Applications
SCOPIX
IX - 179
Applications (cont’d)
•
The Table "Ref.: Harmonic X" indicates, for the harmonic selected:
-
its value as a % of the fundamental
-
its phase in ° with respect to the fundamental
-
its frequency in Hz
-
its RMS voltage in V
Example of harmonic breakdown
•
On CH1, inject a rectangular signal with frequency 100 Hz and
amplitude 5 V.
•
On CH4, inject a triangular signal with frequency 100 Hz and
amplitude 5 V.
•
Use the key
opposite
to perform automatic optimum adjustment of
the channels, the ranges, the time base and the trigger.
"Oscilloscope" Mode
Use the FULL TRACE key to display the 2 traces separately.
"Harmonic Analysis" Mode
•
Press the key
opposite
to activate the "Harmonic Analysis" mode.
•
Use the "Display" menu to select the required harmonic breakdown:
Example: Odd harmonics
•
Use the stylus to select the fundamental (F) or one of the harmonics to
perform the measurement:
Example: Harmonic 3