Section 5
Electrical Test
Abbott VP 2000™ Processor Service Manual
5-41
30-608308/R1—September 2008
14. Push “TEST” - “RUN” - “1” - “ENTER”. A built-in test will
start. When the arm moves past Bath 2, place the power
switch in the OFF position. The power interrupt alarm
should continue to sound. The arm will return to Bath 1.
Push “CLEAR” and the power interrupt alarm should go
OFF.
Check off the following block on the checklist:
POWER INTERRUPT ALARM
15. Place the UUT power switch in the ON position. When the
control panel displays “READY” press “TEST” - “ENTER”
keys. Next, press the following keys and observe the
number symbol that appears in the Bath display. Verify
each number.
Table 5.4:
Bath Display Number Symbols
Key
Number
PROG
13
REVIEW
20
PAUSE
16
RUN
14
AGITATE
19
15
ABORT
21
UP ARROW
22
LEFT ARROW
25
RIGHT ARROW
24
DOWN ARROW
23
1
1
4
4
7
7
CLEAR
12
2
2