Info Agfa HealthCare Publications Service Manual prior to attempting any operation repair or maintenance task on the equipment 2 Strictly observe all safety directions within the Generic Safety Directions and on the product Edition 1 Revision 0 DRYSTAR AXYS Chapter 3 2 2 10 2007 Type ...

Page 109: ... the local Keypad 10 3 2 Infocounter 10 3 2 1 Infocounter Analysis via printed infocounter on Film 11 3 2 2 Infocounter Analysis via Printers Infocounters Report 12 3 2 3 Infocounter Analysis via Service Command Session 12 3 2 4 How to evaluate the Infocounter 14 3 3 Test images 15 3 3 1 Testimage STI1 16 3 3 2 Testimage STI2 18 3 3 3 Testimage SMPTE 19 3 3 4 Testimage Medical_Image 21 3 3 5 Testi...

Page 110: ...ompany Confidential 3 6 2 Printer Bootup Message 45 3 7 ADVP 49 3 7 1 ADVP as DICOM print SCU 50 3 7 2 ADVP as DICOM print SCP Print Emulator 50 3 7 3 ADVP as Communication Recorder 52 4 SOFTWARE INSTALLATION 55 4 1 SW Update Loading a SW Patch 55 4 1 1 Software Update via CF card 56 4 1 2 Software Update via a remote computer 56 4 2 Emergency SW Installation on a new empty Hard Drive 57 4 3 Emerg...

Page 111: ...DM calibration Ink Eraser Pelikan 10 999991176 0 For the dry TH cleaning procedure Isopropyl Alcohol and lint free cloth Commercially available To clean the TH Magnifying glass 8x CM 9 9579 9904 0 To check image quality of films at the lightbox Service key CM 9 0426 6309 0 To override the safety switch Standard ruler 1mm Commercially available For horizontal image adjustment Density Reproduction C...

Page 112: ...ther densitometers are not supported 1 2 Optional tools The optional tools are required in exceptional cases only Tool Part Number Purpose Crossed network cable EB 72010270 or commercially available To get access to the printer homepage in case access via the hospital network is not possible Note Many service PCs as of approx year 2006 do not require a crossed network cable They autodetect the typ...

Page 113: ...cification is the basic requirement for the Service PC Hardware USB port or integrated CF Card reader Serial port Network connection RJ45 Software Browser Internet Explorer 5 0 1 2 2 Recommended configuration of the Service PC The national Agfa GICS department Global Information and Communication Services is responsible for selection and configuration of the Service PC The actual specification is ...

Page 114: ...trol Panel Network Properties Internet Protocol TCP IP Properties Obtain an IP address automatically Refer to Figure 1 Note The path may differ for different operating systems Setting for Windows XP Select Control Panel Network Connections Local Area Connection Properties Internet Protocol TCP IP Properties Tab Alternative Configuration Automatic Private IP address Refer to Figure 2 2 Switch on th...

Page 115: ...1 Infocounter Troubleshooting 3 2 Test images Troubleshooting image faults Confirmation of perfect image quality 3 3 Show error codes View reason meaning and cure of error codes 3 4 Printer Browser Pages Perform DICOM attribute adjustments e g image adjustments 3 6 Logging View the online logging for troubleshooting purposes 3 6 ADVP Test the DICOM communication between modality and printer 3 7 NO...

Page 116: ... username drystar password axys 2 For a complete description of the Key operator functions please refer to the DRYSTAR AXYS Reference manual An overview of all key operator menus is available in section 3 7 Software menus Settings 3 2 Infocounter Purpose The infocounter is used for troubleshooting purposes and for investigation of the printer history before maintenance It contains following inform...

Page 117: ... Figure 6 Acces to the infocoeunter Refer to A Via printed infocounter on film 3 2 1 B Via Printers Infocounters Report 3 2 2 C Via Service Command session 3 2 3 NOTE We recommend to use for Infocounter Analysis method A Via printed infocounter on film as this is the fastest way to access the infocounter 3 2 1 Infocounter Analysis via printed infocounter on Film Required Tools none On the local ke...

Page 118: ... 5 Analyze the infocounter file as described in section 3 2 4 How to evaluate the Infocounter Figure 4 3 2 3 Infocounter Analysis via Service Command Session Required Tools Service PC required to connect to the printer and analyze the infocounter 536632el cdr Printer Service PC 1 Connect the Service PC to the network 2 Start the browser Internet Explorer 5 3 Enter the IP address of the printer in ...

Page 119: ...t Page Software versions Show printer status Enter key operator command session Enter service command session Release remote command session Scaling Setup DICOM profile editor LPD Profile Editor Network configuration Printqueue editor Controller logging Modules Remote sessions Welcome to this DryStar Axys printer Connectivity tools Module repair Service Loggings 1 2 To printer flash card To this c...

Page 120: ...ction 4 1 Infocounters txt evaluation What to check in the infocounter Comment 2 3 Reliability Should normally be not higher than 0 05 Higher fault rates need a thorough investigation 3 3 Hardware Modification History By comparing the status of the device with the available modifications refer to Technical Documentation chapter 8 the exact hardware status can be determined 3 4 Software modificatio...

Page 121: ...STI1 Composed of many different test images to check all aspects of image quality 3 3 1 STI2 Composed of many different test images to check all aspects of image quality 3 3 2 SMPTE Standard test image for evaluation of image quality 3 3 3 Medical_image Test image to check whether stripes or dots which are slightly visible in one of the technical images are also visible in a medical test image 3 3...

Page 122: ...ls Enter Service Command Session 3 Print Testimage 1 Test image STI1 ZIP 2 The test image as displayed in Figure 7 will be printed Figure 7 Evaluation Evaluate the test image at the lightbox as follows 1 Field 1 contains a continuous grey wedge from grey value 0 to 255 Check no single grey steps may be visible in case single grey steps are visible this is called contouring The naming for this effe...

Page 123: ... Check all 256 squares must be visible In addition a contrast difference must be visible between the individual square and the small square inside 3 Field 3 contains a flat field with density value 2 0 Check no stripes in any direction may be visible NOTE In case stripes dots or insufficient density reproduction in the image can be observed refer to the troubleshooting guide see MEDNET GSO Library...

Page 124: ...ce Command Session 3 Print Testimage 1 Test image STI2 ZIP 2 The test image as displayed in Figure 9 will be printed Figure 9 Evaluation Evaluate the test image at the lightbox as follows 1 Field 1 contains white and black lines dots of 1 2 and 3 pixel width Check all lines dots must be visible 2 Field 2 consists of a SMPTE testimage Refer to 3 3 3 3 Field 3 consists of a flatfield with Dmax 1 0 4...

Page 125: ...ation of image quality and for trouble shooting see chapter 3 3 Usage 1 In the key operator menu select 3 Print Image 1 Testimage SMPTE_xxx tif 2 The test image as displayed in Figure 11 will be printed SMPTE_lin TIF Taste LUT linear Dmax Dmax Printer SMPTE_l300 TIF Taste LUT linear Dmax 3 0 SMPTE_kan TIF Taste LUT Kanamori Dmax Dmax Printer Figure 11 Edition 1 Revision 0 DRYSTAR AXYS Chapter 3 2 ...

Page 126: ... 4 consists of a black square 100 black with a smaller grey square inside 95 black The square must show 2 different densities i e inside there must be a smaller square with slightly lower density 5 Field 5 consists of a white square 0 black with a smaller grey square inside 5 black Check The square must show 2 different densities i e inside there must be a smaller square with slightly higher densi...

Page 127: ...ice Command Session 3 Print Testimage 1 Test Image Medical_image ZIP 2 The test image as displayed in Figure 13 will be printed Figure 13 Evaluation Evaluate the test image at the lightbox as follows 1 This medical test image of a chest is the most critical medical test image regarding visibility of print stripes or dots due to its huge bright areas In bright areas the eye is most sensitive Check ...

Page 128: ...d Session 3 Print Testimage 1 Test Image LinCheck ZIP 3 The test image as displayed in Figure 14 will be printed 536632ga cdr Figure 14 Evaluation 4 The test image contains a grey wedge This grey wedge is measured during the density reproduction check Refer to chapter 3 6 Adjustments 536632gb cdr Density measurement area Figure 15 NOTE For check of density reproduction only measure in the density ...

Page 129: ...k whether stripes or dots that are visible in the STI1 or STI2 image are also visible in the testimage medicalimage IMPORTANT To make this test film available at a printer with software 1 6 0_C1 proceed as follows Connect via telnet to the printer 1 At the C prompt enter copy C images service m zip D images service 2 Print the image from Service Command Session menu 3 Print Testimage 4 medicalimag...

Page 130: ... menu on local display 3 5 2 Service Engineer Tools Functions for configuration and troubleshooting of the Controller part of the printer 3 5 3 Specialist Tools Software tools reserved for R D to perform special actions during test phase of the printer Not further described Security Tools Software Tools to configure user access control and secure communication as defined by HIPAA This menu is rese...

Page 131: ...ools Edition 1 Revision 0 DRYSTAR AXYS Chapter 3 2 25 10 2007 Type 5367 100 Agfa Company Confidential Delete Set Emergency View Status Id Patient Name Host Created On Sent Status Priority XYZ CT_A 875 16 02 2005 Printing Normal 0 1 Printing Normal 0 1 XYZ CT_A 876 16 02 2005 XYZ CT_A 877 16 02 2005 Printing Normal 0 46 XYZ CT_A 878 16 02 2005 Calculating Normal 0 1 XYZ CT_A 879 16 02 2005 Waiting ...

Page 132: ...d session Shows list of menus like Key Operator menu on local keypad 2 Release remote command session Makes a reset on the key operator command session to release the connection in case the key operator command session hangs NOTE As long as a key operator or service command session via Printer Browser Pages is active access to the printer via local keypad is not possible The printer will show the ...

Page 133: ...ork configuration Printqueue editor Licenses Controller logging Modules Remote sessions Connectivity tools Module repair Figure 20 Main Purpose Provides functions for configuration and troubleshooting of the Controller part of the printer Username Password mega Sub menus 1 List of errors Shows list of all HC error codes Same content like chapter 3 3 appendix error codes 2 Infocounters Report Shows...

Page 134: ...ferent menus to define printer behavior regarding DICOM communication 13 LPD profile editor Different menus to define printer behavior regarding Postscript communication 14 Network Configuration Allows special network settings to adapt to other network devices 15 Printqueue editor Provides functions for extended queue management 16 Controller logging Provides functions to view filter adapt and sav...

Page 135: ...he device detects each defective motor and sensor and issues a corresponding error code Sub menus By clicking on the assembly the corresponding assembly page is displayed See Figure 22 Comments 1 By pressing activate deactivate the motor can be started 2 To check a sensor activate the sensor manually and press REFRESH Now the changed state of the sensor has to be displayed if the sensor is still o...

Page 136: ...nt resides on MedNet Any printed copy of this document is uncontrolled Figure 22 Service Engineer Example for a Module Page For comments see previous page Figure 23 Service Engineer Example for a Module Page E Label content Edition 1 Revision 0 DRYSTAR AXYS Chapter 3 2 30 10 2007 Type 5367 100 Agfa Company Confidential ...

Page 137: ... 10 2007 Type 5367 100 Agfa Company Confidential 536632el cdr Printer Service PC Figure 24 Printer 536632ek cdr Crossed UTP network cable Figure 25 As an alternative if it is not possible to put the Service PC into the hospital network 1 Use a crossed network UTP cable to connect the network interface of the Service PC directly with the network interface of the printer Note Many service PCs as of ...

Page 138: ...o have the Service PC working only in a subnet get from him the netmask also called subnet mask 2 Enter IP address and subnet mask as displayed in Figure 26 to Figure 29 Case 2 The service PC is not part of the hospital network but connected via crossed UTP network cable to the printer 1 Look up the printer IP address and subnet mask in the printer key operator menu show settings Network 2 Define ...

Page 139: ...des on MedNet Any printed copy of this document is uncontrolled Default Subnet Masks set for NO subnetting Class First 3 digits of IP subnet mask A 0 126 255 0 0 0 B 128 191 255 255 0 0 C 192 223 255 255 255 0 Figure 28 Figure 29 Edition 1 Revision 0 DRYSTAR AXYS Chapter 3 2 33 10 2007 Type 5367 100 Agfa Company Confidential ...

Page 140: ...tings In case the connection to the Internet is configured for a connection via Proxy server this has to be disabled to be able to connect to the printer In the Internet Explorer select Tools Internet Options Select tab Connections In case a proxy server is configured tick the checkbox Bypass Proxy server for local addresses Figure 30 Edition 1 Revision 0 DRYSTAR AXYS Chapter 3 2 34 10 2007 Type 5...

Page 141: ...e 2 Select the desired menu 3 Enter user name and password when prompted mega Enclosed two examples which shall explain the usage of the Printer Browser Pages Example description refer to 1 Adapt the taste LUT for a certain host by creating a host specific DICOM host profile 3 5 5 1 2 Adapt the annotations for all DICOM inputs 3 5 5 2 Service engineer tools Key operator tools Operator tools Specia...

Page 142: ...tar Axys printer HOME General information SERVICE TOOLS List of errors Printer Infocounters report DICOM LPD and HIPAA report QA Report Page Software versions Show printer status Enter key operator command session Enter service command session Release remote command session Scaling Setup DICOM profile editor LPD Profile Editor Network configuration Printqueue editor Controller logging 536732et cdr...

Page 143: ...version of this document resides on MedNet Any printed copy of this document is uncontrolled Figure 34 Host profile editor 1 Select the connected host if present otherwise leave it on generic Figure 35 Edition 1 Revision 0 DRYSTAR AXYS Chapter 3 2 37 10 2007 Type 5367 100 Agfa Company Confidential ...

Page 144: ...he taste LUT will be adapted to Kanamori_like 200 NOTE In case the checkbox is checked before a parameter in the DICOM profile editor this parameter is used no matter whether the modality sends the parameter or not overwrite functionality In case the checkbox is not checked the value defined in the host profile is only used in case it is not sent by the modality Image adaptations should be made pr...

Page 145: ...ot supported Kernel Magnification Pixel Size Perception LUT Decimate Crop behaviour Kanamori Like value Custom value Illumination cd m2 Reflected Ambient Light cd m2 Border Density OD x100 Empty Image Density OD x100 Minimum Density OD x100 Maximum Density OD x100 Header Annotation Contents Font Footer Annotation Window Level Image Orientation Conformance Status Level Nevent enabled Mammo modality...

Page 146: ...essing the HELP button Figure 37 3 5 5 2 Example 2 Define annotations for all inputs In this example the defaults for all hosts independent of the AE title are adapted Here it is only possible to adapt the annotations 1 In the printer homepage select Service Engineer tools Enter service command session 2 Enter username and password 3 Select DICOM profile editor 4 Select Default site annotations ED...

Page 147: ... Refer to Figure 39 to see where annotation 1 is located 2 Enter the annotation text Example St Joseph s Hospital 3 Possibly adapt the font and font size of the annotation 4 In the Annotation Box 1 select Submit 5 Check the checkbox before Annotation 1 6 Select Submit Now independent of the AE title all films will have annotation 1 St Joseph s Hospital Edition 1 Revision 0 DRYSTAR AXYS Chapter 3 2...

Page 148: ...s DOCUMENT CONTROL NOTE The controlled version of this document resides on MedNet Any printed copy of this document is uncontrolled Figure 40 Site Defaults Window Edition 1 Revision 0 DRYSTAR AXYS Chapter 3 2 42 10 2007 Type 5367 100 Agfa Company Confidential ...

Page 149: ...In the Service Engineer Tools go to Controller Logging Step 2 Set tasks PrintJobHandler DCMPrintSCP_0 DCMPrintSCP_1 if available to TRACE Step 3 Produce the fault by sending an image from the modality or by printing a test image Step 4 Reboot the printer Step 5 Get the session files See note 1 Step 6 Get the infocounter file See note 1 Step 7 Save the DICOM LPD and HIPAA report See note 2 Step 8 S...

Page 150: ...Step 3 In the PIDEX tool select Tools logview and open the desired logfile Step 4 In the PIDEX tool go to function Edit find to find the desired information It is also possible to select a single word in the logfile Then all other occurances of this expression are also highlighted NOTE A session file is named e g DSAXYS_1052_SES_21OCT2007_09 55 48 ZIP where the different figures represent followin...

Page 151: ...printer on Now the bootup message appears 4 When the Printer login appears the bootup message is finished 536632ep cdr Figure 41 3 6 2 1 Setup of the HyperTerminal NOTE The hyper terminal setup as displayed here can be used for all AGFA devices Instead of the HyperTerminal following other programs can be recommended to connect to terminal level xt xt exe is an AGFA program which is part of the IMO...

Page 152: ...es on MedNet Any printed copy of this document is uncontrolled Figure 42 2 Enter a name for the connection e g AGFA Figure 43 3 Select COM1 or COM2 for communication port depending on what you are using on the Service PC Figure 44 Edition 1 Revision 0 DRYSTAR AXYS Chapter 3 2 46 10 2007 Type 5367 100 Agfa Company Confidential ...

Page 153: ...dNet Any printed copy of this document is uncontrolled 4 Make the setup of the HyperTerminal session as displayed here i e 9600 Bps 8 data bits no parity 1 stop bit flow control Xon Xoff 5 Select OK Figure 45 6 Select Properties Figure 46 Edition 1 Revision 0 DRYSTAR AXYS Chapter 3 2 47 10 2007 Type 5367 100 Agfa Company Confidential ...

Page 154: ...resides on MedNet Any printed copy of this document is uncontrolled 7 Select settings Figure 47 8 Select VT100 for emulation type 9 Press OK Figure 48 10 Press Enter Now the bootup message or login prompt is visible Figure 49 Edition 1 Revision 0 DRYSTAR AXYS Chapter 3 2 48 10 2007 Type 5367 100 Agfa Company Confidential ...

Page 155: ...hooting tool Usage ADVP can be used for following purposes Purpose Details see 1 Send test images from the service PC to the printer via network 3 7 1 2 Record DICOM print sessions for further troubleshooting 3 7 2 3 Record DICOM print sessions for changing and re sending them to the printer 3 7 3 NOTE ADVP can run on the service PC or any Windows PC in the network The only pre requisite The PC mu...

