Repair, Replacement of Spare Parts
Section 7
7-30
Abbott VP 2000™ Processor Service Manual
30-608308/R1—September 2008
Removal
1. Remove the hoods by gently squeezing inward on the two
(2) metal arms that connect the hoods to the unit.
2. Remove the Top Cover by unscrewing six (6) 6-32 x 1/4
Phillips-Head screws, three (3) on each side.
3. Unplug the connectors at J8, J9, J10, J11 and J12 on the
Computer Board Assembly.
4. Remove the four (4) 6-32 x 3/8 Phillips-Head screws that
fasten the Computer Control Board Assembly to the Drive
Plate. Carefully pull the assembly away from the Drive
Plate to disengage connector P4 (on the back of the
Computer Control Board) from the X-Sensor Board
connector. Then remove the Computer Control Board
Assembly.
Installation
Reassemble by following the removal procedure in reverse.
NOTE:
Make sure the Computer Control Board
Connector, P4, is aligned with the mating connector on
the X-Sensor Board Assembly. Also ensure that the four
(4) mounting holes are aligned with the threaded
standoffs on the Drive Plate before pausing the
Computer Control Board Assembly into position.
CAUTION:
Before sliding the Top Cover back into
position, make sure all wiring, cables and connectors
are replaced snugly, within the unit and not hanging
out where they might be sheared off when replacing
the cover.