![background image](http://html1.mh-extra.com/html/abbott/vp-2000-processor/vp-2000-processor_service-manual_450983103.webp)
Section 5
Level I - User Maintenance
Abbott VP 2000™ Processor Service Manual
5-13
30-608308/R1—September 2008
Oiling the Blower Fan
Oil the Charcoal Filter’s blower fan inside the filter unit every
six (6) months.
1. Open the Charcoal Filter canister by unlatching the top of
the canister.
2. Carefully remove the Charcoal Cassette and set it to the
side.
3. Locate the two (2) oiling tubes on the side of the Blower
Fan in the Charcoal Filter canister.
4. Using a general-purpose 20-weight oil, fill the two (2)
oiling tubes.
5. When the oil tubes are full, set the Charcoal Cassette back
into the Charcoal Filter canister.
6. Replace the top of the Charcoal Filter canister and close
the latches.
Check for obstructions in the tubing from the instrument to
the filter system on an as needed basis.
Inspection of the System
1. Periodically inspect the ducting for tears or degradation.
2. When replacing the filter cassettes, inspect the bottom of
the canister for spilled carbon granules or dust and
vacuum clean if required.