Page 156: ...uration screen 3 Select Test Print Film 4 Select SMPTE or any other test film 5 In the Configuration window fill in Called Printer AE Title and Printer IP address 6 Select OK Now the desired test film is sent to the printer Figure 51 ADVP as Print SCU 3 7 2 ADVP as DICOM print SCP Print Emulator ADVP as DICOM print SCP can be used to check the DICOM communication sent from the modality ADVP in thi...

Page 157: ...window fill in Called Printer AE Title this is the AE title of the printer Called Host AE Title ADVP Listen port usually 104 this is the port the modality sends DICOM messages 7 Select okay Now ADVP is prepared for recording the communication between modality and the printer Figure 52 ADVP as Print Emulator NOTE In case ADVP runs on the printer just select Stop Controller to stop the application s...

Page 158: ...ith the changed attribute s In general following configurations are possible a Service PC is part of the network The modality sends to the service PC the service PC transmits the DICOM image and control data on the fly to the printer Modifying the DICOM data can be done without changing the configuration b Service PC is only temporary integrated in the network to emulate the printer The modality s...

Page 159: ...ect Basic Grayscale in the Start up configuration screen 5 Select Application Comms Recorder Start 6 In the following window fill in Called Printer AE Title this is the AE title of the printer Called Host AE Title ADVP Listen port usually 104 this is the port the modality sends DICOM messages Printer IP address this is the IP address of the printer printer listen port usually 104 7 Select okay Now...

Page 160: ...DICOM data 1 Select Application Comms Recorder Stop 2 Select Application Comms Recorder Modify 3 In the Printer Snapshot edit the desired parameters 4 Select Application Comms Recorder Replay Now the image is re sent to the printer and printed at the printer with the adapted data Figure 54 Resending the DICOM data Edition 1 Revision 0 DRYSTAR AXYS Chapter 3 2 54 10 2007 Type 5367 100 Agfa Company ...

Page 161: ...nter that does not bootup anymore 4 3 NOTE A software upgrade is a functional extension It is usually sold through the Sales and Services Channels A software update is a corrective modification of the basic software version It does not include new functionality SW Updates Patches are available on MedNet 4 1 SW Update Loading a SW Patch This chapter describes the SW Update with a patch NOTE For ins...

Page 162: ...ckup on CF card 4 1 2 Software Update via a remote computer Purpose Software Update Installation of a software patch SW Update Service engineer tools Enter Key operator command session 9 Installation 1 New Software Installation Username mega Username 536632eo cdr Key Operator Installation Choose Dataset Browse Proceed Go Back Reset Values Welcome to this DryStar Axys printer 1 Establish a web brow...

Page 163: ...e programming is interrupted by power loss this specific board may become unusable Do not reboot the printer before requested by the Printer Browser Pages 4 2 Emergency SW Installation on a new empty Hard Drive Purpose This chapter describes the SW Installation on an empty hard drive e g after exchange of the hard drive SW Installation on empty HD Edition 1 Revision 0 DRYSTAR AXYS Chapter 3 2 57 1...

Page 164: ...r Emergency SW Installation on a new empty hard disk without Configuration Restore the printer is in the same state like it has been shipped from factory default IP address 169 254 10 10 Emergency SW FOUND ON EXTERNAL CF card Install emergency SW YES V NO X Emergency SW Install Do you want to preserve config YES V NO X 536632gc cdr Figure 58 NOTE The printer Emergency_Install software can be obtai...

Page 165: ...e the CF card is lost or not readable the complete printer setup has to be redone Emergency SW FOUND ON EXTERNAL CF card Install emergency SW YES V NO X Emergency SW Install Do you want to preserve config YES V NO X 536632gc cdr Figure 59 SW Installation on empty HD Edition 1 Revision 0 DRYSTAR AXYS Chapter 3 2 59 10 2007 Type 5367 100 Agfa Company Confidential 5366_Reg01_017_e cdr 초 2 1 Insert th...

Page 166: ...his specific board may become unusable Do not reboot the printer before requested by a local keypad message NOTE When you choose during Emergency SW Installation to preserve the configuration following steps are executed internally 1 A backup of the printer configuration data is created and copied to RAM 2 The software is copied from CF card to hard disk and inflated 3 The boards are programmed 4 ...

Page 167: ...ential Document Node ID 14349766 eq_03 3_troubleshooting_e_template_v06 Copyright 2011 Agfa HealthCare N V DRYSTAR AXYS Type 5367 100 Purpose of this Document This document describes the systematic troubleshooting of the printer Document History Edition Revision Release Date Changes compared to previous Version 1 0 1 1 03 2011 Replaced previous list of error codes by error codes of software 2 1 0_...

Page 168: ... V All rights reserved Technical modifications reserved AGFA and the Agfa Rhombus are trademarks of Agfa HealthCare N V WARNING Improper operation or service activities may cause damage or injuries INSTRUCTION 1 Read the Generic Safety Directions document see MEDNET GSO General Info Agfa HealthCare Publications Service Manual prior to attempting any operation repair or maintenance task on the equi...

Page 169: ...ntrolled version of this document resides on MedNet Any printed copy of this document is uncontrolled Edition 1 Revision 1 DRYSTAR AXYS Chapter 3 3 3 of 72 03 2011 Type 5367 100 Agfa Company Confidential LIST OF CONTENTS 1 GENERAL TROUBLESHOOTING PROCEDURE 4 2 ERROR CATALOGUE 5 ...

Page 170: ...mage s show stripes dots density or contrast faults For more than 5 minutes the printer shows message Please wait selftest proceeding Refer to table next page See also Troubleshooting Guide See also Troubleshooting Guide See also Troubleshooting Guide Refer to IMPORTANT note next page Refer to IMPORTANT note next page Refer to Troubleshooting Guide refer to Troubleshooting Guide Refer to Troublesh...

Page 171: ... an HC code This HC code can be looked up in key operator menu 7 Service Actions Error Codes Depending on attributes for Type and Action see error descriptions below error messages will be issued as error code or plain text Type Action Error message will be issued as Defect Replace Message MODXX Please replace module Defect See manual Message EXXXXXX texttext Failure None Message WXXXXXX texttext ...

Page 172: ...traymodule connected Type Defect Context E301000 Action SeeManual Reason XS7 or XS8 not connected Cure1 Connect XS7 or XS8 HC301001 uppertray overruled Type Failure Context W301001 Action None Reason Inputtray overruled dmax 2 0 Cure Replace filmpack HC301002 uppertray disabled Type Failure Context W301002 Action None Reason Inputtray disabled Cure Replace filmpack HC301003 uppertraymodule tagread...

Page 173: ...ype Failure Context W301004 Action None Reason1 S11 not connected Cure1 Connect S11 Reason2 Upper tray module defect Cure2 Replace Part MOD01 HC301005 uppertraymodule s24 Type Failure Context W301005 Action None Reason1 S24 not connected Cure1 Connect S24 Reason2 Upper tray module defect Cure2 Replace Part MOD01 HC301006 uppertraymodule takeoversensors working Type Defect Context E301006 Action Se...

Page 174: ...1007 Action None Reason1 S1 not connected Cure1 Connect S1 Reason2 Upper tray module defect Cure2 Replace Part MOD01 HC301016 upper filmcal required Type Failure Context W301016 Action None Reason Upper tray needs calibration Cure Perform upper tray calibration HC301018 upperTag unsupported Type Accidental Context Action None Reason Wrong version in upper tray Cure Replace filmpack HC301019 tag hi...

Page 175: ...ntext W301020 Action None Reason S11 defect Cure Replace Part MOD01 HC301021 uppertraymodule s24 flagproblem Type Failure Context W301021 Action None Reason S24 defect Cure Replace Part MOD01 HC301022 uppertraymodule sensors Type Failure Context W301022 Action None Reason1 one of the takeoverdetection sensors is not functioning Cure1 Remove jam Reason2 Upper tray module defect Cure2 Replace Part M...

Page 176: ...Type Defect Context E302000 Action SeeManual Reason XS9 not connected to lower tray module Cure1 Connect XS9 to lower tray module HC302001 lowertray overruled Type Failure Context W302001 Action None Reason Inputtray overruled dmax 2 0 Cure Replace filmpack HC302002 lowertray disabled Type Failure Context W302002 Action None Reason Inputtray disabled Cure Replace filmpack HC302003 lowertraymodule ...

Page 177: ...ntext W302007 Action None Reason1 S2 not connected Cure1 Connect S2 Reason2 Lower tray module defect Cure2 Replace Part MOD02 HC302016 lower filmcal required Type Failure Context W302016 Action None Reason Lower tray needs calibration Cure Perform lower tray calibration HC302018 lowerTag unsupported Type Accidental Context Action None Reason Wrong version in lower tray Cure Replace filmpack HC3020...

Page 178: ...ray was already low It was used in another printer Cure Replace filmpack when filmcount runs out HC302025 used film lower Type Failure Context W302025 Action None Reason counter of new filmpack in lower tray was already low It was used in another printer Cure Replace filmpack when filmcount runs out MODULE upperpickupmodule Elabel Yes HC303000 upperpickupmodule film pickupposition Type Accidental ...

Page 179: ... Upper pickup homing failed accidently Cure Self resolving accident HC303003 upperpickupmodule film vacuum Type Accidental Context Action None Reason1 Upper input jam Cure1 Remove upper input jam Reason2 vacuum system failing Cure2 clean vacuum tubes filter vacuum cups Reason3 Lower pickup module defect Cure3 Replace Part vacuum pump HC303004 upperpickupmodule filmpickup Type Accidental Context Ac...

Page 180: ...al HC303006 upperpickupmodule s18 high Type Accidental Context Action None Reason Internal jam Cure Remove internal jam HC303007 upperpickupmodule pickup skew Type Accidental Context Action None Reason Upper input jam Cure Remove upper input jam HC303008 upperpickupmodule tphalign Type Accidental Context Action None Reason Upper input jam Cure Remove upper input jam HC303009 upperpickupmodule film...

Page 181: ...on None Reason Upper pickup homing failed accidently Cure Self resolving accident HC303011 upperpickupmodule thpositioning Type Accidental Context Action None Reason Upper input jam Cure Remove upper input jam HC303012 upperpickupmodule elabel present Type Failure Context W303012 Action None Reason1 Upper pickup elabel not connected Cure1 Connect upper pickup elabel Reason2 Upper pickup Elabel def...

Page 182: ...ontext W303014 Action None Reason1 XS1 disconnected Cure Connect XS1 Reason Upper pickup module defect Cure Replace Part MOD03 HC303015 upperpickupmodule working Type Failure Context W303015 Action None Reason1 Upper pickup module defect Cure1 Replace Part MOD03 Reason2 Film wrinkled in upper tray Cure2 Remove wrinkled film HC303016 upperpickupmodule passed high Type Failure Context W303016 Action...

Page 183: ...lure Context W303021 Action SeeManual Reason1 Upper pickup motor not connected Cure1 reconnect upper pickup motor Reason2 Upper pickup module defect Cure2 Replace Part MOD03 HC303022 upperpickupmodule m7 shortcircuit Type Defect Context E303022 Action SeeManual Reason Upper pickup module defect Cure Replace Part MOD03 HC303023 upperpickupmodule s14 Type Failure Context W303023 Action None Reason1 ...

Page 184: ... None Reason1 Upper pickup pickup sensor S15 malfunctioning Cure1 check upper pickup pickup sensor S15 Reason2 Upper pickup module defect Cure2 Replace Part MOD03 HC303025 upperpickupmodule s18 Type Failure Context W303025 Action None Reason1 Upper pickup passed sensor S18 malfunctioning Cure1 check upper pickup passed sensor S18 Reason2 Upper pickup module defect Cure2 Replace Part MOD03 MODULE l...

Page 185: ... Reason Lower input jam Cure Remove lower input jam HC304002 lowerpickupmodule film homeposition Type Accidental Context Action None Reason Lower input jam Cure Remove lower input jam HC304003 lowerpickupmodule film vacuum Type Accidental Context Action None Reason1 Lower input jam Cure1 Remove lower input jam Reason2 vacuum system failing Cure2 clean vacuum tubes filter vacuum cups Reason3 Lower ...

Page 186: ...ial HC304005 lowerpickupmodule s19 low Type Accidental Context Action None Reason Lower input jam Cure Remove lower input jam HC304006 lowerpickupmodule s19 high Type Accidental Context Action None Reason Internal jam Cure Remove internal jam HC304007 lowerpickupmodule pickup skew Type Accidental Context Action None Reason Lower input jam Cure Remove lower input jam HC304008 lowerpickupmodule tpha...

Page 187: ...ion None Reason Lower input jam Cure Remove lower input jam HC304010 lowerpickupmodule homeposition Type Failure Context W304010 Action None Reason Lower pickup homing failed accidently Cure Self resolving accident HC304011 lowerpickupmodule thpositioning Type Accidental Context Action None Reason Lower input jam Cure Remove lower input jam HC304012 lowerpickupmodule elabel present Type Failure Co...

Page 188: ...owerpickupmodule elabel config Type Failure Context W304013 Action None Reason Lower pickup elabel defect Cure Replace Part MOD04 HC304014 lowerpickupmodule connected Type Failure Context W304014 Action None Reason1 XS4 disconnected Cure Connect XS4 Reason lower pickup module defect Cure Replace Part MOD04 HC304015 lowerpickupmodule working Type Failure Context W304015 Action None Reason1 Lower pi...

Page 189: ...ilure Context W304016 Action None Reason1 Lower pickup past sensor jammed Cure1 Free lower pickup past sensor Reason2 Lower pickup past sensor defect Cure2 Replace Part MOD04 HC304021 lowerpickupmodule m6 connection Type Failure Context W304021 Action SeeManual Reason1 Lower pickup motor not connected Cure1 connect lower pickup motor Reason2 Lower pickup module defect Cure2 Replace Part MOD04 HC30...

Page 190: ...home sensor S16 malfunctioning Cure1 check lower pickup home sensor S16 Reason2 Lower pickup module defect Cure2 Replace Part MOD04 HC304024 lowerpickupmodule s17 Type Failure Context W304024 Action None Reason1 Lower pickup pickup sensor S17 malfunctioning Cure1 check lower pickup pickup sensor S17 Reason2 Lower pickup module defect Cure2 Replace Part MOD04 HC304025 lowerpickupmodule s19 Type Fai...

Page 191: ...oolingfan defect Cure Replace Part MOD16 HC305001 thmodule thpos fail Type Accidental Context Action None Reason Internal jam Cure Resolve jam HC305002 thmodule thermalheadheater Type Failure Context W305002 Action SeeManual Reason1 Connector P3 on PSU not connected Cure1 Connect P3 on PSU Reason2 Connectors P2 and P3 on OSB not connected Cure2 Connect connectors P2 and P3 Reason3 Connectors P6 on...

Page 192: ...AXYS Chapter 3 3 26 of 72 03 2011 Type 5367 100 Agfa Company Confidential Part MOD09 Reason7 TH defect Cure7 Replace Part MOD05 HC305003 thmodule elabel present Type Failure Context W305003 Action None Reason1 TH Elabel not connected Cure1 Connect TH elabel Reason2 Elabel defect Cure2 Replace Part MOD05 HC305004 thmodule elabel config Type Failure Context W305004 Action None Reason Elabel defect C...

Page 193: ...n3 Connectors P3 on TH not connected Cure3 Connect connector P3 on TH Reason4 Connector P2 on TH not connected Cure4 Connect connector P2 on TH Reason5 Connector P1 on TH not connected Cure5 Connect connector P1 on TH Reason6 Connector P43 on MOD08 baseboard Devil or Devius not connected Cure6 Connect connector P43 on MOD08 baseboard Devil or Devius Reason7 PSU defect Cure7 Replace Part MOD09 Reas...

Page 194: ...tion 1 Revision 1 DRYSTAR AXYS Chapter 3 3 28 of 72 03 2011 Type 5367 100 Agfa Company Confidential Cure3 Connect connector P1 on TH Reason4 Connector P43 on MOD08 baseboard Devil or Devius not connected Cure4 Connect connector P43 on MOD08 baseboard Devil or Devius Reason5 PSU defect Cure5 Replace Part MOD09 Reason6 OSB defect Cure6 Replace Part OSB Reason7 TH defect Cure7 Replace Part MOD05 ...

Page 195: ...nnectors P3 on TH not connected Cure3 Connect connector P3 on TH Reason4 Connector P2 on TH not connected Cure4 Connect connector P2 on TH Reason5 Connector P1 on TH not connected Cure5 Connect connector P1 on TH Reason6 Connector P43 on MOD08 baseboard Devil or Devius not connected Cure6 Connect connector P43 on MOD08 baseboard Devil or Devius Reason7 PSU defect Cure7 Replace Part MOD09 Reason8 O...

Page 196: ...9 Action SeeManual Reason TH module heater defect Cure Replace Part MOD05 HC305010 thmodule voltage low Type Failure Context W305010 Action SeeManual Reason1 Serial connection to powersupply fails Cure1 power down the printer and power up again Reason2 PSU module defect Cure2 Replace Part MOD09 Reason3 TH module defect Cure3 Replace Part MOD05 IMPORTANT This fault may have following additional rea...

Page 197: ...esolution Type Defect Context E305012 Action SeeManual Reason TH has the wrong resolution for this printer OR this software is installed on the wrong printer Cure Replace Part MOD05 HC305021 thmodule device thermistors Type Defect Context 05 Action Replace Reason1 Connector P3 on PSU not connected Cure1 Connect P3 on PSU Reason2 Connectors P2 on TH not connected Cure2 Connect connector P2 Reason3 ...

Page 198: ...00 Agfa Company Confidential MODULE outputmodule Elabel Yes HC306000 outputmodule connected Type Defect Context E306000 Action SeeManual Reason1 XS2 not connected Cure1 Connect XS2 Reason2 XS3 not connected Cure2 Connect XS3 HC306001 outputmodule cdm init Type Failure Context W306001 Action None Reason Output module defect Cure Replace Part MOD06 HC306002 outputmodule intensity scan Type Accidenta...

