Section 4
Pre-Installation Checklist
Abbott VP 2000™ Processor Service Manual
4-11
30-608308/R1—September 2008
Site Environmental Requirements
Miscellaneous
Operation temperature:
15-30 C (59-86 F)
Operating altitude:
Max 2000 m (6562 ft)
Operation humidity:
15-85% relative (non-condensing)
Pollution degree:
2
Over Voltage Category:
II
NOTE
—Avoid placing instrument near air vents or drafts
Uninterruptible Power Supply
(UPS) Requirements
It is recommended that the customer use the following UPS:
Additional Questions
Does the customer require a certificate of installation (IQ/OQ)? _________________________________________________________
Action plan to resolve inadequacies or concerns:
Customer Signature/Title and Date
Abbott Representative Signature and Date
Fax completed report to:
Abbott VP 2000 Processor
Product Manager
1.224.361.7505
Questions? Call: 1.224.361.7375 or, if in Europe, the Middle East, or Africa, contact the ESS
Service and Support team in Delkenheim, Germany
Page 3 of 3