Section 6
Maintenance Diagnostic Faults
Abbott VP 2000™ Processor Service Manual
6-33
30-608308/R1—September 2008
System Will Not Execute a Program: Fault 11.0
1. Double-sided unit:
Check that the carriers installed in the front and rear of
the system are the same type, i.e., 600 ml or 12 ml.
All units:
Review the program to confirm that the bath positions
are valid for the type of carriers installed.
•
Example: A single-sided HMS/MS-100 has a total of six
(6) bath positions - a program directing the Robotic
Arm to go to bath eight (8) would be invalid.
•
Example: A double-sided HMS/DS-50 has a total of 22
bath positions, but a unit with an MS-50 carrier in the
front and DS-100 carrier in the rear would have a total
of 14 vessels - a program directing the Robotic Arm to
go to bath 16 would be invalid.
2. Review program according to the
Abbott VP 2000
Processor User Guide
.
3. Enter new program; shut machine OFF for 24 hours, then
turn machine ON an review program.
4. Power ON, top cover removed.
WARNING: Electrical Shock Hazard.
Test bucket
select sensors according to the Sensor Test Program in
this section.
5. Power OFF, top cover removed.
Test by substituting known good units for computer
board and test control module.