Page 199: ...e outputsensor hi Type Accidental Context Action None Reason Output film jam Cure Remove film jam HC306004 outputmodule printingFail Type Accidental Context Action None Reason1 Internal film jam Cure1 Remove film jam Reason2 S18 or S19 is sticking Cure2 replace this upper or lower tray HC306005 outputmodule film scan Type Accidental Context Action None Reason Internal film jam Cure Remove jam HC30...

Page 200: ...rscan fail Type Accidental Context Action None Reason Internal film jam Cure Remove film jam HC306008 outputmodule outputhome fail Type Accidental Context Action None Reason Internal film jam Cure Remove film jam HC306009 outputmodule cdmhome fail Type Accidental Context Action None Reason Internal film jam Cure Remove film jam HC306010 outputmodule elabel present Type Failure Context W306010 Acti...

Page 201: ...ype Failure Context W306011 Action None Reason Output elabel defect Cure Replace Part MOD06 HC306012 outputmodule cdm home Type Accidental Context Action None Reason CDM homing accidently failed Cure Self resolving accident HC306013 outputmodule home Type Accidental Context Action None Reason1 Output homing accidently failed Cure1 Self resolving accident Reason2 Output module defect Cure2 Replace ...

Page 202: ...ure Context W306015 Action None Reason CDM intensity is unstable Cure Replace CDM module MOD06 HC306020 outputmodule working Type Defect Context 06 Action Replace Reason Output module defect Cure Replace Part MOD06 HC306021 outputmodule m4 connection Type Defect Context E306021 Action SeeManual Reason1 Output motor not connected Cure1 connect output motor Reason2 Output module defect Cure2 Replace...

Page 203: ...ction Type Defect Context E306023 Action SeeManual Reason1 CDM motor not connected Cure1 connect CDM motor Reason2 Output module defect Cure2 Replace Part MOD06 HC306024 outputmodule cdm m5 shortcircuit Type Defect Context E306024 Action SeeManual Reason Output module defect Cure Replace Part MOD06 HC306025 outputmodule s5 Type Defect Context 06 Action Replace Reason Output module defect Cure Repl...

Page 204: ...le defect Cure Replace Part MOD06 HC306028 outputmodule cdm working Type Defect Context 06 Action Replace Reason Output module defect Cure Replace Part MOD06 MODULE drumdrivemodule Elabel No HC307001 drumdrivemodule m1 connection Type Defect Context E307001 Action SeeManual Reason1 M1 not connected Cure1 Connect M1 Reason2 Drum drive module defect Cure2 Replace Part MOD07 IMPORTANT This fault may ...

Page 205: ... 100 Agfa Company Confidential HC307002 drumdrivemodule s9 Type Defect Context E307002 Action SeeManual Reason1 S9 not connected Cure1 Connect S9 Reason2 Drum drive module defect Cure2 Replace Part MOD07 HC307004 drumdrivemodule move Type Accidental Context Action None Reason Drum drive move failed accidently Cure Selfresolving accident HC307020 drumdrivemodule working Type Defect Context 07 Actio...

Page 206: ...ct Context E307021 Action SeeManual Reason Drum drive module defect Cure2 Replace Part MOD07 MODULE devilmodule Elabel Yes HC308000 devilmodule dsh connected Type Defect Context E308000 Action SeeManual Reason1 Connector P9 not connected Cure1 Connect P9 Reason2 MOD08 baseboard Devil or Devius module defect Cure2 Replace Part MOD08 HC308001 devilmodule psu connected Type Defect Context E308001 Act...

Page 207: ...1 Connect P8 Reason2 MOD08 baseboard Devil or Devius module defect Cure2 Replace Part MOD08 HC308003 devilmodule th connected Type Defect Context E308003 Action SeeManual Reason1 Connector P43 not connected Cure1 Connect P43 Reason2 Connector P1 on TH not connected Cure2 Connect P1 on TH Reason3 MOD08 baseboard Devil or Devius module defect Cure3 Replace Part MOD08 Reason4 TH defect Cure4 Replace ...

Page 208: ...eeManual Reason1 Connector P21 not connected Cure1 Connect P21 Reason2 MOD08 baseboard Devil or Devius module defect Cure2 Replace Part MOD08 HC308005 devilmodule p20 connected Type Defect Context E308005 Action SeeManual Reason1 Connector P20 not connected Cure1 Connect P20 Reason2 MOD08 baseboard Devil or Devius module defect Cure2 Replace Part MOD08 HC308006 devilmodule p41 connected Type Defec...

Page 209: ... Reason2 MOD08 baseboard Devil or Devius module defect Cure2 Replace Part MOD08 HC308008 devilmodule p1 connected Type Defect Context E308008 Action SeeManual Reason1 Connector P1 not connected on MOD08 baseboard Devil or Devius Cure1 Connect P1 on MOD08 baseboard Devil or Devius Reason2 MOD08 baseboard Devil or Devius module defect Cure2 Replace Part MOD08 HC308009 devilmodule elabel present Type...

Page 210: ... Type Failure Context W308010 Action None Reason MOD08 baseboard Devil or Devius Elabel defect Cure Replace Part MOD08 HC308011 devilmodule power 3v3 Type Defect Context E308011 Action SeeManual Reason1 MOD08 baseboard Devil or Devius module defect Cure1 Replace Part MOD08 Reason2 PSU module defect Cure2 Replace Part MOD09 HC308012 devilmodule power 12v Type Defect Context E308012 Action SeeManual...

Page 211: ...on SeeManual Reason1 MOD08 baseboard Devil or Devius module defect Cure1 Replace Part MOD08 Reason2 PSU module defect Cure2 Replace Part MOD09 HC308014 devilmodule power 24v Type Defect Context E308014 Action SeeManual Reason1 MOD08 baseboard Devil or Devius module defect Cure1 Replace Part MOD08 Reason2 PSU module defect Cure2 Replace Part MOD09 HC308015 devilmodule power actuator Type Defect Con...

Page 212: ...1 on TH Reason2 Connector P43 not connected on MOD08 baseboard Devil or Devius Cure2 Connect P43 on MOD08 baseboard Devil or Devius Reason3 Connector P3 not connected on MOD08 baseboard Devil or Devius Cure3 Connect P3 on PSU Reason4 PSU module defect Cure4 Replace Part MOD09 Reason5 MOD08 baseboard Devil or Devius module defect Cure5 Replace Part MOD08 Reason6 TH module defect Cure6 Replace Part ...

Page 213: ... sn mismatch Type Defect Context E308019 Action SeeManual Reason SN mismatch Cure Call service HC308020 devilmodule power sensor Type Defect Context E308020 Action SeeManual Reason1 MOD08 baseboard Devil or Devius module defect Cure1 Replace Part MOD08 Reason2 PSU module defect Cure2 Replace Part MOD09 HC308021 devilmodule p42 connected Type Defect Context E308021 Action SeeManual Reason1 Connecto...

Page 214: ... continuous printing this can be cured by waiting until temperature comes down otherwise it has to be made sure that the environment temperature comes down Reason2 Cooling of MOD12 CPU board Thanos or Tyndall not efficient because of blocked airflow Cure2 Make sure that enough airflow is available for the MOD12 CPU board Thanos or Tyndall by insuring that air inlets and outlets are completely free...

Page 215: ...ontext 08 Action Replace Reason MOD08 baseboard Devil or Devius module defect Cure Replace Part MOD08 HC308027 devilmodule fpga Type Defect Context 08 Action Replace Reason MOD08 baseboard Devil or Devius module defect Cure Replace Part MOD08 HC308028 devilmodule m2 driver Type Defect Context 08 Action Replace Reason MOD08 baseboard Devil or Devius module defect Cure Replace Part MOD08 HC308029 de...

Page 216: ...ontext 08 Action Replace Reason MOD08 baseboard Devil or Devius module defect Cure Replace Part MOD08 HC308031 devilmodule m5 driver Type Defect Context 08 Action Replace Reason MOD08 baseboard Devil or Devius module defect Cure Replace Part MOD08 HC308032 devilmodule f1 driver Type Defect Context 08 Action Replace Reason MOD08 baseboard Devil or Devius module defect Cure Replace Part MOD08 HC3080...

Page 217: ...text W308034 Action None Reason Diagnose not complete Cure Reboot printer with closed trays and covers HC308035 devilmodule m6 driver Type Defect Context 08 Action Replace Reason MOD08 baseboard Devil or Devius module defect Cure Replace Part MOD08 HC308036 devilmodule m7 driver Type Defect Context 08 Action Replace Reason MOD08 baseboard Devil or Devius module defect Cure Replace Part MOD08 HC308...

Page 218: ...ntext 08 Action Replace Reason MOD08 baseboard Devil or Devius module defect Cure Replace Part MOD08 HC308039 devilmodule v1 driver Type Defect Context 08 Action Replace Reason MOD08 baseboard Devil or Devius module defect Cure Replace Part MOD08 HC308040 devilmodule v2 driver Type Defect Context 08 Action Replace Reason MOD08 baseboard Devil or Devius module defect Cure Replace Part MOD08 HC31202...

Page 219: ...rial Type Defect Context 12 Action Replace Reason MOD12 cpu board Thanos or Tyndall defect Cure Replace Part MOD12 HC312022 devilmodule tyndall ethernet Type Defect Context 12 Action Replace Reason MOD12 cpu board Thanos or Tyndall defect Cure Replace Part MOD12 HC312023 devilmodule tyndall portex Type Defect Context 12 Action Replace Reason Software problem Cure Reinstall software HC312024 devilm...

Page 220: ... Defect Context 18 Action Replace Reason1 RAM defect Cure1 Replace Part MOD18 Reason2 MOD12 cpu board Thanos or Tyndall defect Cure2 Replace Part MOD12 HC312025 devilmodule tyndall clock Type Defect Context 12 Action Replace Reason MOD12 cpu board Thanos or Tyndall defect Cure Replace Part MOD12 MODULE powersupplymodule Elabel Yes HC309000 powersupplymodule temperature Type Defect Context E309000 ...

Page 221: ...ilure Context W309001 Action None Reason1 PSU Elabel not connected Cure1 Connect PSU elabel Reason2 PSU module Elabel defect Cure2 Replace Part MOD09 HC309002 powersupplymodule elabel config Type Failure Context W309002 Action None Reason PSU module Elabel defect Cure Replace Part MOD09 HC309003 powersupplymodule powerth current Type Defect Context E309003 Action SeeManual Reason1 TH power cables ...

Page 222: ...ntial HC309004 powersupplymodule powermotor current Type Defect Context E309004 Action SeeManual Reason1 Actuators power cables short circuited Cure1 Check actuators power cables Reason2 PSU module defect Cure2 Replace Part MOD09 HC309005 powersupplymodule temperature warning Type Failure Context W309005 Action None Reason PSU temperature high Cure Cool down the PSU HC309020 powersupplymodule powe...

Page 223: ...ompany Confidential HC309022 powersupplymodule powermotor Type Defect Context 09 Action Replace Reason1 Actuators power cables short circuited Cure1 Check actuators power cables Reason2 PSU module defect Cure2 Replace Part MOD09 HC309024 powersupplymodule powerlogic Type Defect Context 09 Action Replace Reason PSU module defect Cure Replace Part MOD09 HC309025 powersupplymodule power3v3 Type Defec...

Page 224: ...al HC309026 powersupplymodule power12v Type Defect Context 09 Action Replace Reason PSU module defect Cure Replace Part MOD09 HC309027 powersupplymodule powern12v Type Defect Context 09 Action Replace Reason PSU module defect Cure Replace Part MOD09 MODULE displaykeypadmodule Elabel Yes HC310000 displaykeypadmodule elabel present Type Failure Context W310000 Action None Reason1 Display Elabel not ...

Page 225: ...idential HC310001 displaykeypadmodule elabel config Type Failure Context W310001 Action None Reason Display Elabel defect Cure Replace Part MOD10 MODULE inputsensormodule Elabel No HC311000 inputsensormodule s4 flagproblem Type Failure Context W311000 Action None Reason Internal jam input sensor S4 isn t responding Cure Remove jam HC311001 inputsensormodule s13 flagproblem Type Failure Context W31...

Page 226: ... Confidential HC311002 inputsensormodule connected Type Defect Context E311002 Action SeeManual Reason XS5 not connected Cure Connect XS5 HC311003 inputsensormodule s4 Type Failure Context W311003 Action None Reason1 S4 not connected Cure1 Connect S4 Reason2 S4 bracket module defect Cure2 Replace Part MOD11 HC311004 inputsensormodule s13 Type Failure Context W311004 Action None Reason1 S13 not con...

Page 227: ...0 Agfa Company Confidential HC311005 inputsensormodule sensors Type Failure Context W311005 Action None Reason one of the inputsensors is not functioning Cure Remove jam HC311020 inputsensormodule working Type Defect Context 11 Action Replace Reason S4 bracket module is defective Cure Power off printer and replace Part MOD10 MODULE camdrivemodule Elabel No HC313000 camdrivemodule home Type Acciden...

Page 228: ...3001 camdrivemodule m2 connection Type Defect Context E313001 Action SeeManual Reason1 Motor M2 not connected Cure1 Connect motor M2 Reason2 Cam shaft module defect Cure2 Replace Part MOD13 HC313002 camdrivemodule s7 Type Defect Context E313002 Action SeeManual Reason1 Sensor S7 not connected Cure1 Connect sensor S7 Reason2 Cam shaft module defect Cure2 Replace Part MOD13 HC313020 camdrivemodule w...

Page 229: ... 03 2011 Type 5367 100 Agfa Company Confidential HC313021 camdrivemodule m2 shortcircuit Type Defect Context E313021 Action SeeManual Reason Cam shaft module defect Cure Replace Part MOD13 MODULE gearboxmodule Elabel No HC314000 gearboxmodule home Type Accidental Context Action None Reason Internal jam Cure Remove internal jam HC314001 gearboxmodule pickup passed Type Accidental Context Action Non...

Page 230: ... Agfa Company Confidential HC314002 gearboxmodule takeover film close Type Accidental Context Action None Reason Internal jam Cure Remove internal jam HC314003 gearboxmodule film inputsensors Type Accidental Context Action None Reason Internal jam Cure Remove internal jam HC314004 gearboxmodule m8 connection Type Defect Context E314004 Action SeeManual Reason1 Motor M8 not connected Cure1 Connect ...

Page 231: ...oxmodule s3 Type Defect Context E314005 Action SeeManual Reason1 Sensor S3 not connected Cure1 Connect sensor S3 Reason2 Roller drive module defect Cure2 Replace Part MOD14 HC314006 gearboxmodule connected Type Defect Context E314006 Action SeeManual Reason1 Connector XS12 not connected Cure1 Connect Connector XS12 Reason2 gearboxmodule drive module defect Cure2 Replace Part MOD14 HC314020 gearbox...

Page 232: ...8 shortcircuit Type Defect Context E314021 Action SeeManual Reason Roller drive module defect Cure Replace Part MOD14 MODULE suspensiondrivemodule Elabel No HC315000 suspensiondrivemodule move Type Accidental Context Action None Reason Suspension move 10 or 14 inch failed accidently Cure Self resolving accident HC315001 suspensiondrivemodule m3 connection Type Defect Context E315001 Action SeeManu...

Page 233: ... Action SeeManual Reason1 Sensor S21 not connected Cure1 Connect sensor S21 Reason2 Suspension drive module defect Cure2 Replace Part MOD15 HC315003 suspensiondrivemodule s22 Type Defect Context E315003 Action SeeManual Reason1 Sensor S22 not connected Cure1 Connect sensor S22 Reason2 Suspension drive module defect Cure2 Replace Part MOD15 HC315004 suspensiondrivemodule connected Type Defect Conte...

Page 234: ...ondrivemodule working Type Defect Context 15 Action Replace Reason Suspension drive module defect Cure Replace Part MOD15 HC315021 suspensiondrivemodule m3 shortcircuit Type Defect Context E315021 Action SeeManual Reason Suspension drive module defect Cure Replace Part MOD15 MODULE coolingfanmodule Elabel No HC316000 coolingfanmodule thf1 connection Type Defect Context E316000 Action SeeManual Rea...

Page 235: ...Coolingfan not connected Cure1 Connect coolingfan Reason2 Coolingfan module defect Cure2 Replace Part MOD12 MODULE pumpandvalvesmodule Elabel No HC317000 pumpandvalvesmodule m9 connection Type Defect Context E317000 Action SeeManual Reason1 Motor M9 not connected Cure1 Connect motor M9 Reason2 XS10 not connected Cure2 Connect XS10 Reason3 Pump and valves module defect Cure3 Replace Part MOD17 HC31...

Page 236: ...C317002 pumpandvalvesmodule v2 connection Type Defect Context E317002 Action SeeManual Reason1 XS10 not connected Cure1 Connect XS10 Reason2 Pump and valves module defect Cure2 Replace Part MOD17 HC317020 pumpandvalvesmodule m9 shortcircuit Type Defect Context E317020 Action SeeManual Reason Pump and valves module defect Cure Replace Part MOD17 HC317021 pumpandvalvesmodule v1 shortcircuit Type Def...

Page 237: ...esmodule v2 shortcircuit Type Defect Context E317022 Action SeeManual Reason Pump and valves module defect Cure Replace Part MOD17 MODULE eboxfanmodule Elabel No HC320000 eboxfanmodule eboxfan connection Type Defect Context E320000 Action SeeManual Reason1 EBox fan module not connected Cure1 Connect EBox fan module Reason2 EBox fan module defect Cure2 Replace Part MOD20 HC320020 eboxfanmodule ebox...

Page 238: ... of this document is uncontrolled Edition 1 Revision 1 DRYSTAR AXYS Chapter 3 3 72 of 72 03 2011 Type 5367 100 Agfa Company Confidential MODULE bothtraysmodule Elabel No HC321001 pickupmodule alltrays disabled Type Defect Context E321001 Action SeeManual Reason All inputtrays have detected a mallfunction Cure Check other active HC codes to get detailed info ...

Page 239: ...d 10 2007 printed in Germany Agfa Company Confidential Document Node ID 14349630 eq_03 4_codings_e_template_v04 Copyright 2007 Agfa HealthCare N V DRYSTAR AXYS Type 5367 100 Purpose of this document This document lists the jumper settings of the printer and describes the meaning of the LEDs on the different printed circuit boards Document History Edition Revision Release Date Changes 1 0 10 2007 I...

