Technical Specifications
X - 206
SCOPIX
Technical Specifications (cont'd)
Oscilloscope Mode
Display
Characteristics
OX 7102
III
- OX 7104
III
OX 7202
III
- OX 7204
III
Display
screen
LCD 5.7’’ TFT (color display)
LED
back-lighting
Light
Continuous adjustment
Resolution
1/4 VGA, i.e. 320 pixels horizontally x 240 pixels vertically
Screen saver
Delay can be selected in the Util Menu Configuration
15‘, 30’, 1 hour or none
Display modes
Vector
Envelope
Averaging
All acquisition
Acquired samples, interpolated samples, averaging, linear interpolation
between 2 acquired samples.
Display of min. and max. on each abscissa, acquired in several bursts
Range of factors: none, 2, 4, 16, 64
Display of all the samples acquired in a burst with linear interpolation
Graticule
Complete or Edges
Indications on screen
Triggering
Traces
Trigger level position (with coupling and overshoot indicator)
Position of the Trigger point on the bargraph and on the top edge of the
screen (with overshoot indicators)
Trace identifiers, activation of the traces
Position, Sensitivity
Ground reference
High and low overshoot indicators if traces are off screen
Miscellaneous
1/10th probe calibration
signal
Form: rectangular
Amplitude:
≈
0- 3 V
Frequency:
≈
1kHz
Dual insulation / channels: 600V CAT III, 1000V CAT II
Connect the cold point of the probe to the cold point of the
probe calibration output.
Autoset
Search time
Frequency range
Range of amplitude
Cyclic ratio limits
< 5 s
> 30 Hz
15 mVpp to 400 Vpp
from 20 to 80 %
www.
ShopAEMC
.com
Shop for AEMC products online at:
1.877.766.5412
Summary of Contents for OX 7102 III
Page 117: ...Multimeter Mode SCOPIX V 117 www ShopAEMC com Shop for AEMC products online at 1 877 766 5412...
Page 155: ...Recorder Mode SCOPIX VII 155 www ShopAEMC com Shop for AEMC products online at 1 877 766 5412...
Page 199: ...Applications SCOPIX IX 199 www ShopAEMC com Shop for AEMC products online at 1 877 766 5412...