Oscilloscope Mode - The "Measure" Menu
SCOPIX
IV - 85
Oscilloscope Mode (cont'd)
It is possible to select two permanent measurements.
The "
"
symbol indicates the measurement(s) that will be indicated
in the status area.
Activation of the automatic measurements causes two markers (+) to
appear on the curve, if at least one period is visible on the screen.
The display order corresponds to the chronological order of the
selection and the markers are assigned to the first measurement
selected.
Automatic measurements in the status area can be deleted by means of
this menu, by erasing the selected measurements (no "
" symbol in the
automatic measurements table).
Reference memory
difference
Activation of the "Reference memory difference" option is a way of
calculating the deviations, for all the automatic measurements, between
the selected trace and the memorized reference trace (see §. Memory
Menu).
Example
Calculation performed and displayed on one of the 20 automatic
measurements:
Vpp (Ref. memory difference) = Vpp (Trace 1) – Vpp (Trace 1
Ref 1)
The calculation is performed in the same way for all the measurements.
This option is only active if a reference trace is present.
It must correspond to the trace on which you wish to perform
automatic measurements
(
Example: Trace 1 and Trace 1
Ref. 1).
Condition: the reference trace must have the same characteristics as
the associated trace (sensitivity and time base)
20 automatic
measurements
Vmin
minimum peak voltage
Vmax
maximum peak voltage
Vpp
peak-to-peak voltage
Vlow
established low voltage
Vhigh
established high voltage
Vamp
amplitude
Vrms
rms voltage
Vrms_c
Rms voltage performed on an integer number of cycles
Vavg
average voltage
Over+
positive overshoot
Trise
rise time
Tfall
fall time
W+
positive pulse width (at 50 % of Vamp)
W-
negative pulse width (at 50 % of Vamp)
P
period
F
frequency
DC
cyclic ratio
N
number of pulses
Over-
negative overshoot
Sum
summon of instantaneous values of the signal
www.
ShopAEMC
.com
Shop for AEMC products online at:
1.877.766.5412
Summary of Contents for OX 7102 III
Page 117: ...Multimeter Mode SCOPIX V 117 www ShopAEMC com Shop for AEMC products online at 1 877 766 5412...
Page 155: ...Recorder Mode SCOPIX VII 155 www ShopAEMC com Shop for AEMC products online at 1 877 766 5412...
Page 199: ...Applications SCOPIX IX 199 www ShopAEMC com Shop for AEMC products online at 1 877 766 5412...