Section 6
Maintenance Diagnostic Faults
Abbott VP 2000™ Processor Service Manual
6-45
30-608308/R1—September 2008
No Heat in Dryer: Fault 19.0
1. Power OFF, center cover and AC power barrier removed.
Check that the two (2) connectors from the dryer are
plugged into the harness shroud.
2. Power OFF, center cover and AC power barrier removed.
Unplug the dryer connector and check that the resistance
between Pin 1 and Pin 2 of the plug is 43
3.
3. Power ON, center cover and AC power barrier removed.
WARNING: Electrical Shock Hazard.
Move Robotic
Arm to Bath 1 in the down position. Unplug the dryer
connector. Check for 120V AC between Pin 1 and Pin 2
of the harness.
4. Power ON, top cover removed.
WARNING: Electrical Shock Hazard.
Use digital volt
meter (set for DC volts). Check voltage across R26 on
computer board [2.94 VDC].
5. Power ON, top cover removed.
WARNING: Electrical Shock Hazard.
Use digital volt
meter (set for DC volts).