Maintenance and Diagnostic Procedures
Section 5
5-46
Abbott VP 2000™ Processor Service Manual
30-608308/R1—September 2008
27. Place the Robotic Arm in the up position at front of the
slide stainer, verify multimeter readings at:
28. Place the Robotic Arm in the down position at the front of
the slide stainer, verify multimeter readings at:
29. Place the Robotic Arm in the up position at the rear of the
slide stainer, verify multimeter readings at:
30. Place the Robotic Arm in the down position at the rear of
the slide stainer, verify multimeter readings at:
U15
Pin 11 Of5v
.05v
U15
Pin 17 Of < .2v
U15
Pin 16 Of5v
.05v
U15
Pin 15 Of < .2v
U15
Pin 11 Of
< .2v
U15
Pin 17 Of5v
.05v
U15
Pin 16 Of5v
.05v
U15
Pin 15 Of < .2v
U15
Pin 11 Of5v
.05v
U15
Pin 17 Of < .2v
U15
Pin 16 Of < .2v
U15
Pin 15 Of5v
.05v
U15
Pin 11 Of < .2v
U15
Pin 17 Of5v
.05v
U15
Pin 16 Of < .2v
U15
Pin 15 Of5v
.05v