Section 5
Level II - Field Service Maintenance
Abbott VP 2000™ Processor Service Manual
5-15
30-608308/R1—September 2008
Lubricate
Y-Drive Gear
(Refer to Section 7, Repair,Replacement of
Spare Parts, Y-Axis Shaft Lubrication
Procedure)
Filter Bearing
(Refer to Oiling the Blower Fan in this
section)
Inspect
Hoods
Sensor Bar (inspect wiring for
corrosion; replace as necessary)
Interlock Switches
Replace
Fluid System Tubing
(Refer to Section 7, Repair, Replacement of
Spare Parts, Fluid-In Control Assembly)
Drain Tube O-ring
(Refer to Section 7, Repair, Replacement of
Spare Parts, Drain Tube O-Ring)
Filter Cassette
(Refer to Replacing the Charcoal Cassette in
this section)
Test
Manual Movement of Robotic
Arm
(Refer to Section 5, Maintenance and
Diagnostic Procedures)
Perform Maintain Test
(Refer to the Abbott VP 2000 Processor User
Guide, Section 6, Service and Maintenance,
Manually Moving the Robotic Arm)
Basin Temperature Check
(Refer to Heated Basin Temperature Check
in this section)
Drying Station Temperature
Check
Refer to Drying Station Temperature Check
in this section)
Table 5.2:
Preventive Maintenance Tasks
Action
Item
Done/P
assed
Failed
Comments