![ABB NGC8206 User Manual Download Page 54](http://html.mh-extra.com/html/abb/ngc8206/ngc8206_user-manual_4215054.webp)
2101510 Rev. AG
Page 2–35
Part Number
Description
•
Particulate/coalescing filter
•
Liquid/vapor separator
2104677-001
Designed to remove condensing sample line liquid or water that has penetrated the
sample probe. The system expels the liquid or water downward into the bottom of the
Armstrong trap drainer via inertial separation. Sample gas is extracted from the fitting at
the top of the Armstrong accumulator. The pressurize accumulator expels collecting liquid
from the bottom by a modulation float valve design that triggers liquid build-up in the trap
accumulator.
•
Balston housing/ filter
•
Liquid separator Armstrong drainer/trap
•
0-30 psi gauge
•
0-1600 cc/min adjustable metered bypass
2.16.2 Mounting Brackets
Two sample conditioning system mounting brackets are available: a single stream
bracket or a multiple stream bracket (
) for up to three modules.
See
Figure 2-25 Single and Multiple Stream Sample Conditioning Assemblies