Page 240: ...ce activities may cause damage or injuries INSTRUCTION 1 Read the Generic Safety Directions document see MEDNET GSO General Info Agfa HealthCare Publications Service Manual prior to attempting any operation repair or maintenance task on the equipment 2 Strictly observe all safety directions within the Generic Safety Directions and on the product Edition 1 Revision 0 DRYSTAR AXYS Chapter 3 4 2 10 2...

Page 241: ...s document is uncontrolled Edition 1 Revision 0 DRYSTAR AXYS Chapter 3 4 3 10 2007 Type 5367 100 Agfa Company Confidential LIST OF CONTENTS 1 CODINGS 4 1 1 Jumper Setting 73 GB Maxtor Hard Disk 4 1 2 Jumper Setting 73 GB Hitachi Hard Disk 5 2 FUSES 6 3 LEDS 6 3 1 DEVIL Board 7 3 2 THANOS Board plugged onto the DEVIL Board 8 3 3 Power Supply 9 3 4 Front Panel 10 3 5 Rear panel 11 ...

Page 242: ...r J1 Setting Function 1 2 open ID Bit 3 3 4 open ID Bit 2 5 6 open ID Bit 1 7 8 open ID Bit 0 9 10 open Not used 11 12 open Termination Power Table 1 Jumper J3 Setting Function 1 2 open ID Bit 3 3 4 open ID Bit 2 5 6 open ID Bit 1 7 8 open ID Bit 0 9 10 open Busy LED 11 12 open Enable write protect 13 14 open enable stagger spin 15 16 open Enable delay spin 17 18 open Force single ended bus mode T...

Page 243: ...36634BC CDR 1 12 1 24 C3 C4 Figure 2 Jumper C3 Setting Function 1 2 open ID Bit 3 3 4 open ID Bit 2 5 6 open ID Bit 1 7 8 open ID Bit 0 9 10 open Disable Autostart 11 12 open Force single ended bus mode 13 14 open Enable delayed start 15 16 open Write Protect 17 18 open Reserved 19 20 open Reserved 21 22 open LED 23 24 open Term Power to SCSI bus Table 3 Edition 1 Revision 0 DRYSTAR AXYS Chapter 3...

Page 244: ...4 Setting Function 1 2 open ID Bit 3 3 4 open ID Bit 2 5 6 open ID Bit 1 7 8 open ID Bit 0 9 10 open Not used 11 12 open Write Protect Table 4 2 Fuses The printer has no fuses that can be replaced If one of the fuses in the power supply is defective the whole Power supply unit must be replaced refer to section 3 5 3 LEDs The following tables show the status of the LED indications and the correspon...

Page 245: ... 3V3 D5 green 3 3V present P_24V D9 green 24V present 12V D10 green 12V present IMGCTRL_CONFDONE D12 yellow Off FPGA config done ON FPGA not configured IMGCTRL_1 D13 yellow Software controllable UP_LED0 D14 yellow Software controllable UP_LED1 D15 yellow Software controllable IDE_DASP D16 yellow IDE activity 12V D17 green 12V present 5V D18 green 5V present CP_ACTLED D19 yellow Disc activity contr...

Page 246: ...2 yellow FPGA LED1 Software controllable LD3 yellow FPGA LED1 Software controllable LD5 green VCC supply voltage ok LD6 green 3 3V supply voltage ok LD8 green 1V8 supply voltage ok LD9 green 1V8 DDR2 supply voltage ok LD10 green VCORE supply voltage ok LD11 yellow 1V2 supply voltage ok LD12 yellow PCI Bus status LD13 yellow Fire Wire Link active LD14 yellow Fire Wire cable connected LD15 yellow CP...

Page 247: ...3 3 V not present 5 V On Off Green 5 V present 5 V not present 12 V On Off Green 12 V present 12 V not present 12 V On Off Green 12 V present 12 V not present 24 V On Off Green 24 V present 24 V not present 13 27 V On Off Green voltage between 13 V and 27 V voltage 13 V When the temperature inside the power supply exceeds the permissible limit printing can continue for up to 120 seconds After this...

Page 248: ...gfa Company Confidential 3 4 Front Panel Led Color Function D1 green 12V power present D2 D25 green red See table below D3 green 5V power present Color Light Status Receive Jobs Printing possible Constant Ready stand by No No Startup sequence No No Receiving jobs Yes Yes Calculating Yes Yes Printing Yes Yes Calibrating Yes No Installation No No Green Blinking Key operator Yes No Yes No Service mod...

Page 249: ...ument is uncontrolled 3 5 Rear panel 536634aa cdr Figure 4 Name Led Color Function NTRAFFIC D1 yellow Network activity NGOOD_LINK D2 yellow Network connected 1 GBit_speed D3 yellow High speed network IDE_DASP D16 yellow IDE activity Note These LEDs are part of the DEVIL board refer to 3 1 Edition 1 Revision 0 DRYSTAR AXYS Chapter 3 4 11 10 2007 Type 5367 100 Agfa Company Confidential ...

Page 250: ......

Page 251: ...ed in Germany Agfa Company Confidential Document Node ID 14350805 eq_03 5_replacements_e_template_v07 Copyright 2010 Agfa HealthCare N V DRYSTAR AXYS Type 5367 100 Purpose of this Document This document describes the availability of replacement instructions The replacement instructions are enclosed to the printer spare parts as print out Document History Edition Revision Release Date Changes compa...

Page 252: ...marks of Agfa HealthCare N V WARNING Improper operation or service activities may cause damage or injuries INSTRUCTION 1 Read the Generic Safety Directions document see MEDNET GSO General Info Agfa HealthCare Publications Service Manual prior to attempting any operation repair or maintenance task on the equipment 2 Strictly observe all safety directions within the Generic Safety Directions and on ...

Page 253: ...Company Confidential Document Node ID 14350417 eq_03 6_adjustments_e_template_v06 Copyright 2008 Agfa HealthCare N V DRYSTAR AXYS Type 5367 100 Purpose of this document This document describes the adjustments calibrations and image quality checks of the printer Document History Edition Revision Release Date Changes compared to previous Version 1 0 1 1 06 2008 Added section 2 5 Compliance Film Cali...

Page 254: ...re N V All rights reserved Technical modifications reserved AGFA and the Agfa Rhombus are trademarks of Agfa HealthCare N V WARNING Improper operation or service activities may cause damage or injuries INSTRUCTION 1 Read the Generic Safety Directions document see MEDNET GSO General Info Agfa HealthCare Publications Service Manual prior to attempting any operation repair or maintenance task on the ...

Page 255: ...ration 14 2 5 Compliance Film Calibration as part of Afssaps EUREF Calibration Procedure 18 2 5 1 Checking the Densitometer Calibration 21 2 5 2 Checking the Densitometer Configuration in the Printer 21 2 5 3 Resetting the Compliance Film Calibration Internal Densitometer Calibration and Thermal Print Head Calibration 22 2 5 4 Performing Internal Densitometer Calibration Film Calibrations and Ther...

Page 256: ... is 1 5 mm for DT2B DT2C films or 1 mm for DT 2 Mammo Films 1 1 1 1 Horizontal Film Alignment Procedure WHEN TO BE PERFORMED The horizontal film alignment is necessary when The TPH has been replaced and the blank frame at the left and right hand side on a SMPTE test image is o 1 5 mm for DT2B DT2C Films o 1 mm for DT 2 Mammo Films PURPOSE Centering the image on the film Compensate for mechanical t...

Page 257: ...Session 3 Select 6 Calibration 4 Horizontal Alignment 4 Select upper tray or lower tray A test film as shown in Figure 1 will be printed 5 Measure the left and right blank border of the test film 6 Enter the values in mm typically approx 10 mm in the browser screen 7 Select Accept Changes Via Local Keypad 1 In the key operator menu select menu 7 Service Actions 2 Select 6 Calibration 4 Horizontal ...

Page 258: ...owser User Interface Measure left white border via local display Measure right white border 0 0 mm 0 0 mm Welcome to this DryStar AXYS Left white border mm Right white border mm Service Calibration Horizontal alignment Accept changes Discard changes 0 0 0 0 536736ac cdr Figure 1 Verification 1 On the local keypad select key operator menu 3 Print Image 1 Testimage upper or lower tray SMPTE_xxx tif ...

Page 259: ...r replacement of the Thermal print head TPH During the troubleshooting procedure 2 1 Internal densitometer CDM calibration When the density reproduction check indicates to calibrate the CDM During the troubleshooting procedure 2 2 Print Head TH calibration During the troubleshooting procedure During the TH cleaning procedure Refer to section 3 2 in this chapter 2 3 Smart Profile Calibration During...

Page 260: ... 2 5 After TH replacement At least each 5000 films adjustable maintenance action Refer to chapter 9 Maintenance In case of a drift of the maximum density During the troubleshooting procedure PURPOSE Determination of the correlation between control of the thermal head and corresponding densities on film REQUIRED TOOLS None REQUIRED TIME 10 min NOTE The film calibration is by default executed automa...

Page 261: ...asured Dmax 3 Repeat film calibration for the second input tray 536336as cdr Figure 2 NOTE The calibration is successful in case the measured Dmax is max 2 different from the target Dmax The target Dmax depends on the default densitometer which is configured in Key operator menu 2 Change Settings 5 Image Quality 1 General Refer to table below Example Selected default densitometer X Rite 341 Measur...

Page 262: ...acturer WHEN TO BE PERFORMED The calibration of the internal densitometer CDM via an external densitometer is necessary when the density reproduction check for a specific material is not OK after a film calibration During the troubleshooting procedure PURPOSE Adjust the CDM to another calibrated external densitometer To compensate for CDM measurement drifts REQUIRED TOOLS One of following calibrat...

Page 263: ...tion Internal Densitometer Calibration Welcome to this DryStar Axys printer Macbeth TR924 Gretag X Rite 301 X Rite 310 Start Calibration Go back Reset Values Welcome to this DryStar Axys printer SUCCEEDED Enter below the measurements of the printer calibration sheet Step Previous values Step Previous values 00 0 25 01 0 25 02 0 26 03 0 30 04 0 54 05 0 94 06 1 43 07 1 96 08 2 50 09 3 00 10 3 18 11 ...

Page 264: ...ble Vertical stripes are unacceptable in case they are visible in the medical service test image REQUIRED TOOLS One of following calibrated densitometers McBeth TR 924 Gretag D200 2 X Rite 301 X Rite 310 X Rite 331 X Rite 341 Lint free cloth Isopropyl alcohol INSTRUCTIONS 1 Print the Medical service test image 2 Clean the thermal head with a cleaning tissue lightly moistened with Isopropyl alcohol...

Page 265: ...inter Service Calibration Printhead Calibration Upper Tray Lower Tray Select a 14 film for optimum quality Welcome to this DryStar Axys printer 536736ag cdr Select Densitometer Service Calibration Thermal Printhead Calibration Welcome to this DryStar Axys printer Macbeth TR924 Gretag X Rite 301 X Rite 310 Start Calibration Go back Reset Values Welcome to this DryStar Axys printer Enter below the m...

Page 266: ... respect to the first printed medical image replace the Thermal Print head refer to Section 3 5 7 Perform a Film calibration Refer to section 2 1 Figure 5 Medical service test image NOTE With Print Head Calibration only macro TPH banding can be solved Sharp and or small 10 mm discontinuities can not be solved by a Print Head Calibration 2 4 Smart Profile Calibration WHEN TO BE PERFORMED The smart ...

Page 267: ...ervice Engineer Tools Enter Service Command Session 3 Select 6 Calibration 5 Smart Profile Calibration 4 Select the densitometer which will be used for measurement 5 Select the film tray with 14x17 14x14 or 14x11 films A test film as shown in Figure 6 is printed 6 Measure the 66 squares on the printed film Write down the values on the foreseen squares on the film with a waterproof pen Figure 6 7 E...

Page 268: ...beth TR924 Gretag X Rite 301 X Rite 310 Start Calibration Go back Reset Values Welcome to this DryStar Axys printer Smart Profile Calibration Smart Profile Calibration Select the input tray with 14 x17 film format Smart Profile Calibration Step Step Step Step Step 00 01 02 03 04 05 06 07 08 09 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 ...

Page 269: ...Engineer Tools Enter Service Command Session 3 Print Image 1 Testimage upper or lower tray STI2 ZIP The STI2 test image will be printed 2 From the same menu print the medical_image zip IMPORTANT Select the tray with 14 width films 3 Check both images at the light box for dark stripes in the area of the 10 film edge On the STI2 test image dark stripes may be visible in the area of the 10 film edge ...

Page 270: ...ed in following cases Before the Afssaps EUREF mammo acceptance test Each 6 months if the printer is under control of Afssaps EUREF and it did not pass the Afssaps EUREF calibration check see section 2 5 6 Afssaps Agence française de sécurité sanitaire des produits de santé EUREF European Reference Organisation for Quality Assured Breast Screening and Diagnostic Services PREREQUISITE 1 The complia...

Page 271: ...rk interface and browser Internet Explorer Calibrated densitometer MacBeth TR 924 X Rite 331 341 X Rite 301 X Rite 310 or Gretag D200 2 Lint free cloth Isopropyl alcohol Waterproof pen Density reproduction check Excel sheets available via MEDNET GSO Library Afssaps EUREF Calibration Check Excel Sheet available via MEDNET GSO Library Connectivity Release Document for the corresponding modality to D...

Page 272: ...nter installation date today DRC ok Deviations 10 Check calibration of densitometer Details see section 2 5 1 Check configured default densitometer Details see section 2 5 2 Used densitometer default densitometer in the printer Reset CDM TH CFC calibrations Details see section 2 5 3 Print TG18 QPC pattern Measure 18 density values Enter values in EUREF AFSSAPS GSDF check xls Details see 2 5 6 Perf...

Page 273: ...1 Show settings 3 Network DICOM 2 Write down the printer IP address 3 Leave the key operator menu 4 Confirm that the printer is in ready state i e no films are in the printer queue 5 Disconnect the network cable at the wall side 6 Connect the network cable to the service PC See NOTE below 7 Adapt the service PC network address if required For details refer to chapter 3 2 Tools section Connection o...

Page 274: ... calibration for lower tray depends on which film tray is using DT2M film material If both trays contain DT2M film material select both trays one after another 4 Leave the service command session 5 In the service tools page select Reset calibration data 6 Select Default CDM and Default Profile Note that Default Profile button is not available if no Print Head Profile has been performed in the fiel...

Page 275: ...DT2M films The calibration film as shown in Figure 12 is printed and the measured Dmax is shown INPUT TRAY FILM CALIBRATION OK Dmax 3 69 XR310 Default densitometer 536736az cdr Measured Dmax 3 Repeat the film calibration a 2nd time 4 Repeat the film calibration two times for the second input tray if it keeps DT2M films too Figure 12 2 5 4 2 Performing an internal Densitometer Calibration 1 Select ...

Page 276: ... Axys printer 536736ak cdr Select Densitometer Service Calibration Internal Densitometer Calibration Welcome to this DryStar Axys printer Macbeth TR924 Gretag X Rite 301 X Rite 310 Start Calibration Go back Reset Values Welcome to this DryStar Axys printer SUCCEEDED Enter below the measurements of the printer calibration sheet Step Previous values Step Previous values 00 0 25 01 0 25 02 0 26 03 0 ...

Page 277: ... shown INPUT TRAY FILM CALIBRATION OK Dmax 3 69 XR310 Default densitometer 536736az cdr Measured Dmax 3 Repeat the film calibration for the second input tray if it keeps DT2M films too Figure 14 2 5 4 4 Performing a Print Head Calibration followed by two Film Calibrations NOTE The print head calibration has to be performed 2 times if films of width 10 and 14 are available INSTRUCTIONS 1 Clean the ...

Page 278: ...on Internal Densitometer Calibration Thermal Print Head Calibration Horizontal Alignment Smart profile calibration Go back Service Calibration Welcome to this DryStar Axys printer Service Calibration Printhead Calibration Upper Tray Lower Tray Select a 14 film for optimum quality Welcome to this DryStar Axys printer 536736ag cdr Select Densitometer Service Calibration Thermal Printhead Calibration...

Page 279: ...are Measured density Target density Density diff Check 100 21 21 0 0 90 56 55 4 0 6 0 80 91 89 8 1 2 0 70 127 124 2 2 8 0 Density Reproduction OK 60 161 158 6 2 4 0 No CDM calibration needed for Mammo film 50 195 193 2 0 40 229 227 4 1 6 0 30 263 261 8 1 2 0 20 298 296 2 1 8 0 10 333 330 6 2 4 0 0 368 365 3 0 When entering the measured density values the decimal point appears in a different spot e...

Page 280: ...55 342 479 0 457 3 9 6 1200 36 13 348 614 0 440 0 484 0 396 6 2 08 34 270551 12 7 1440 51 35702 390 683 0 399 2 6 7 1440 54 76 398 640 0 410 0 451 0 369 7 1 9 51 357016 11 8 1680 75 48349 439 328 0 380 1 8 8 1680 81 09 448 666 0 388 0 426 0 349 8 1 73 75 483485 10 9 1920 111 1691 490 605 0 382 3 1 9 1920 117 99 498 692 0 371 0 408 0 334 9 1 56 111 16915 9 10 2160 163 9521 544 108 0 384 7 2 10 2160...

Page 281: ...described steps you performed already neglect the instructions in this screen 3 5 Select the tray with DT2 Mammo films If both trays keep DT2 Mammo films of different format perform the compliance film calibration for one tray and repeat it for the other tray 4 6 Select the configured densitometer Refer to section 2 5 1 5 7 Select Start calibration 8 Enter Dmin and Dmax measured during step 2 5 6 ...

Page 282: ...e the service command session 14 Repeat step 2 5 6 Performing the Afssaps EUREF Calibration Check If this time all Deviation compared to DICOM values are 10 the Afssaps EUREF calibration procedure is finished If one or more Deviation compared to DICOM value s are 10 repeat steps 2 5 6 and 2 5 7 up to three times See IMPORTANT note below IMPORTANT If after three repetitions one or more Deviation co...

