Section 6
Fault Diagnostics
Abbott VP 2000™ Processor Service Manual
6-11
30-608308/R1—September 2008
No Y-movement of Robotic Arm
• Robotic Arm
• Y-motor Assembly
• Power Supply
• Computer Control Board
No motion in one direction
• Power Supply
• Check wiring harness
• Drive Belt
• Motor
Robotic Arm stops operation during a
program
• Try again
• Computer Control Board
• X-Sensor Board
Water Bath does not fill
• Source Connection
• Installation of Standpipe
Flooding
• Check drain for kinks/blockage
• Check elevation of drain tube
• Check all tubing connections
No water flow
• Source Connection
• Solenoid regulator
• Power Supply
• Computer Control Board
Chatter of Squeal with fluid flows
Adjust regulator
No recirculation
• Pump Assembly
• Power Supply
• Computer Control Board
No power to dryer
• Temperature Setting
• Drying Assembly
• Power Supply
• Computer Control Board
• Thermal Fuse Air Intake to Dryer
Table 6.1:
Symptom and Faults
Fault
Check/Replace