Technical Specifications
SCOPIX
X - 209
Technical Specifications (cont'd)
Accessories
HX0072 - AmpFLEX Probe
Standards applied
IEC 61010-2-032 : 2002
EN 61326-1 (07/1997) + A1 (10/1998) + A2 (09/2001)
Reference conditions
Only one conductor inserted in the flexible toroid
Conductor position:
centred
Clamping :
∅
240 mm
Temperature :
from 18°C to 28°C
Relative humidity:
from 20 % to 75 %
Frequency range:
40 Hz to 400 Hz
Start-up before measurement:
1min
External DC magnetic field:
< 40 A/m
No external AC magnetic field
No external electric field
Sinusoidal signal
Use conditions
Altitude < 2000 m, indoors
Range for use
from 1 A to 3500 A
RMS
Specified range
from 5 A to 3000 A
RMS
Accuracy in the
measurement range
1 % ± 0.5 A
50 Hz dephasing
1.3° max. (1° typ.)
Residual current
at I = 0 A (noise)
1.5 A
RMS
max. (0.5 A
RMS
typ.)
Bandwidth at -3 dB
10 Hz to 200 kHz
Power-up and to-idle
time
1.5 µs
Residual DC current
20 A max. (invisible with AC coupling)
Delay time
1.2 µs max.
Frequency derating
3000 A if 10 Hz < Freq. < 10 kHz
50A if Freq. = 200 kHz
Electromagnetic
immunity at 10 V/m
error < 3 % of measurement extent
Operating temperature -10°C to +55°C
HX0073 - MiniAmpFLEX
Probe
Standards applied
IEC 61010-2-032 : 2002
EN 61326-1 (07/1997) + A1 (10/1998) + A2 (09/2001)
Reference conditions
Only one conductor inserted in the flexible toroid centred
Conductor position:
centred
Clamping :
∅
35 mm
Temperature :
from 18°C to 28°C
Relative humidity:
from 20 % to 75 %
Frequency range:
40 Hz to 400 Hz
Start-up before measurement:
1min
External DC magnetic field:
< 40 A/m
No external AC magnetic field
No external electric field
Sinusoidal signal
Use conditions
Altitude < 2000 m, indoors
Range for use
from 0.2 A to 350 A
RMS
Specified range
from 1 A to 300 A
RMS
Accuracy in the
measurement range
1 % ± 70 mA
50 Hz dephasing
1.3° max. (1° typ.)
Residual current at
I = 0 A (noise)
0.2 A
RMS
max. (0.1 A
RMS
typ.)
BW min.at -3 dB
from 10 Hz to 2.5 MHz (from 10 Hz to 3 MHz typ.)
Power-up and to-idle
time
< 110 ns
Residual DC current
2 A max. (invisible with AC coupling)
Delay time
600 ns max.
Frequency derating
300 A if 10Hz < Freq. < 100 kHz
10 A if Freq. > 1 MHz
Electromagnetic
immunity at 10 V/m
error < 3 % of measurement extent
Operating temperature -10°C to +55°C
www.
ShopAEMC
.com
Shop for AEMC products online at:
1.877.766.5412
Summary of Contents for OX 7102 III
Page 117: ...Multimeter Mode SCOPIX V 117 www ShopAEMC com Shop for AEMC products online at 1 877 766 5412...
Page 155: ...Recorder Mode SCOPIX VII 155 www ShopAEMC com Shop for AEMC products online at 1 877 766 5412...
Page 199: ...Applications SCOPIX IX 199 www ShopAEMC com Shop for AEMC products online at 1 877 766 5412...