Page 283: ...66 1 83 2 00 2 16 2 33 2 50 2 66 2 83 3 00 Service Calibration Compliance film calibration Select densitometer Macbeth TR924 Gretag X Rite 301 X Rite 310 Start calibration Go back Reset values 5367_SIB04_02 CDR Service Calibration Dmin 0 20 Dmax 3 68 Reflected Ambient Light 1 View Box Illumination 4000 Number of steps 18 EUREF Start calibration Go back Reset values Service Calibration Compliance f...

Page 284: ...ilable via MEDNET path GSO library Hardcopy General Hardcopy Info General Services Info Application Info Density Reproduction Check Excel Sheet PURPOSE With the density reproduction check the difference for a specific film type between optical density defined in the sensitometric calibration file and the printed optical density measured with an external densitometer is checked This is a plausibili...

Page 285: ... 0 Density Lower limit Upper limit 0 0 0 30 3 3 45 4 4 60 5 5 75 6 6 90 7 7 X rite 310 105 8 8 X rite 331 120 9 9 X rite 341 135 9 9 X rite 301 150 9 9 Mac Beth TR924 165 9 9 Gretag D200 2 180 10 10 195 10 10 210 10 10 225 10 10 240 10 10 255 11 11 270 11 11 285 11 11 300 11 11 315 12 12 Density difference SMPTE image DT2B film DSAXYS 15 10 5 0 5 10 15 0 50 100 150 200 250 300 350 Target Density D...

Page 286: ... test image Perform Dry TH Cleaning Procedure Clean TH with Isopropyl alcohol wet cleaning Vertical stripes visible Perform Resistor Check Perform TH profile calibration Dry TH cleaning procedure Dismount TH Mount TH Use ink eraser for cleaning Refer to figure next page Print STI2 test image New vertical stripes visible No Service Engineer Tools Enter Service Command Session 3 Print Testimage 1 Te...

Page 287: ...ions o Wet cleaning the last cleaning action must be carried out in one way o Dry cleaning Dismount the thermal head before dry cleaning Make a linear cleaning movement parallel to the resistor line Refer to Figure 21 General do not push on the IC cover of the thermal head during cleaning 536410cd cdr TH resistor line TH top view linear cleaning movement parallel to TH resistor line Ink eraser e g...

Page 288: ... it s not necessary to perform a dry cleaning The Print Head can be replaced right away The resistor check is performed in 2 or more steps 1 A rough resistor measurement of all the resistor values If large resistor drifts show up a message will show up to replace the Print Head 2 A fine resistor measurement in one of 44 predefined areas of the image see Figure 22 The Service Engineer can enter thi...

Page 289: ...MENT CONTROL NOTE The controlled version of this document resides on MedNet Any printed copy of this document is uncontrolled Edition 1 Revision 1 DRYSTAR AXYS Chapter 3 6 37 06 2008 Type 5367 100 Agfa Company Confidential Figure 22 Print Head resistor check test film ...

Page 290: ......

Page 291: ...this document is uncontrolled 10 2007 printed in Germany Agfa Company Confidential Document Node ID Node ID eq_03 7_menus setting_e_template_v04 Copyright 2007 Agfa HealthCare N V DRYSTAR AXYS 5367 100 Purpose of this Document This document lists the menus for service of the printer Document History Edition Revision Release Date Changes 1 0 10 2007 Initial version Referenced Documents Document Tit...

Page 292: ...3 7 2 10 2007 5367 100 Agfa Company Confidential WARNING Improper operation or service activities may cause damage or injuries INSTRUCTION 1 Read the Generic Safety Directions document see MEDNET GSO General Info Agfa HealthCare Publications Service Manual prior to attempting any operation repair or maintenance task on the equipment 2 Strictly observe all safety directions within the Generic Safet...

Page 293: ...des on MedNet Any printed copy of this document is uncontrolled Edition 1 Revision 0 DRYSTAR AXYS Chapter 3 7 3 10 2007 5367 100 Agfa Company Confidential LIST OF CONTENTS 1 PRINTER CONTROLS 4 1 1 Controls at the local Keypad 5 2 KEY OPERATOR MENU VIA LOCAL KEYPAD 6 3 KEY OPERATOR BROWSER PAGES 7 4 SERVICE BROWSER PAGES 11 ...

Page 294: ... At the right side of the display a LED indicates the status of the Drystar Axys Calibrating Yes No Installation No No Key operator Yes No Yes No Red Blinking Service mode No No Input tray empty Yes Yes No Film jam Yes No Warning Yes Yes Constant Error status No No Color Light Status Receive jobs Printing OK Action Green Constant Ready standby Yes Yes Proceed Blinking Startup sequence No No Wait R...

Page 295: ...try in a menu Up key T o move the cursor to the previous entry field T o scroll upwards T o increment the number in a n alpha numerical entry field Down key T o move the cursor to the next entry field T o scroll downwards T o decrement the number in a n alpha numerical entry field Left key T o scroll backwards through multiple choices within a field T o move the entry position in a n alpha numeric...

Page 296: ...s Installed datasets T est image From CF Card Time and date Printer AE title Hostname Domain name Printer IP address Netmask Router IP address Remove modality Add modality General Film calibration Warning messages Clean therm head Language Beep on empty tray General setting Service actions Quality Control Installation Repair history Maintenance Error messages Printer installation Calibration Resto...

Page 297: ...nter key operator command session Release remote command session Please wait while the printer is initializing for Remote Session Welcome to this DryStar AXYS printer Welcome to this DryStar AXYS printer Service engineer tools Key operator tools Operator tools Specialist tools Security Tools HELP Welcome to this DryStar AXYS printer 1 2 3 4 5 6 7 8 9 Show settings Change settings Print image Save ...

Page 298: ...Print Image 3 3 1 Welcome to this DryStar AXYS printer 1 2 To Printer Flash Card To this Computer Go back Key Operator Save configuration 4 Welcome to this DryStar AXYS printer 1 2 3 From CF Card From Remote Computer Frm dd mm yy hh mm Go back Key Operator Restore configuration 5 Welcome to this DryStar AXYS printer 1 2 Film Calibration Therm Head clean Go back Key Operator Calibration 6 Welcome t...

Page 299: ...ges Reset Values Film view Tag film format Tag film type DT2B 14x17 2 2 Key Operator Change input tray settings Upper input tray Normal Back Welcome to this DryStar AXYS printer 2 3 1 2 3 Network settings Remove modality Add modality Go back Printer AE title Hostname Domain name Printer IP Address Netmask Router IP address Key Operator Change network settings dicom Key Operator Change network sett...

Page 300: ...install please make backup 1 2 OK make backup SKIP backup Go back Key Operator Installation Key Operator Installation Choose Dataset Browse Proceed Go Back Reset Values Welcome to this DryStar AXYS printer Welcome to this DryStar AXYS printer Welcome to this DryStar AXYS printer yes no Date 20 NOV 2004 Time 09 01 Initials Max D 3 1 DD 0 2 Mid Density 1 1 Base Fog 0 2 New reference Comment Data of ...

Page 301: ...oftware versions Show printer status Enter key operator command session Enter service command session Release remote command session Reboot Drystar Reset Calibration Data Scaling Setup DICOM profile editor LPD Profile Editor Network configuration Printqueue editor Controller logging Modules Remote sessions Connectivity tools Module repair see page 7 see page 20 S1 S2 S3 S4 S5 S6 S7 S8 S9 S10 S11 S...

Page 302: ...tting 50 ______________ Status Blemish heating enabled CDM device enabled Online diagnosis enabled Security disabled ______________ Storage Removable storage not present Removable storage size Removable storage free Fixed c storage size 122 MB 128045056 bytes Fixed c storage free 0 bytes INFO COUNTER drystar Axys 1 Site and system info Date 18 Nov 2007 Time 12 42 15 1 1 Device info Product name dr...

Page 303: ...und the default site profile will be used Setup Dicom Print SCP Servers Edit the default site profile Validate profiles Auto Dicom Configuration New View Edit Remove Clone HOME HELP DICOM Input Setup S9 S10 S10 1 S10 3 S10 2 class com agfa med odin printerserver CalibrationRestore Default uppertray Return to the default filmcalibration values Current calibration will be copied to a backup Default ...

Page 304: ...rity LOW Polarity normal Film Orientation portrait Trim NO Film Size ID Medium Type Print even if Film Size ID Medium type not supported YES Print even if Application type not supported NO Initial allocated memory size per printjob MBytes 100 Kernel Interpolation Type cubicHighRes smoothFactor 0 7 Magnification Magn max Decimate Crop behaviour disabled Pixel Size ratio 1 0 1 0 Perception LUT kanam...

Page 305: ...Print Priority Polarity Film Orientation Trim Film Size ID Medium Type Print even if Film Size ID Medium type not supported Kernel Magnification Pixel Size Perception LUT Decimate Crop behaviour Kanamori Like value Custom value Illumination cd m2 Reflected Ambient Light cd m2 Border Density OD x100 Empty Image Density OD x100 Minimum Density OD x100 Maximum Density OD x100 Header Annotation Conten...

Page 306: ...t HOME HELP LPD Server Setup S11 1 portrait Grayscale Grayscale normal r0 pagesizeDependent bluebased Optimal Row _14x17 These settings can be overruled by the LPD protocol You can prevent this by checking the checkboxes at the front Profile Name Medium Size Medium Type Layout Number of copies Film Orientation Region Of Interest Pixel Size ratio Polarity Image Orientation Input LUT Usertaste Gray ...

Page 307: ...er translates a given hostname into an IP address Test DNS Enter a hostname Resolve Add Changes require reboot HOME HELP Network Configuration Remove UTP 10 OK Factory Defaults Clear Counters Basic IP Routing DNS Ethernet Reboot Device Ethernet Current line properties 100 Mbit Full Duplex Link partner other side 100FD 100HD 10FD 10HD autonegotiation Ethernet MAC address 08 00 66 80 44 09 Setup Typ...

Page 308: ... lookup and rotation DCMPrintToPrintJob converts Dicom Printjob into printqueue job PrintJobHandler is listener for jobavailable events and converts the available jobs in a bitmap for ColorConvertOpImage implements the conversion from one ColorSpace to another Page calculates the position of imageboxes in a page ComposeOpImage implements the operation to Compose multiple source images into one ima...

Page 309: ...are Menus Settings DOCUMENT CONTROL NOTE The controlled version of this document resides on MedNet Any printed copy of this document is uncontrolled Edition 1 Revision 0 DRYSTAR AXYS Chapter 3 7 19 10 2007 5367 100 Agfa Company Confidential ...

Page 310: ...y Control Installation leave remote session Service Main 1 2 3 4 5 6 7 8 9 Service engineer tools Key operator tools Operator tools Specialist tools Security Tools HELP Welcome to this DryStar AXYS printer HOME General information SERVICE TOOLS List of errors Printer Infocounters report DICOM LPD and HIPAA report QA Report Page Software versions Show printer status Enter key operator command sessi...

Page 311: ...YS printer Welcome to this DryStar AXYS printer Welcome to this DryStar AXYS printer Welcome to this DryStar AXYS printer Active Current settings Factory settings Cycles to go Maintenance Action Thermal head cleaning required Replace thermal head anti dust filter Replace thermal head end of life Film calibration upper tray required Film calibration lower tray required Service maintenance required ...

Page 312: ...5 Service Change Image quality settings Service Change Image quality settings off off Macbeth TR924 on on Gretag X Rite 301 X Rite 310 Accept changes Discard changes Reset Values Accept changes Discard changes Reset Values Welcome to this DryStar Axys printer Welcome to this DryStar Axys printer Film view Tag film format Tag film type Film view Tag film format Tag film type If RF tag is missing or...

Page 313: ...ware Installation Printer Installation Install Site Data Go back Service Installation 9 1 Before install please make backup 1 2 OK make backup SKIP backup Go back Service Installation Service Installation Choose Dataset Browse Proceed Go Back Reset Values Welcome to this DryStar AXYS printer Welcome to this DryStar AXYS printer Welcome to this DryStar AXYS printer 1 2 3 New Software Installation P...

Page 314: ...This page is intentionally left blank to enable print out of even and odd pages on duplex printers ...

Page 315: ...se patch_e_template_v07 Copyright 2011 Agfa HealthCare N V DRYSTAR AXYS Type 5367 100 Purpose of this document This document lists all customer released software versions with their features limitations and known problems Document History Edition Revision Release Date Changes compared to previous Version 1 2 1 3 03 2011 Added software version 2 1 0_C2 and 2 1 0_C3 See sections 7 and 8 Referenced D...

Page 316: ...arks of Agfa HealthCare N V WARNING Improper operation or service activities may cause damage or injuries INSTRUCTION 1 Read the Generic Safety Directions document see MEDNET GSO General Info Agfa HealthCare Publications Service Manual prior to attempting any operation repair or maintenance task on the equipment 2 Strictly observe all safety directions within the Generic Safety Directions and on t...

Page 317: ... 3 SOFTWARE PATCH 1 6 2_C1 8 3 1 Solved Bugs 8 3 2 New Features 9 3 2 1 New Feature Compliance Film Calibration 10 3 2 2 New Feature Activation of a new 10x12 Layout for GE Modalities 11 3 2 3 New Feature Possibility to read out future ISO Format RF Tags 12 4 SOFTWARE PATCH 1 6 3_C1 12 4 1 Solved Bug 12 5 SOFTWARE RELEASE 2 0 0_C2 13 5 1 New Features of Software 2 0 0_C2 13 5 2 Solved Bug in Softw...

Page 318: ...ware release can be downloaded 1 1 Supported Features All features as specified Printing diagnostic data according to the DICOM standard with a resolution of 508dpi Printing on Drystar DT2B DT2C and DT2 Mammo blue base films Supported film sizes o DT2B DT2C 14x17 14x14 14x11 10x12 8x10 o DT2 Mammo 14x11 10x12 8x10 1 2 Limitations Software 1 6 0_C1 has following limitations Only one module spare pa...

Page 319: ...s it is possible that a false error message MOD08 please replace module is shown MOD08 is the Devil CPU after start up of the printer A reset of the printer solves the problem The operator browser menu does not allow to delete images or set into emergency status It is possible however to delete print jobs in the operator menu of the local keypad refer to Drystar Axys user manual For TH profile cal...

Page 320: ...fects or otherwise minor improvements to performance without changing the software s basic design structure or functionality On MedNet GSO Library always the latest available software release can be downloaded 2 1 Solved Software Bugs With software patch 1 6 1_C1 following software bugs are solved Symptom 1 Sporadic input film jams for films from the upper tray mainly 14 films Error message at the...

Page 321: ...up timing The last digit in the spare part number indicates the spare part revision at release of this document When ordering the actual revision of the spare part is delivered NOTE Upper pickup modification kit CM 3617030 can be used for Drystar Axys and Drystar 5302 Symptom 2 When printing calibration films Film calibration Internal Densitometer Calibration Thermal Print Head Calibration the pri...

Page 322: ...nce without changing the software s basic design structure or functionality On MedNet GSO Library always the latest available software release can be downloaded 3 1 Solved Bugs In Software Patch 1 6 2_C1 following software bugs are solved Symptom 1 When overlapping the two Mammo films at the lightbox there is too much dazzling light between the two films as the clear border at the chest wall side ...

Page 323: ...browser key operator menu it is not possible to print DICOM images Following modalities are affected Phillips MXVIEW Phillips MX8000 Fuji ADR series Fuji Eyepix Adding these modalities manually via DICOM host profile does not give any problem Cause Software problem in software 1 6 2_C1 3 2 New Features Software patch 1 6 2_C1 contains following new features Compliance film calibration Details see ...

Page 324: ...l Documentation Drystar Axys DD DIS033 07E Adjustments and Calibrations edition 1 revision 1 and higher revisions Service Calibration Film calibration 1 Internal densitometer calibration 2 Thermal print head calibration 3 Horizontal alignment 4 Smart profile calibration 5 Compliance film calibration 6 Go back This page is generated on 21 05 2008 12 33 33 Step Target valuesMeasured values Step Targ...

Page 325: ...s active for all other modalities too It is installed via browser menu service tools GE Mammo layout installation See Figure 3 HOME General information SERVICE TOOLS List of errors Printer Infocounters report DICOM LPD and HIPAA report QA Report Page Software versions Show printer status Enter key operator command session Enter service command session Release remote command session Reboot Drystar ...

Page 326: ...atch 1 6 3_C1 Type of new Software Software Update Release Date 07 2008 Printer Serial Numbers SN 1423 plus 1420 Availability MedNet GSO Library Prerequisites None NOTE A software upgrade is a functional extension An update software patch is an internal change designed to correct bugs fix defects or otherwise minor improvements to performance without changing the software s basic design structure ...

Page 327: ...contains following new features Software supports both printers DRYSTAR 5302 and DRYSTAR AXYS Improved RF tag locking mechanism identical to RF tag locking mechanism of DRYSTAR 550X software 6 0 Support of Compliance film calibration which may be required to pass the Afssaps EUREF acceptance test For DRYSTAR AXYS this feature was available already as of software 1 6 3_C1 however only for Mammo pri...

Page 328: ...n pressing the ON OFF switch the printer cannot be switched off Access via network is not possible to reboot remotely After unplugging and reconnecting the mains cable the printer usually starts up normally Cause CPU Driver problem Solution Install software 2 0 0_C2 to prevent the problem 5 3 Known Software Issue Symptom Login problems with the Firefox browser In some cases the login to the servic...

Page 329: ...n An update software patch is an internal change designed to correct bugs fix defects or otherwise minor improvements to performance without changing the software s basic design structure or functionality On MedNet GSO Library always the latest available software release can be downloaded 6 1 New Feature of Software 2 0 0_C3 Compared to software 2 0 0_C2 the RF tag locking mechanism is further imp...

Page 330: ... is an internal change designed to correct bugs fix defects or otherwise minor improvements to performance without changing the software s basic design structure or functionality On MedNet GSO Library always the latest available software release can be downloaded 7 1 New Feature of Software 2 1 0_C2 Software 2 1 0_C2 supports new hardware Devius module Tyndal module SATA hard disk for DRYSTAR 5302...

Page 331: ...formance without changing the software s basic design structure or functionality On MedNet GSO Library always the latest available software release can be downloaded 8 1 New Feature of Software 2 1 0_C3 By installing software 2 1 0_C3 the Devius board is re programmed to revision 9 This is a solution for the warning message RF Tag Problem after 10 printed films of a new film pack For details see D...

