Recorder Mode - The “Trigger” Menu
VII - 144
SCOPIX
Recorder Mode (cont’d.)
Display
(fault capture,
file capture)
Two cases arise:
•
the horizontal zoom is enabled,
•
the horizontal zoom is disabled.
Horizontal zoom
disabled
Modification of display:
•
The cursors are no longer displayed.
•
The measurements displayed do not have the same significance.
•
All the measurements are related to the current recording.
•
“Trigger time” and “Sample value at trigger time” measurements are only
displayed if the trigger has taken place.
•
Maximum and minimum searches cannot be enabled (see §. Display Menu,
p. 146).
•
The automatic measures cannot be enabled (see §. Measurement Menu,
p. 148).
Horizontal zoom
enabled
The display is identical to normal mode.
•
The measurements displayed have the same significance as in normal mode
(see §. Display, p. 146).
•
Maximum and minimum searches are possible (see §. Display Menu, p. 146).
•
Automatic measurements can be enabled (see §. Measurement Menu, p. 148).
Value of
last
acquired
sample
Value of
sample at
time of
trigger
Current
recording
Acquisition time
of first viewable
sample
Trigger time
Acquisition
time of last
acquired
sample
faults
www.
ShopAEMC
.com
Shop for AEMC products online at:
1.877.766.5412
Summary of Contents for OX 7102 III
Page 117: ...Multimeter Mode SCOPIX V 117 www ShopAEMC com Shop for AEMC products online at 1 877 766 5412...
Page 155: ...Recorder Mode SCOPIX VII 155 www ShopAEMC com Shop for AEMC products online at 1 877 766 5412...
Page 199: ...Applications SCOPIX IX 199 www ShopAEMC com Shop for AEMC products online at 1 877 766 5412...