Section 7
Sensor Bar Assembly
Abbott VP 2000™ Processor Service Manual
7-17
30-608308/R1—September 2008
Sensor Bar Assembly
Prior to Disassembly move the Robotic Arm so that it is over
the Drying Station and leave it in the up position.
WARNING:
Always disconnect the system power at the
plug to the wall socket before beginning disassembly of
the unit.
Removal
1. Remove the Front Hood by gently squeezing inward on
the two (2) metal arms that connect the Front Hood to the
unit.
2. Open the Heater Module Cover and disconnect the Sensor
Bar Interface Cable.
Figure 7.5: Sensor Bar Interface Cable
3. Remove the two (2) 6-32 x 1/4 Phillips-Head screws from
the right-side of the Heater Module.