Page 332: ......

Page 333: ...ersion of this document resides on MedNet Any printed copy of this document is uncontrolled 10 2007 printed in Germany Agfa Company Confidential Document Node ID 14349767 Copyright 2007 Agfa HealthCare N V DRYSTAR AXYS Type 5367 100 The Frequently Asked Questions FAQ are integrated in the troubleshooting guide Refer to chapter 3 3 of this service manual ...

Page 334: ......

Page 335: ...YS Type 5367 100 Purpose of this document This document contains Reference diagrams for identification and location of assemblies inside the machine including an alphanumeric reference list Wiring diagrams Document History Edition Revision Release Date Changes compared to previous Version 1 1 1 2 03 2011 Added new hardware Devius PCB Tyndall CPU DDR2 RAM and SATA Hard disk in the reference diagram...

Page 336: ...fa HealthCare N V WARNING Improper operation or service activities may cause damage or injuries INSTRUCTION 1 Read the Generic Safety Directions document see Agfa HealthCare Library General Info Agfa HealthCare Publications Service Manual prior to attempting any operation repair or maintenance task on the equipment 2 Strictly observe all safety directions within the Generic Safety Directions and o...

Page 337: ...ocument resides on the Agfa HealthCare Library Any printed copy of this document is uncontrolled Edition 1 Revision 2 DRYSTAR AXYS Chapter 4 3 03 2011 Type 5367 100 Agfa Company Confidential LIST OF CONTENTS 1 REFERENCE DIAGRAMS 4 1 1 Components Location 4 1 2 List of Components 7 2 WIRING DIAGRAMS 9 ...

Page 338: ...ry Any printed copy of this document is uncontrolled 1 Reference Diagrams 1 1 Components Location S12 S24 S11 S 1 in upper tray S2 in lower tray S13 S10 CDM S5 M5 M4 S6 CDM LED Board S4 RF1 in Lower Tray RF2 in Upper Tray M1 S9 536704aa_1 cdr Figure 1 Edition 1 Revision 2 DRYSTAR AXYS Chapter 4 4 03 2011 Type 5367 100 Agfa Company Confidential ...

Page 339: ... M8 F2 M6 in Upper pickup tray M7 in Lower pickup tray S14 S15 in Upper pickup tray S16 S17 in Lower pickup tray S18 S19 in Upper Lower pickup tray 536704aa_3 cdr S21 S22 M3 V1 V2 M9 SCSI SATA Hard Disk Devius Tyndall DDR2 RAM SATA hard disk introduced for SN 3177 SN 3180 SN 3494 to SN 3537 SN 3539 SN 3574 SN 3672 Figure 2 Edition 1 Revision 2 DRYSTAR AXYS Chapter 4 5 03 2011 Type 5367 100 Agfa Co...

Page 340: ...Any printed copy of this document is uncontrolled Key6 Thanos Tyndall Power Supply OSB 536704aa_2 cdr TPH RAM DDR2 RAM Devil Devius Devius Tyndall DDR2 RAM SATA hard disk introduced for SN 3177 SN 3180 SN 3494 to SN 3537 SN 3539 SN 3574 SN 3672 Figure 3 Edition 1 Revision 2 DRYSTAR AXYS Chapter 4 6 03 2011 Type 5367 100 Agfa Company Confidential ...

Page 341: ...ut Module E3 E label Lower Pickup Module E4 E label Power Supply E5 E label TPH module E6 E label 6 Display Module E7 E label 7 Upper Pickup Module F1 Thermal Head fan Figure 2 F2 CPU cooling fan E box Figure 2 Hard disk Drive Hard disk Drive 73 GB or more Figure 2 KEY6 PCB Keypad module Figure 3 M1 Drum motor Figure 1 M2 Cam drive motor Figure 2 M3 Suspension drive motor Figure 2 M4 Output roller...

Page 342: ...gure 2 S8 Top cover closed detection Figure 2 S9 Print drum drive rotation detection Figure 1 S10 Densitometer home position Figure 1 S11 Take over passed detection Figure 1 S12 Overheat sensor Figure 1 S13 Input sensor TPH Figure 1 S14 Upper pickup home position Figure 2 S15 Upper pickup pickup position Figure 2 S16 Lower pickup home position Figure 2 S17 Lower pickup pickup position Figure 2 S18...

Page 343: ...ge 2 Page 3 Page 4 The following circuit diagrams are attached Description 1 Wiring Diagrams Drystar Axys with Devil PCB Main Page Diagram 1 8 2 Wiring Diagrams Drystar Axys with Devius PCB Main Page Diagram 2 8 3 Wiring Diagram Lower Pickup Assembly Diagram 3 8 4 Wiring Diagram Lower Tray Assembly Diagram 4 8 5 Wiring Diagram Output Assembly Diagram 5 8 6 Wiring Diagram Pump Valves Assembly Diagr...

Page 344: ... Prot earth 16 PS P1 X X DEV P11 1 1 RD DEV P11 2 2 RD HD 4 3 RD DEV P11 4 4 BK DEV P11 5 5 BK HD 3 6 BK DEV P11 7 7 VT DEV P11 8 8 BK DEV P11 9 9 BU DEV P11 10 10 BU DEV P11 11 11 BU DEV P11 12 12 BK DEV P11 13 13 BK DEV P11 14 14 BK DEV P11 15 15 VT DEV P11 16 16 BK DEV P11 17 HD 1 17 YE DEV P11 18 HD 2 18 BK DEV P11 19 19 WH DEV P11 20 20 BK DEV P11 21 21 GN BU22 MINI FIT JR DEV P11 22 22 OG P1...

Page 345: ...HQ HG RQ 2XWSXW 2 9 JUHHQ HG RQ 2XWSXW 2 2YHUWHPSHUDWXUH UHG HG RQ 2YHUWHPSHUDWXUH SURWHFWLRQ PRGH S WUDFWRU 3 73 8 B 9 3 7 3 8 B 7 3 9 3 26 6HQVRU 6 QSXW VHQVRU 73 26 6HQVRU 6 QSXW VHQVRU 73 6 3 6HQVRU 6 7UDQVSRUW UROOHU RSHQ GHWHFWLRQ 6 3 6HQVRU 6 3ULQW GUXP GULYH 6 3 6HQVRU 6 VXVSHQVLRQ SRVLWLRQ 6 3 6HQVRU 6 VXVSHQVLRQ SRVLWLRQ 6 3 6HQVRU 6 73 KRPH SRVLWLRQ 6HQVRU 6 0 0 0RWRU 0 6XVSHQVLRQ PRWRU...

Page 346: ...6 6 6 6 5 6 6 6 97 67 02 8B 6 6 6 8 RZHUB38 6 6 6 2 6 6 6 5 6 6 6 8 6 6 6 67 02 8B 6 6 6 0 0 0 6 6 6 5 6 6 6 0 0 0 5 0 0 0 6 6 6 6 6 6 6 6 6 6 6 6 5 6 6 6 6 6 6 97 6 6 6 6 6 6 8 RZHUB38 2 RZHUB38 5 RZHUB38 RZHUB38 6 6 6 6 6 6 1 6 6 6 6 6 6 1 8 0 1 7 6 6 6 6 6 D 9 3 9 3 5 9 3 9 3 5 9 3 97 9 3 97 9 3 2 9 3 5 9 3 9 3 9 3 5 9 3 97 9 3 67 0 1 7 9 3 97 6 6 6 6 6 6 6 6 6 1 8 6 6 6 1 6 3 6HQVRU 6 0 H DEHO...

Page 347: ...U 6 RZHUBWUD 5 1 B 5 0 2 67 5 67 02 60 6 6 6 6 6 6 5 6 6 6 97 6 6 6 5 5 67 5 5 67 5 5 5 67 5 5 67 5 5 67 8 6 6 6 8 6 6 6 6 6 D 9 3 5 9 3 97 9 3 9 3 9 3 5 9 3 9 3 9 3 8 9 3 8 9 3 97 6 6 6 6 6 6 5 6 6 6 8 6 6 6 97 5 67 5 6 6 6 6 6 6 5 6 6 6 6 6 6 8 6 6 6 8 6 3 6HQVRU 6 5 7 5 5 38 RZHU PDJD LQH HPSW GHWHFWLRQ 2 5 75 66 RQQHFWLRQ EHWZHHQ DVVHPEOLHV 0O 5HFHSWDFOH RXVLQJ 0DWHV ZLWK 0O UDZHU FRQQHFWRU 5 ...

Page 348: ... X 20 X XS3bXS3 22 21 BK BU22 70066 XS3bXS3 21 22 BK XS3 XS3a DEV P9 1 1 YE DEV P9 2 2 BK DEV P9 3 3 WH DEV P9 5 4 BU DEV P9 6 5 VT DEV P9 8 6 BU DEV P9 9 7 VT DEV P9 12 8 BU DEV P9 11 9 VT DEV P9 10 10 BK DEV P9 14 11 OG DEV P9 15 12 WH DEV P9 13 13 RD DEV P9 16 14 BK DEV P1 30 15 RD DEV P1 28 16 VT DEV P1 29 17 BK DEV P1 18 18 BK DEV P1 17 19 VT DEV P1 16 20 RD XS2aXS2 18 21 VT ST22 70107 DEV P1...

Page 349: ...RUELGGHQ 6KHHW QDPH 3803 9 9 6 60 5 1 5 0 U VWDU 6 3XPSB B9DOYHVBDVP 9 8SSHU PDJD LQH YDFXXP YDOYH 9 RZHU PDJD LQH YDFXXP YDOYH 5 1 B 5 0 3 5 1 B 5 0 3 0 0 UG EN 9DFXXP SXPS 6 6 E 9 3 8 9 3 1 9 3 9 3 1 9 3 8 0 1 7 5 9 3 1 6 6 D 0 UG 5 0 EN 9 9 9 67 0 1 7 5 9 6 6 E 9 3 1 8 0 1 7 5 9 3 3803 9 9 6 66 38 9DFXXP SXPS 8SSHU PDJD LQH YDFXXP YDOYH RZHU PDJD LQH YDFXXP YDOYH 3803 9 9 6 66 03 2011 Edition 1...

Page 350: ...6 6 0 0 0 5 0 0 0 6 6 6 6 6 6 6 6 6 6 6 6 5 6 6 6 6 6 6 97 6 6 6 6 6 6 97 8SSHUB38 2 8SSHUB38 5 8SSHUB38 8SSHUB38 6 6 6 6 6 6 1 6 6 6 6 6 6 1 8 0 1 7 6 6 6 6 6 D 9 3 9 3 5 9 3 9 3 5 9 3 97 9 3 97 9 3 2 9 3 5 9 3 9 3 9 3 5 9 3 97 9 3 67 0 1 7 9 3 97 6 6 6 6 6 6 8 6 6 6 1 6 6 6 1 8SSHUB38 6 6 6 2 6 6 6 5 6 6 6 8 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 5 6 6 6 97 67 02 8B 6 6 6 97 0 0 0 6 6 6 5 67 02 8B 6 6 6 ...

Page 351: ... 67 8 6 6 6 8 6 6 6 6 6 D 9 3 5 9 3 97 9 3 9 3 9 3 5 9 3 9 3 9 3 8 9 3 8 9 3 6 D 6 97 6 6 D 9 3 9 3 97 9 3 5 9 3 9 3 97 9 3 5 6 D 6 97 8 9 3 97 6 6 E 6 6 97 6 5 6 6 97 6 5 6 E 6 8 6 E 6 8 6 E 6 6 E 6 97 6 6 E 6 5 6 6 E 6 5 8 6 E 6 6 E 6 97 5 67 5 6 6 6 6 6 6 5 6 6 6 6 6 6 8 6 6 6 8 6 6 6 6 6 6 5 8 6 6 6 97 6 6 6 26 6HQVRU 6 26 6HQVRU 6 6 3 6HQVRU 6 833 5 75 66 RQQHFWLRQ EHWZHHQ DVVHPEOLHV 0O 5HFHS...

Page 352: ......

Page 353: ...thorized personnel is not permitted and may be unlawful DOCUMENT CONTROL NOTE The controlled version of this document resides on Agfa HealthCare Library Any printed copy of this document is uncontrolled Agfa Company Confidential 09 2011 Printed in Germany Document Node ID 18050603 Edition 1 Revision 2 spare_parts_e_template_v09 Copyright 2011 Agfa HealthCare N V ...

Page 354: ...ncontrolled WARNING Hazards may be introduced because of component failure or improper operation Replace defective parts with Agfa HealthCare original spare parts Use only tools and measuring instruments which are suitable for the procedure Only approved Agfa HealthCare accessories must be used For a list of compatible accessories contact your local Agfa HealthCare organization or www agfa com INS...

Page 355: ...cal waste disposal service or producer distributor of this equipment If your equipment contains removable batteries or accumulators please dispose of these separately according to local regulations Contact Spare Parts ordering Spare Parts returns Europe Overseas Orderprocessing europe matrium net Orderprocessing overseas matrium net Worldwide Returns matrium net NOTE Agfa reserves the right to del...

Page 356: ...E PAGES 10 11 LEISTUNGSEINHEIT POWER UNIT MODULE DE PUISSANCE PAGES 16 17 ANTRIEB DRIVE ENTRAÎNEMENT PAGES 8 9 DD DIS274 07M Spare Parts List DOCUMENT CONTROL NOTE The controlled version of this document resides on Agfa HealthCare Library Any printed copy of this document is uncontrolled Edition 1 Revision 2 Chapter 5 4 Drystar Axys Type 5367 100 Agfa Company Confidential 09 2011 ...

Page 357: ...een arrow to navigate back to the overview of the modules How to navigate the Spare Parts List online with the Acrobat Reader DOCUMENT CONTROL NOTE The controlled version of this document resides on Agfa HealthCare Library Any printed copy of this document is uncontrolled Edition 1 Revision 2 Chapter 5 5 Drystar Axys Type 5367 100 Agfa Company Confidential 09 2011 ...

Page 358: ...a HealthCare Library Any printed copy of this document is uncontrolled 04 07 04 03 MOD 01 02 MOD 02 05 06 01 MOD 10 05 05 5367_chap05_01_m cdr 08 DD DIS274 07M Spare Parts List Edition 1 Revision 2 Chapter 5 6 Drystar Axys Type 5367 100 09 2011 COVERS Agfa Company Confidential ...

Page 359: ...ROHS MOD 02 LOWER TRAY W O TAG ROHS MOD 02 RÉCEPTACLE EN TÔLE EN BAS W O TAG ROHS MOD 02 CM 3492672 3 OBERE ABLAGE W O TAG ROHS MOD 01 UPPER TRAY W O TAG ROHS MOD 01 RÉCEPTACLE EN TÔLE EN HAUT W O TAG ROHS MOD 01 CM 3492812 4 OBERE HINTERE OBERE VERKLEIDUNG ROHS TOP COVER REAR TOP COVER ASSY ROHS COUVERCLE EN HAUT ROHS CM 3492822 5 LINKE RECHTE VORDERE ABDECKUNG ROHS LEFT RIGHT FRONT COVER ROHS CO...

Page 360: ...rary Any printed copy of this document is uncontrolled 5367_chap05_02_m cdr 01 MOD 06 04 05 MOD 14 06 MOD 13 07 09 MOD 11 08 02 MOD 03 02 MOD 04 03 MOD 16 10 11 12 DD DIS274 07M Spare Parts List Edition 1 Revision 2 Chapter 5 8 Drystar Axys Type 5367 100 09 2011 DRIVE Agfa Company Confidential ...

Page 361: ...2695 5 GETRIEBE RIEMEN ROHS MOD 14 GEAR BOX BELT ROHS MOD 14 ENGRENAGE COURROIE ROHS MOD 14 CM 3492762 6 NOCKENWELLENANTRIEBSMODUL ROHS MOD 13 CAM DRIVE MODUL ROHS MOD 13 MODULE D ENTRAÎNENMENT ROHS MOD 13 CM 3492872 7 TPH FLACHKABEL ROHS TPH FLAT CABLE ROHS CABLE PLAT TPH ROHS CM 3097371 8 FILMSENSOR ROHS S5 FILM SENSOR ROHS S5 DETECTEUR FILM ROHS S5 CM 3492803 9 EINGABESENSORMODUL ROHS MOD 11 IN...

Page 362: ...gfa HealthCare Library Any printed copy of this document is uncontrolled 03 06 05 MOD 15 01 MOD 17 02 MOD 07 07 S9 04 5367_chap05_03_m cdr DD DIS274 07M Spare Parts List Edition 1 Revision 2 Chapter 5 10 Drystar Axys Type 5367 100 09 2011 DRIVE LEFT Agfa Company Confidential ...

Page 363: ...LVES ROHS MOD 17 CM 3492702 2 WALZENANTRIEBSMODUL ROHS MOD 07 DRUM DRIVE MODULE ROHS MOD 07 MODULE D ENTRAÎNEMENT DU CYLINDRE ROHS MOD 07 CM 3492711 3 DRUCKTROMMEL ROHS PRINT DRUM MODULE ROHS TAMBOUR D IMPRESSION ROHS CM 3433521 4 HAUPTANTRIEBSRAD WALZENANTRIEBSRIEMEN ROHS MAIN DRIVE PULLEY DRUM BELT ROHS ROUE D ENTRAÎNEMENT COURROIES D ENTRAÎNEMENT DU TAMBOUR ROHS CM 3492742 5 ANTRIEBSMODUL ROHS ...

Page 364: ..._chap05_04_m cdr 01 MOD 05 02 03 MOD 20 EXCEPTIONS CONCERNING INTRODUCED HARDWARE SEE SIB13 DD DIS341 10E UP TO SN3671 04 MOD 08 07 MOD 18 08 MOD 18 06 MOD 12 05 09 IN COMBINATION WITH THANOS BOARD ONLY DD DIS274 07M Spare Parts List Edition 1 Revision 2 Chapter 5 12 Drystar Axys 5367 100 09 2011 DRIVE POWER UNIT UP TO SN3671 Agfa Company Confidential ...

