Section 6
Maintenance Diagnostic Faults
Abbott VP 2000™ Processor Service Manual
6-23
30-608308/R1—September 2008
Test Control Module Lights All Eights, Arm Does Not Move: Fault 4.0
1. Power ON, top cover removed.
WARNING: Electrical Shock Hazard.
Remove one
wire from each switch to disconnect interlock switch
from system.
2. Power OFF, top cover removed.
Check that all plug-in connectors [J1, J2, J3, J7, J8, J9, J10,
J11, J12 and J13] are correctly aligned and fully seated.
3. Power OFF, top cover removed.
Test by substituting knows good units for computer
board, power supply board, and test control module.