Section 7
Fluid-In Control Assembly
Abbott VP 2000™ Processor Service Manual
7-11
30-608308/R1—September 2008
3. Remove the Drip Tray located in the lower center of the
unit by grasping the flange on the tray from the rear of the
unit and lifting upward and out.
4. Remove the AC Power Barrier. Loosen three (3) 6-32 x 1/4
Phillips-Head screws located at the ends of the Power
Barrier, and slide the Barrier back and up to remove.
5. Unplug the Fluid-In control solenoid connector at the
System Harness Shroud.
6. Disconnect the green ground wire on the Fluid-In solenoid
by removing the 6/32 hex nut that fastens it to the base
pan.
7. Remove the flexible plastic tubing from the Solenoid
Valve Output Fitting.
8. Cut the cable tie fastening the Fluid-In plastic tube to the
bottom of the pan.
9. Leaving the Fluid-In line attached, remove the two (2)
6-32 hex nuts that secure the Fluid-In Control Assembly to
the base pan, and remove the assembly.
Installation
Reassembly by following the removal procedure in reverse.
NOTE:
When reinstalling the flexible tubing, moisten
the inside diameter of the tube to ease installation.
Ensure that the tubing is fully seated on it barbed
fitting.