Applications
SCOPIX
IX - 181
Applications (cont’d)
11. Min / Max
Acquisition
Detection of a wrong representation due to under-sampling
Examination of an amplitude-modulate signal
•
Using the key shown
opposite
, select the "Oscilloscope" mode.
•
Inject on the CH1 input an amplitude-modulated sinusoidal signal at a.
frequency of 15Hz.
•
Inject on the CH4 input a sinusoidal signal at 300 Hz and 3 V
DC
.
for synchronization of the signal on CH1.
•
Adjust the scan speed:
using the scrollbar in the time base window: 1 ms/div
or using the keys
opposite.
•
Validate the CH1 signal: Vert Menu Display Trace 1
or using the key
opposite.
or on the display of the CH1 trace parameters.
•
Adjust the CH1 sensitivity:
Vert
Menu
CH1 Sensitivity/coupling: 200 mV/div
or using the keys
opposite.
•
Ditto for CH4 with a sensitivity of: 500 mV/div
•
Adjust the trigger parameters:
Trig.
Menu
Parameter Main:
Source:
CH4
Coupling: AC
Front: + or using the key
opposite.
•
Adjust the trigger mode:
Trig.
Menu
Auto Mode:
or using the SGLE REFR key.
•
Start acquisition using the RUN/HOLD key (RUN mode).
•
Use the FULL TRACE key to display the 2 traces separately.
Examination of the signals
www.
ShopAEMC
.com
Shop for AEMC products online at:
1.877.766.5412
Summary of Contents for OX 7102 III
Page 117: ...Multimeter Mode SCOPIX V 117 www ShopAEMC com Shop for AEMC products online at 1 877 766 5412...
Page 155: ...Recorder Mode SCOPIX VII 155 www ShopAEMC com Shop for AEMC products online at 1 877 766 5412...
Page 199: ...Applications SCOPIX IX 199 www ShopAEMC com Shop for AEMC products online at 1 877 766 5412...