Page 3–6
2101510 Rev. AG
6)
Insert gas probe into pipeline coupling (see
7)
Using the correct tool, tighten probe. Securely tighten so there is no gas leakage.
Do not over-tighten.
8)
Install shut-off valve on secondary side of sample probe, if desired.
Figure 3-3 Sample Probe
Figure 3-4 Sample Probe Insertion
3.4 Stand Alone Installation
If installing a NGC8201 using a free-standing pipe, use this procedure to install the
pipe. Before beginning, review the procedure and the materials required for
installation.
THREADED NECK 3/4" NPT.
PRESSURE RELIEF VALVE
PRESSURE GUAGE
1/4" OUTLET PORT
3/4" NPT
GAS FLOW
SAMPLE PROBE
CUSTOMER INSTALLED COUPLING
PRESSURE GAUGE
OUTLET PORT 1/4" NPT
RELIEF VALVE