Oscilloscope Mode - The "Trigger" Menu
IV - 74
SCOPIX
Oscilloscope Mode (cont'd)
Example
Signal injected on CH1: a train of five 6 V
DC
pulses at a frequency of 20 kHz
separated by 500 µs.
The trigger is set on the descending edge.
The first edge activates the trigger. It is not included in the count.
The trigger is triggered on the third descending edge of the pulse train.
The Holdoff stabilizes the signal by disabling the trigger for 232 µs.