Section 7
Sensor Bar Interface Cable
Abbott VP 2000™ Processor Service Manual
7-19
30-608308/R1—September 2008
Sensor Bar Interface Cable
Prior to Disassembly move the Robotic Arm so that it is over
the Drying Station and leave it in the up position.
WARNING:
Always disconnect the system power at the
plug to the wall socket before beginning disassembly of
the unit.
Removal
1. Remove the Hoods by gently squeezing inward on the two
(2) metal arms that connect the Hoods to the unit.
Remove the Carrier-Tray Assembly.
2. Remove the Drip Tray, located in the center of the unit, by
grasping the flange on the tray from the rear of the unit
and lifting upward and out.
3. Remove the AC Power Barrier. Loosen three (3) 6-32 x 1/4
Phillips-Head screws located at the ends of the barrier, and
slide the barrier back and up to remove.
4. Unplug the Heater Module from the System Harness
Shroud.
5. Disconnect the green ground wire by removing the 6-32
hex nut securing the wire to the base pan.
6. Remove the two (2) 4-40 hex nuts securing the Heater
Module to the front base. Slide the Heater Module forward
and lift away from the front base.
7. Disconnect the Sensor Bar Interface Cable from the Sensor
Bar.