Applications
IX - 172
SCOPIX
Applications (cont’d)
If necessary:
•
Modify the trigger level using the stylus:
- by moving the symbol T (Trigger) on the screen. The value of the trigger
level is indicated in the bottom right-hand corner of the screen.
- or through the trigger parameters menu:
Trig. Menu Parameter Main Level
•
Modify the vertical positioning of the curve:
- by using the stylus to move symbol 1, on the left of the screen.
- or by using the keys
opposite.
The key opposite is used to make these adjustments automatically.
HX0030 probe
compensation
Adjust the low-frequency compensation of the probe so that the signal plateau is
horizontal (see figure below).
Over-compensated probe
Compensated probe
Under-compensated probe
Adjust the screw on the Probix HX0030 probe to adjust the compensation.
www.
ShopAEMC
.com
Shop for AEMC products online at: