Description of the instrument
II - 18
SCOPIX
Description of the instrument (cont'd)
Charging the
battery
*
Once the battery is located in its compartment, follow the external power supply
start-up instructions.
•
To speed up recharging of the battery, switch off the power to the
oscilloscope by a long press on the ON/OFF button.
•
Battery fully discharged: during the first 10 minutes, the
CHARGE
LED of the
external power supply flashes, indicating that the battery is slow-charging.
The
CHARGE
LED then lights up when it switches to fast-charging.
During the loading of the battery (after 15 min), the startup of the
oscilloscope causes the premature stop of the load. It is possible to re-
start the loading while disconnecting, then reconnecting the charger.
The LED goes out when the battery is fully charged (approx. 2h30).
Charging the
battery while
using the
oscilloscope
When the oscilloscope is connected to the mains via the external power supply, it
is possible to perform slow-charging of the battery.
The
CHARGE
LED of the external power supply lights up. Full recharging of a
totally discharged battery then takes approx. five hours and a half. The LED goes
out when the battery is fully charged.
*
Frequently recharging a battery when it is not fully discharged reduces its
life span.
Powering the
instrument with
the battery
When the oscilloscope is powered by a battery, an charge status indicator is
included in the display area of the current value:
the battery is 100% charged
the battery is 80% charged
the battery is 60% charged, etc.
this symbol indicates that there are only a few minutes more,
so you are advised to recharge it or switch to the mains supply.
The battery is totally flat and the screen is about to be
deactivated. You must either recharge the battery or connect the
instrument to the external power supply.
If you change battery, a complete battery charge and discharge cycle (until the
instrument is automatically deactivated when discharging has finished) is
essential to calibrate this indicator.
To maintain the battery in good condition, use the oscilloscope at least
until the level
before recharging.
www.
ShopAEMC
.com
Shop for AEMC products online at:
1.877.766.5412
Summary of Contents for OX 7102 III
Page 117: ...Multimeter Mode SCOPIX V 117 www ShopAEMC com Shop for AEMC products online at 1 877 766 5412...
Page 155: ...Recorder Mode SCOPIX VII 155 www ShopAEMC com Shop for AEMC products online at 1 877 766 5412...
Page 199: ...Applications SCOPIX IX 199 www ShopAEMC com Shop for AEMC products online at 1 877 766 5412...