Page 365: ...L THEMIS THANOS RAM REPLACEMENT ASSY MOD 08 BIS FN3671 DEVIL THEMIS THANOS RAM REPLACEMENT ASSY MOD 08 UP TO SN3671 DEVIL THEMIS THANOS RAM REPLACEMENT ASSY MOD 08 JUSQU AU NS3671 CM 6037350 5 DEVIL REPLACEMENT ASSY BIS FN3671 DEVIL REPLACEMENT ASSY UP TO SN3671 DEVIL REPLACEMENT ASSY JUSQU AU NS3671 CM 6009773 6 TYNDALL PCB MIT RAM MOD 12 SW 2 1 X TYNDALL PCB WITH RAM MOD 12 SW 2 1 X TYNDALL PCB ...

Page 366: ...t is uncontrolled 5367_chap05_04_m cdr 01 MOD 05 02 04 MOD 08 06 MOD 12 07 MOD 18 03 MOD 20 08 05 EXCEPTIONS CONCERNING INTRODUCED HARDWARE SEE SIB13 DD DIS341 10E FROM SN3672 DD DIS274 07M Spare Parts List Edition 1 Revision 2 Chapter 5 14 Drystar Axys Type 5367 100 09 2011 DRIVE POWER UNIT FROM SN3672 Agfa Company Confidential ...

Page 367: ... KÜHLGEBLÄSE MOD 20 VILIS COOLING FAN MOD 20 VENTILATEUR VILIS MOD 20 CM 6009784 4 DEVIUS TYNDALL RAM MODUL MOD 08 AB FN3672 SW 2 1 X DEVIUS TYNDALL RAM MODULE MOD 08 FROM SN3672 SW 2 1 X MODULE DEVIUS TYNDALL RAM MOD 08 A PARTIR DE NS3672 SW 2 1 X CM 6009762 5 DEVIUS PCB AB FN3672 SW 2 1 X DEVIUS PCB FROM SN3672 SW 2 1 X DEVIUS PCB A PARTIR DE NS3672 SW 2 1 X CM 6009773 6 TYNDALL PCB MIT RAM MOD ...

Page 368: ... Box Belt 06 Name Female vacuum connector Male vacuum connector Guide latch Vertical Alignment Gear box spring Gear box belt Quantity 2 2 2 1 1 Purpose Vacuum connector inside printer Counterpart to male vacuum connector at pickup module Vacuum connector at pickup module Vertical alignment guide latch left and right hand side Tension spring for gear box belt Belt for gear box 5367_chap05_05_m CDR ...

Page 369: ...ENTATION ROHS MOD 09 CM 3260514 2 SCSI HARDDISK 73GB ROHS BIS FN3671 SCSI HARDDISK 73GB ROHS UP TO SN3671 SCSI HARDDISK 73GB ROHS JUSQU AU NS3671 CM 6009730 3 SATA HARDDISK AB FN3672 SATA HARDDISK FROM SN3672 SATA HARDDISK A PARTIR DE NS3672 CM 3392031 4 SCSI HARDDISK KABEL ROHS BIS FN3671 SCSI HARDDISK CABLE ROHS UP TO SN3671 CABLE SCSI HARDDISK ROHS JUSQU AU NS3671 CM 6009740 5 SATA HARDDISK KAB...

Page 370: ...ntralized decentralized Stockpilling M Maintenance Parts required to maintain a machine Kit M should be available as case stock Includes all parts required to perform a maintenance according to the maintenance checklist Quantity covers one single maintenance L Local stock Extremely expensive or bulky parts Kit L should be part of the local central warehouse Determine quantity depending on costs an...

Page 371: ...for the following machine type s Device Name Type Number Specification 5367 0100 Drystar AXYS EYZ4E DOCUMENT CONTROL NOTE The controlled version of this document resides on Agfa HealthCare Library Any printed copy of this document is uncontrolled Accessory Overview Following accesssories are separately available Accessory Order Number MOBILE KIT EX2DV MAMMO LICENCE EY8RN ...

Page 372: ......

Page 373: ...rinted in Germany Agfa Company Confidential Document Node ID 24708127 eq_08_modifications_e_template_v07 Copyright 2008 Agfa HealthCare N V DRYSTAR AXYS Type 5367 100 Purpose of this Document This document provides an overview of all product modifications which may be of importance for Service They are listed in chronological sequence Document History Edition Revision Release Date Changes 1 0 05 2...

Page 374: ...of Agfa HealthCare N V WARNING Improper operation or service activities may cause damage or injuries INSTRUCTION 1 Read the Generic Safety Directions document see MEDNET GSO General Info Agfa HealthCare Publications Service Manual prior to attempting any operation repair or maintenance task on the equipment 2 Strictly observe all safety directions within the Generic Safety Directions and on the pr...

Page 375: ...evision 0 DRYSTAR AXYS Chapter 8 3 05 2008 Type 5367 100 Agfa Company Confidential OVERVIEW OF MODIFICATIONS Timeline Modification Referenced Documents 2008 March Introduction of Upper Pickup Modification Kit incl Software 1 6 1_C1 SB No 02 DD DIS037 08E Enclosure DD DIS399 08E February Changed Compact Flash Card 128 MB CM 343356 1 to 1 GB None 2007 December Changed Input Sensor Module None ...

Page 376: ...put Sensor Module CM 349280 1 Reason Improvement of production process Product s Serial Number s SN 1069 December 2007 Part Number s Input Sensor Module CM 349280 2 Referenced Document s None 2 Changed Compact Flash Card 128 MB CM 343356 1 to 1 GB Reason Change of type of compact flash card for printer backup due to purchasing reasons Product s Serial Number s Printers and spare parts delivered as...

Page 377: ...al 3 Introduction of Upper Pickup Modification Kit incl Software 1 6 1_C1 Reason Improvement of film run Product s Serial Number s The Upper Pickup Modification Kit was introduced in production as of SN 1188 Software 1 6 1_C1 was introduced in production as of SN 1188 plus SNs 1066 1161 1172 1173 1176 1178 March 2008 Part Number s Upper Pickup Modification Kit CM 361703 0 Referenced Document s SB ...

Page 378: ......

Page 379: ...ction of this document by unauthorized personnel is not permitted and may be unlawful DOCUMENT CONTROL NOTE The controlled version of this document resides on MedNet Any printed copy of this document is uncontrolled 09 2007 printed in Germany Agfa Company Confidential Document Node ID 14349768 eq_09_maintenance_e_template_v03 Copyright 2007 Agfa HealthCare N V DRYSTAR AXYS Type 5367 Edition 1 Revi...

Page 380: ...e 5367 Agfa Company Confidential WARNING Improper operation or service activities may cause damage or injuries INSTRUCTION 1 Read the Generic Safety Directions document see MEDNET GSO General Info Agfa HealthCare Publications Service Manual prior to attempting any operation repair or maintenance task on the equipment 2 Strictly observe all safety directions within the Generic Safety Directions and...

Page 381: ...is uncontrolled Edition 1 Revision 0 DRYSTAR AXYS Chapter 9 3 09 2007 Type 5367 Agfa Company Confidential Purpose of this document This document describes all routines and tests to be carried out during maintenance Document History Edition Revision Release Date Changes 1 0 09 2007 Initial Version Referenced Documents Document Title n a n a ...

Page 382: ... 8 2 2 2 Infocounter Analysis via Printers Infocounters Report 9 2 2 3 Infocounter analysis via Service Command Session 9 2 2 4 How to evaluate the Infocounter 11 2 3 Image Quality Check 12 3 MAINTENANCE POINTS POWER OFF 13 3 1 Check of general Condition 13 3 2 Check of Condition of Belts and Gears 13 3 3 Visual Check of the Cables 14 3 4 Cleaning of the Suction Cups 14 3 5 Cleaning of Electronics...

Page 383: ... 09 2007 Type 5367 Agfa Company Confidential 4 3 SW Patch Installation 21 4 4 Film Calibration 21 4 5 Density Reproduction Check 22 4 6 Test Image Evaluation 24 4 7 Host System Image Quality Check 24 5 COMPLETION OF THE MAINTENANCE 25 5 1 Backup Data on CF Card 25 5 2 Reset of Maintenance Action counters and relative Counters in the Infocounter 26 5 3 Customer Conversation 26 MAINTENANCE CHECKLIST...

Page 384: ...ith the maintenance instructions Additional maintenance steps can be required according to national regulations 1 1 Maintenance Frequency The maintenance has to be carried out every 25 000 printed films or every 12 months whichever comes first 1 2 Required Time Time required for the maintenance 2 hours 1 3 Required Tools Description Order Number Comment Calibrated densitometer recommended types ar...

Page 385: ... cloth Commercially available Ink eraser Pelikan 10 9 9999 1176 0 Vacuum cleaner 220 240V 50Hz CM 9 9999 0895 0 Dirt bags for vacuum cleaner 10 x CM 9 9999 0896 0 1 5 Required Spare Parts The following spare parts are part of the Maintenance assortment CM 0 5367 0999 733 Description Order Number Case of Exchange Usage Air Filter CM 343350 X Exchange each maintenance NOTE Always have the latest Dry...

Page 386: ... 2 2 Infocounter Analysis Three possibilities exist to get access to the infocounter see also Figure 3 Access to the infocounter Refer to A Via printed infocounter on film 2 2 1 B Via Printers Infocounters Report 2 2 2 C Via Service Command session 2 2 3 NOTE We recommend to use for Infocounter Analysis method A Via printed infocounter on film as this is the fastest way to access the infocounter 2...

Page 387: ...dress window 4 Select menu Service Engineer Tools Printer Infocounters Report 5 Analyze the infocounter file as described in section 2 2 4 How to evaluate the Infocounter Figure 1 2 2 3 Infocounter Analysis via Service Command Session Required Tools Service PC required to connect to the printer and analyze the infocounter 536632el cdr Printer Service PC 1 Connect the Service PC to the network 2 St...

Page 388: ...sions Show printer status Enter key operator command session Enter service command session Release remote command session Scaling Setup DICOM profile editor LPD Profile Editor Network configuration Printqueue editor Controller logging Modules Remote sessions Welcome to this DryStar Axys printer Connectivity tools Module repair Service Loggings 1 2 To printer flash card To this computer Go back Wel...

Page 389: ...ers txt evaluation What to check in the infocounter Comment 2 3 Reliability Should normally be not higher than 0 05 Higher fault rates need a thorough investigation 3 3 Hardware Modification History By comparing the status of the device with the available modifications refer to Technical Documentation chapter 8 the exact hardware status can be determined 3 4 Software Modification History By checki...

Page 390: ...e maintenance The next step is only required if the customer is using films with 14 and 10 width in mixed mode see infocounter 4 At the lightbox check both STI2 and medicalimage for dark print stripes in the 10 film edge area If print stripes are visible in the STI2 and medicalimage perform a smart profile calibration Refer to section 4 1 If print stripes are visible in the STI2 image no smart pro...

Page 391: ...isual Check of the Cables 3 3 Cleaning of the Suction Cups 3 4 Cleaning of Electronics Compartment 3 5 Cleaning of Power Supply Air Inlet 3 6 Check and Cleaning of Drum 3 7 Cleaning of Thermal Print Head 3 8 Electrical Check 3 9 3 1 Check of general Condition 1 Perform a visual check of Power and network cables Covers The Output tray The upper and lower Input tray 2 Disconnect the power and networ...

Page 392: ...ter The insulation may not show any sign of wear or damage 3 4 Cleaning of the Suction Cups 1 Pull out the upper pickup unit 2 Clean the suction cups with a lint free cloth lightly moistened with Isopropyl alcohol 3 Insert the upper pickup unit 4 Repeat this step for the lower pickup unit 536709da cdr 1 Figure 7 536609cc cdr 2 536709db cdr 3 Figure 8 Figure 9 3 5 Cleaning of Electronics Compartmen...

Page 393: ... 3 7 Check and Cleaning of Drum 536609df cdr Figure 12 1 Open the top cover 2 Open the Hold down bracket 3 Open the TH Module 4 Remove the flat belt 5 Check the drum for dirt or damage by turning at the drum drive pulley 6 In case signs of damage are visible also check the STI2 test image printed under step 2 3 Image Quality Check A damage will be visible in the flatfield area of the STI2 test ima...

Page 394: ...Head 536609cg cdr Figure 13 1 Clean the TH resistor line with a lint free cloth lightly moistened with Isopropyl alcohol The last cleaning action must be carried out in one way For the Isopropyl alcohol observe the rule As less water is in the alcohol as better We recommend using a 90 Isopropyl alcohol 2 Remount all covers 3 Insert both input trays CAUTION Thermal Print Head is fragile Risk of dam...

Page 395: ...eck is split up in two steps Step Task Details see 1 Perform the basic protective earth test 3 9 1 2 Perform the electrical check according to National Regulations 3 9 2 3 9 1 Basic Protective Earth Test 536631JN cdr 2 1 Use a multimeter that is capable to measure a resistance 1 Ohm 2 Put one probe at the ground connector of the mains input 3 Put the other probe at a few metallic parts the custome...

Page 396: ...stomer could touch The electrical resistance in this case has to be 0 3 Ohm Refer to Figure 15 Figure 15 IMPORTANT If the electrical resistance of one measurement exceeds the limits 0 2 Ohm without mains cable 0 3 Ohm with mains cable check the printer for loose grounding connections 3 9 2 Electrical Check according to National Regulations 1 Perform the electrical check according to National Regul...

Page 397: ...print stripes in the 10 film edge area are visible in the STI2 and medicalimage Refer to section 2 3 1 Connect the service PC to the printer 2 Enter the IP address into the browser window 3 Enter Service Engineer Tools 4 Select Enter Service Command Session 7 Calibration 5 Smart Profile Calibration 5 Select the densitometer you use for measurement 6 Select the film tray with 14x17 14x14 or 14x11 f...

Page 398: ...te 301 X Rite 310 Start Calibration Go back Reset Values Welcome to this DryStar Axys printer Smart Profile Calibration Smart Profile Calibration Select the input tray with 14 x17 film format Smart Profile Calibration Step Step Step Step Step 00 01 02 03 04 05 06 07 08 09 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 5...

Page 399: ...till it shows the READY screen takes approx 7 minutes 2 Open top cover 3 Check display message It must show Close top cover 536609fa cdr Figure 18 4 3 SW Patch Installation 1 Install the latest SW patch in case a recommended SW patch for the printer is available Refer to chapter 3 8 Software Releases Patches 4 4 Film Calibration 1 Select in key operator menu 6 Calibration 1 Film Calibration 2 Foll...

Page 400: ...d with an external densitometer is checked This is a plausibility check for the CDM In case the measured density values are out of the specified range the CDM requires a re calibration This procedure must be carried out with every used film type 1 In the browser pages select menu Service Engineer Tools Enter Service Command Session 3 Print Testimage 2 Print out LinCheck zip and measure the density...

Page 401: ...0 276 5 6 5 0 0 300 305 5 0 When entering the measured density values the decimal point appears in a different spot e g measured value 0 23 value to enter 23 0 Density Lower limit Upper limit 0 0 0 30 3 3 45 4 4 60 5 5 75 6 6 90 7 7 X rite 310 105 8 8 X rite 331 120 9 9 X rite 341 135 9 9 X rite 301 150 9 9 Mac Beth TR924 165 9 9 Gretag D200 2 180 10 10 195 10 10 210 10 10 225 10 10 240 10 10 255 ...

Page 402: ... direction Point artifacts Surface artifacts 6 Evaluate the test images at follows 1 If no artifacts are seen on the STI STI2 image continue with next step 2 If artifacts seen on STI STI2 image aren t visible on the medicalimage continue with next step 3 In other case i e artifacts which are visible in STI STI2 are visible on medicalimage too refer to Technical Documentation chapter 3 3 Troublesho...

Page 403: ...ompletion of the maintenance comprises following tasks Task Details see Backup Data on CF Card 5 1 Reset of Maintenance Action counters and relative Counters in the Infocounter 5 2 Customer Conversation 5 3 5 1 Backup Data on CF Card 1 Open the upper film tray and take out the CF card 2 Insert the CF card in the printer CF card reader 3 In the Key operator menu select 4 Save configuration 4 Follow...

Page 404: ...hanges See 2 in Figure 23 5 When leaving the Service menu select Preventive maintenance as service action Active Current settings Factory settings Cycles to go Maintenance Action Thermal head cleaning required Replace thermal head anti dust filter Replace thermal head end of life Film calibration upper tray required Film calibration lower tray required Service maintenance required 0 19030 52520 18...

Page 405: ...counter analysis Read out and analyzation of the infocounter Image quality check Printing of STI1 STI2 and medical test image for comparison with test images at final system check Maintenance points Power off DONE Check of general condition Visual check of power and network cable Visual check of covers and film magazines Check of condition of belts and gears Visual check of all belts and gears in ...

Page 406: ...function of safety switch SW patch installation Installation of SW patch in case a recommended SW patch for the printer is available Film calibration Calibration of the films in both input trays Density Reproduction Check Density Reproduction Check to confirm proper operating CDM Execution of CDM calibration if not okay Test image evaluation Printing of STI1 STI2 and medical test image Comparison ...

Page 407: ...AR AXYS Chapter 9 3 09 2007 Type 5367 Agfa Company Confidential Completion DONE Backup Backup of machine specific parameters Reset of Maintenance action counters and relative infocounter Reset of Maintenance action counters Reset of relative infocounter Confirmation of the maintenance in the printer service menu Customer conversation Explanation of results of preventive maintenance ...

Page 408: ...ONTROL NOTE The controlled version of this document resides on MedNet Any printed copy of this document is uncontrolled Edition 1 Revision 0 DRYSTAR AXYS Chapter 9 4 09 2007 Type 5367 Agfa Company Confidential Remarks Date Signature Service Technician Customer ...

Page 409: ......

Page 410: ...cument is uncontrolled Agfa Company Confidential Document No DD DIS032 07E Copyright 2007 Agfa HealthCare N V All rights reserved Technical modifications reserved Published by Agfa HealthCare N V Tegernseer Landstraße 161 D 81539 München Germany AGFA and the Agfa Rhombus are trademarks of Agfa HealthCare N V ...

