Oscilloscope Mode - The "Measure" Menu
IV - 88
SCOPIX
Oscilloscope Mode (cont'd)
Unattached manual
cursors
Used for linking or not linking the manual measurement cursors (1 and 2)
to the reference trace.
When the "Unattached cursors" menu is selected, cursors 1 and 2 can
be moved freely over the whole screen.
•
The "
"
symbol indicates that the "Unattached cursors" menu is active.
•
To deactivate this menu, deselect it by pointing with the stylus.
In the case of "Automatic measurements" and manual measurement
activation:
If …
then …
… the
manual
cursors and the
automatic
markers are
displayed together,
… the
automatic
measurements
are performed on the portion of the
trace defined between the
manual
cursors.
… the portion defined between
the
manual
cursors is too
small [in this case, the fixed
markers (+) will not be
displayed],
… the
automatic
measurements
selected are impossible, so "-.--" is
indicated in the measurement
display area.
Deselect the automatic measurements to validate the manual
measurements (dt, dv).
www.