Section 6
Maintenance Diagnostic Faults
Abbott VP 2000™ Processor Service Manual
6-37
30-608308/R1—September 2008
Figure 6.13: Fluid Will Not Recirculate
FLUID WILL NOT RECIRCULATE
KEYPAD RECIRCULATE INDICATOR
LIGHT ON
ACTIVATE FLOW BUTTON
(RED LIGHT WILL GO ON)
ROBOTIC ARM IN BATH 2 & DOWN
CHECK CONNECTORS
REPLACE CONTROL
MODULE
MOVE TO BATH 2 & DOWN
REALIGN & PLUG IN
YES
NO
YES
NO
YES
NO
PASS
FAIL
POWER SUPPLY BOARD
TEST BY SUBSTITUTION
COMPUTER BOARD
TEST BY SUBSTITUTION
REPLACE PUMP ASSY