2101510 Rev. AG
Page 3–23
8)
Tighten both nuts.
Figure 3-24
Single and Multiple Stream Sample Conditioning Assemblies
3.14 Sample Line Connections
Following the installation of the sample conditioning module(s), install the sample
tubing from the sample probe to the sample conditioning system and NGC feed-
through assembly.
If the sample conditioning module and NGC are located
inside an ENC82, review the following sub section,
to NGC Inside of the ENC82, for
information pertaining to this installation.
3.14.1 Materials
•
1/8” SST chromatography grade transport tubing. Length of tubing to be
determined by the technician based on the distance from the sample probe to
the sample conditioning module and the number of sample streams.
•
2 ea. ferrule and nut (for each sample stream)
•
1 ea. ¼” NPT to 1/8” reducer or other size as determined from the sample
probe output port (for each sample stream)
•
1 ea. sample conditioning module transport tubing (supplied with sample
conditioning module).
3.14.2 Instructions
Be certain that the ends of stainless steel tubing are open
and not restricted.