Page 411: ...led Edition 1 Revision 0 05 2008 printed in Germany Agfa Company Confidential Document Node ID 22088298 service_bulletin_cover_e_template_v01 Copyright 2008 Agfa HealthCare N V DRYSTAR AXYS Type 5367 100 Please note Service Bulletins are not part of the Service Manual for Download Please download the Service Bulletins from the GSO Library MEDNET GSO Æ Hardcopy Æ Dry Printers Æ Drystar Axys Æ Servi...

Page 412: ... CONTROL NOTE The controlled version of this document resides on MedNet Any printed copy of this document is uncontrolled Edition 1 Revision 0 DRYSTAR AXYS Chapter 10 2 05 2008 Type 5367 100 Agfa Company Confidential intentionally left blank ...

Page 413: ...Y NOTE Use dissemination distribution or reproduction of this document by unauthorized personnel is not permitted and may be unlawful DOCUMENT CONTROL NOTE The controlled version of this document resides on MedNet Any printed copy of this document is uncontrolled 12 2007 printed in Germany Agfa Company Confidential Document Node ID 14350164 eq_11_install planning_e_template_v04 Copyright 2007 Agfa...

Page 414: ...ice activities may cause damage or injuries INSTRUCTION 1 Read the Generic Safety Directions document see MEDNET GSO General Info Agfa HealthCare Publications Service Manual prior to attempting any operation repair or maintenance task on the equipment 2 Strictly observe all safety directions within the Generic Safety Directions and on the product Edition 1 Revision 1 DRYSTAR AXYS Chapter 11 2 12 2...

Page 415: ...cument This document contains all planning data including the required measures to be carried out on site prior to the machine delivery Document History Edition Revision Release Date Changes compared to previous version 1 0 1 1 12 2007 Corrected wrong outlet and cable specification for US Wrong NEMA 6 15p outlet cable Correct NEMA 5 15p outlet cable See section 8 Adapted throughput and access time...

Page 416: ...INTENANCE 11 6 SYSTEM INTEGRATION 12 6 1 Network integration parameters 12 6 2 Modality Setup Parameters 13 6 3 Setup of Security and Privacy optional 16 6 3 1 List of Key operator Service people and Specialists with access rights 16 6 3 2 Audit Logging Server and NTP Server parameters 17 6 3 3 Availability of certificates 18 6 4 Service Host Parameters 18 6 5 Example for System Integration 19 7 A...

Page 417: ...king with the latest copy of the document Edition 1 Revision 1 DRYSTAR AXYS Chapter 11 5 12 2007 5367 100 Agfa Company Confidential 9 6 Interface Data 25 9 7 Service Data 25 9 8 Dimensions and Weight 26 9 8 1 Dimensions packed machine 26 9 8 2 Dimensions closed Machine 26 9 8 3 Dimensions opened Machine 26 9 8 4 Weight 26 10 SAFETY STANDARDS AND REGULATIONS 27 ...

Page 418: ...rameters in advance to prevent a delay during installation 6 7 Get up to date Connectivity Release Documents for the modalities which shall print on the printer To be prepared for modality specific adjustments at the printer 6 2 8 Compare the required ambient and climatic conditions with the condition at the installation place To prevent installation of the printer in an inappropriate environment ...

Page 419: ...Remark for an Installation in a HIPAA conform environment In Figure 1 two additional optional modules are not listed 1 Audit logging server This is a computer in the networking allowing to store all communication between the different devices in the network 2 NTP Network Time Protocol Server This is a computer providing time synchronization It is useful for audit logging so that all messages of al...

Page 420: ... cable 5 m long 2 x RJ45 connection incl extension piece 1 Reference manual English 1 User Manual English Chinese simplified Spanish 1 CD with all user manuals and Plug Play installation manuals in all languages 1 packing list 1 Plug and Play Installation manual English 25 films 14x17 DT 2B for test purposes 25 films 8x10 DT 2B for test purposes 2 sets film type labels 1 Chinese warning sticker 1 ...

Page 421: ...a metal frame for fixing the printer on a table or floor To be used in earthquake areas and in mobile usage May not be used in military vessels 536711al cdr Mammo license EY8RN Contains the mammo license key When entering this key the printer is able to print Dmax up to 3 80 O D what is required for mammo printing NOTE Accessory Remote Service In general it is possible to diagnose and repair the p...

Page 422: ...are working with the latest copy of the document Services Service Delivery SRSS 4 Machine dimensions Closed machine 72 8 cm 28 7 67 6 cm 26 6 7 1 5 c m 2 8 2 536711jo cdr Figure 2 Opened machine 536711jp cdr 76 cm 29 9 1 1 6 c m 4 5 7 536711jn cdr Figure 3 Figure 4 Edition 1 Revision 1 DRYSTAR AXYS Chapter 11 10 12 2007 5367 100 Agfa Company Confidential ...

Page 423: ... space for installation and repair is not available the printer can be carried at a different place The printer weight is 90 kg 198 lb It can in general be carried by three people CAUTION The device is heavy 90 kg 198 lb Risk of injuries At least three people have to lift the device from one position to the other Use proper foot and hand protection when lifting the device The table might break and...

Page 424: ... 6 4 For each of these steps various parameters are required in advance They can be entered in the following tables 6 1 Network integration parameters To connect the printer to the hospital network the following parameters are required by the hospital network administrator Parameter Default Value Printer IP address 169 254 10 10 Hostname dsaxys Subnet mask 255 255 0 0 Router IP address Printer AE ...

Page 425: ...e LUT Problem The workstation at the PACS system has one AE title only The DRYSTAR AXYS does not recognize for which type of modality CR or CT the user wants to print Solution Create one logical printer DRYSTARAXYS_CR and one DRYSTARAXYS_CT Create two different host profiles for the two logical printers one with Kanamori LUT the other with linear taste LUT In the workstation setup both printers Th...

Page 426: ...y AE_title Modality 2 nickname of the modality max 8 characters letters figures Parameter Value Modality brand Modality type Modality AE_title Modality 3 nickname of the modality max 8 characters letters figures Parameter Value Modality brand Modality type Modality AE_title Modality 4 nickname of the modality max 8 characters letters figures Parameter Value Modality brand Modality type Modality AE...

Page 427: ...al Info Connectivity Application To Agfa Application Info Agfa ISRxyz pdf only via Intranet not via Internet The ISR also lists the Connectivity Release Document numbers Example see Figure 6 PHILIPS CR DRYSTAR 4500 M 5500 5300 EASYVISION RAD 4 2 Image ETHERNET II DICOM 3 0 Control ETHERNET II DICOM PRINT MANAGEMENT SCU Status Released for installation Connectivity Release Document Number Document ...

Page 428: ...es 6 3 3 NOTE The parameters for Security and Privacy have to be determined and entered by the hospital security responsible person Please hand out the pages of this subsection 6 3 to the hospital security responsible person 6 3 1 List of Key operator Service people and Specialists with access rights The list of Key operators is limited to 50 The Key Operator has access to the Key Operator webpage...

Page 429: ... max 20 characters Network password min 4 characters max 10 characters Service person name Password 1 2 3 4 The list of Specialists is limited to 1 The Specialist has access to the specialist web pages Following conventions are valid Username min 4 characters max 20 characters Network password min 4 characters max 10 characters Specialist name Password 1 6 3 2 Audit Logging Server and NTP Server p...

Page 430: ...ces which are according to the HIPAA specifications have Agfa certificates already loaded All certificates issued by Agfa have a validity period of 10 years 6 4 Service Host Parameters NOTE The AGFATEK LINK Service Host is optional in the network It is an optional part of the SRSS Secure Remote Service System For more information to the SRSS refer to MEDNET GSO Library Technical Services Service D...

Page 431: ... cdr IP address Hostname Subnet mask Router IP address Printer AE Title Domain name 192 9 200 201 dsaxys 255 255 255 0 DSAXYS Nickname Modality Brand Modality Type Calling AE Title Digitizer AGFA CR QS QS Ultrasound ACUSON ASPEN US_ASPEN Nickname Modality Brand Modality Type Calling AE Title Figure 7 7 Ambient conditions 7 1 During operation Ambient temperature min max 15 C 59 F 30 C 86 F Relative...

Page 432: ...r transport must be in accordance with EN60721 3 2 1993 class 2K4 Ambient temperature during storage min max 25 C 13 F 55 C 131 F Relative humidity during storage min max 10 95 Absolute humidity during storage min max 0 1 g m 35 g m Rate of temperature change during storage max 1 C min Air pressure during storage Min max 70 kPa 106 kPa Ambient temperature during transport min max 40 C 40 F 70 C 15...

Page 433: ...alled at a suitable point in the circuit in compliance with VDE 664 VDE Double earthed pin outlet in compliance with DIN 49441 and with CEE 7 standard cover V II Mains connection in the installation room UL Earthed contact outlet for the NEMA 5 15 P plug Euro Version CEE7 7 H05VV F3G 3 x 2 5 mm with IEC C13 plug female cable length 3 m Mains cable part of the delivery US CAN Version NEMA 5 15P SJT...

Page 434: ...opy of the document 8 1 Electrical Connections at the Printer 536611JM cdr according to IEC 601 Service Interface RS232 Interface male DTE 9 pin sub D nullmodem Network connection RJ45 UTP for 100BaseT network Equipotential plug Ground plug 6 mm metric thread 100 240 V AC power supply Figure 9 Edition 1 Revision 1 DRYSTAR AXYS Chapter 11 22 12 2007 5367 100 Agfa Company Confidential ...

Page 435: ...4x17 14x14 14x11 10x12 8x10 Throughput films hour 14x17 75 sheets hour Throughput films hour 8x10 mammo 130 sheets hour Access time1st film 14 x 17 65 secs Access time1st film 8 x 10 mammo 40 secs Printable area pixels Refer to 9 2 1 resolution DPI 508 50 µm Pixel size Bits per pixel number of grey steps 12 4096 Daylight film loading yes Films per film magazine 100 films Number of film magazines 2...

Page 436: ...880 5860 4880 5280 5860 4300 11INX14IN 5376 6922 5376 6342 6922 4796 14INX14IN 6882 6882 6882 6302 6882 6302 14INX17IN 6922 8368 6922 7788 8368 6342 Table 1 Pixel Matrices Non Mammo Printing Diagnostic Area for Mammo Printing NO annotations WITH annotations Portr Landsc Portrait Landscape Film Size Width Height Width Height Width Height 8INX10IN 3828 4958 3828 4378 4958 3248 10INX12IN 4892 5810 48...

Page 437: ...ency Hz 50 60 Power consumption W 70 standby 250 printing average 530 peak 9 4 Climatic Data Heat emission kJ h 252 standby 900 printing average 1908 maximum 9 5 Sound Emission Data Noise level in standby in operation 53 dB A maximum 62 dB A maximum 9 6 Interface Data Image control data interface DICOM 3 0 Print SCP Maximum number of simultaneous DICOM associations 10 Network connection RJ45 UTP 1...

Page 438: ...cked machine Length depth cm inch 90 35 4 Width cm inch 87 34 2 Height cm inch 69 27 2 Height incl Pallet cm inch 84 33 1 9 8 2 Dimensions closed Machine Width cm inch 72 8 28 6 Length depth cm inch 71 5 28 2 Height with output tray cm inch 67 6 26 6 Height without output tray cm inch 53 6 21 1 9 8 3 Dimensions opened Machine Width cm inch 72 8 28 6 Length depth cm inch with film trays opened 116 ...

Page 439: ... international standards and regulations Medical Devices Directive 93 42 EEC CFR Part 21 Safety standards IEC 60601 1 A1 A2 EN 60601 1 A1 A2 UL 60601 1 2003 CSA 22 2 No 601 1 M90 GB4943 2001 Radio interference regulations interference suppression FCC Rules 47 CFR part 15 subpart B class A FCC rules 47 CFT part 15 subpart C class B IEC 60601 1 2 CISPR 11 class A CISPR 22 class A IEC 61000 4 3 IEC 6...

Page 440: ......

Page 441: ..._____ Item Refer to okay 1 All parts which are required additionally to the scope of delivery e g power cables are available at the installation site 3 1 3 The planned installation place has enough space for installation operation and repair 4 and 5 4 The required system integration parameters are available 6 5 The required connectivity Release Documents are available 6 2 6 The Host System is inst...

Page 442: ...ment is uncontrolled Agfa Company Confidential Document No DD DIS031 07E re N V ons reserved ublished by re N V 161 GFA and the Agfa Rhombus are trademarks of Agfa HealthCare NV Copyright 2007 Agfa HealthCa All rights reserved Technical modificati P Agfa HealthCa Tegernseer Landstraße D 81539 München Germany A ...

Page 443: ...ny Agfa Company Confidential Document Node ID 14350512 eq_12_glossary_e_template_v04 Copyright 2007 Agfa HealthCare N V DRYSTAR AXYS Type 5367 100 Purpose of this document This document contains explanations of product specific terms and abbreviations used in the service documentation for DRYSTAR AXYS Document History Edition Revision Release Date Changes 1 0 10 2007 Initial Version Referenced Doc...

Page 444: ...367 100 Agfa Company Confidential WARNING Improper operation or service activities may cause damage or injuries INSTRUCTION 1 Read the Generic Safety Directions document see MEDNET GSO General Info Agfa HealthCare Publications Service Manual prior to attempting any operation repair or maintenance task on the equipment 2 Strictly observe all safety directions within the Generic Safety Directions an...

Page 445: ...distinction between AE title and IP address it is recommended to use only lower case characters for the hostname Preferably use the same name as for the AE title but in lowercase Adressable Area This area on film on which the thermal head can print The adressable area is smaller than the complete film as the thermal head may not print over the film i e on the drum See also diagnostic area AGFATEK ...

Page 446: ...irst two serial interfaces of a computer CPU Devil Board Printed circuit board to control the main functions of the printer CPU Term Short description D DA Converter Digital Analog Converter DCE Data Communications Equipment Term out of the RS232 specification DCE equipment sends data on pin 3 and receives on pin 2 The RS232 interface from modems or PC s is almost always DCE To connect a DCE with ...

Page 447: ...t to a network node by calling it via name instead of the IP address In the internet for example the IP address 19 5 6 121 can be contacted by the name johnson com To identify a device in the network via a name it has to be registered in a DNS server The network system administrator determines the domain name for DRYSTAR AXYS DPI Dots Per Inch Pixels per Inch 1 inch 2 54 cm DRA Direct Remote Acces...

Page 448: ...the printer gets a table with certain density values for certain control values Film type Film of a certain producer or a similar producer but with different characteristics sensitivity contrast FireWire Serial bus standard also known as IEEE1394 Data rate up to 400 Mb s Flash EPROM An EPROM in which data have to be deleted in fixed blocks instead of separate bytes Fog level Lowest optical density...

Page 449: ...devices a host system is the device at which the printer is going to be connected to Hyper Terminal A Windows program used for serial data communication between two computers Term Short description I ICM Image Composition Map Complete image data for a complete film These image data correspond to the several pixels on the film that is going to be printed IEC International Electro technical Commissi...

Page 450: ...WW Term Short description K KEY6 Assy Printed circuit board to control the printer keypad Term Short description L LCD Display Liquid Crystal Display By supplying a voltage the liquid crystals are arranged such that a contrast is visible LED Light Emitting Diode Diode which emits green yellow or red light LUT Look Up Table contrast reference curve e g Kanamori LUT Term Short description M Mammogra...

Page 451: ... up an image on the film E g a 10x12 film contains 4940 x 5900 pixels adressable area PMS DICOM Print Management Service A device that provides DICOM PMS provider makes printing of DICOM images possible A device using DICOM PMS user can control a DICOM printer DRYSTAR Axys is a DICOM PMS provider Portex AGFA own operating system for medical devices Postscript Imperative programming language for th...

Page 452: ...s known by the printer it is possible for the printer to send information via the network back to devices which are located in another segment of the network RS232 Interface Computer interface to convert computer internal parallel data into serial data and vice versa The interface is defined via RS232 Recommended Standard RS232 cable Cable following the RS232 standard Recommended Standard to conne...

Page 453: ...PG are more and more superseding TIF TH Abbreviation of Thermal Print Head Thanos board Board on the Devil Board Responsible for operating system and image processing Thermal Print Head Resistor line that transfers the image to the film Term Short description U USB Universal Serial Bus PC bus system to attach peripherals like scanner keyboards etc to PC s Data rate 12 Mb s and up to 480 Mb s for U...

Page 454: ...This page is intentionally left blank to enable print out of even and odd pages on duplex printers ...

Page 455: ...DRYSTAR AXYS Type 5367 100 1st Edition r Orde no I 0 DD D S 33 07E ...

Page 456: ......

Page 457: ...Node ID 16099429 Copyright 2008 Agfa HealthCare N V Purpose of this Document This document describes how to add the order list for documentation print add comments Document History Edition Revision Release Date Changes compared to previous Version 1 1 1 2 06 2008 Added how to integrate the order list for documentation in a printed manual See section 1 ...

Page 458: ...reating a paper manual Print the order list and put it behind the cover sheet 2 Printing single Pages Chapters or the complete Manual IMPORTANT Preferably print this manual double sided This PDF manual contains empty pages at the end of several chapters to have the next chapter starting with an uneven page number when printed doubles sided If printed one sided dispose these empty pages Preferably ...

Page 459: ...ick the bookmark of the desired chapter 2 Write down or remember the shown PDF page number See Figure 1 Figure 1 PDF page number in the Adobe reader toolbar 3 Go the end of the section or desired range of pages 4 Select Print 5 Select the page range 6 Select Reduce to printer margins and Auto rotate and Center 7 Select OK Figure 2 Print dialogue for printing single pages or chapters Edition 1 Revi...

Page 460: ... print the complete service manual proceed as follows 1 Select Print 2 Select All 3 Select shrink to printable area and Auto rotate and Center see NOTE above 4 Select OK Figure 3 Print dialogue for printing the complete manual 3 Adding Comments If you open this file in an Adobe Reader version 7 the comment toolbar will show up This allows adding comments to highlight or underline text and many mor...

Page 461: ...menu Comment Markup 2 Select the desired comments Press the CTRL key for multiple selections 3 Select Options Export Selected Comments 4 Save the file with any name 3 2 Importing Comments 1 In the drop down menu Comment Markup select Show comments List 2 Select Options Import Comments 3 Browse for the comments file and press select NOTE The imported comments possibly appear on different pages if t...

Reviews: