May 2014 – TD 293 OPERATING MANUAL – SmartGuard
Page 21
3.2.5
Start recording using AADI Real-Time Collector
•
Return to Main menu
•
Open
Control Panel
and select the
Recorder Panel
tab
•
Set the recording intervals for each individual group if enabled in Deployment Settings, refer
chapter 4.1
•
Start individual groups or Start all Groups
3.2.6
Verify data and recording operation using AADI Real-Time Collector
•
Return to the Main menu
•
Click the spreadsheet image/text viewer located under the Data Visualization header to view
real-time data in a stylesheet format.
o
Use the tabs in the top row to select the recording group to display data from one group
or scan trough all active groups.
o
Tick Auto Refresh to receive updated data
•
Click the graph image under the Data Visualization header to use the chart viewer.
o
Select up to 4 parameter to display on the Y axis
3.3
Communication settings
Communication settings can then be changed in
Device Configuration,
User Maintenance, Real-Time Server
heading, to access the
User Maintenace
you need to enter the password, refer chapter 4.3. Most
communication settings should not be changed. Default values are settings for communication with AADI
Real-Time Collector.
Settings for serial connection:
To access this settings you need to add a Real-Time COM Port in the Communication menu, refer chapter
6.6.
Baud Rate:
The default baud rate is 11500
.
Select a baud rate in the range 2400 to 115200 (the baud rate
must equal the receiver baud rate e.g. the AADI Real- Time Collector).
Data Bits
: Set the number of Data Bits to 7 or 8. Set the value to
8
when the receiver is the AADI Real-Time
Collector.
Stop Bit
: Select between 1, 1.5 and 2 stop bits. Set the value to
1
when the receiver is the AADI Real-Time
Collector.
Parity
:
Select between None, Even and Odd parity. Set the value to
None
when the receiver is the AADI Real-
Time Collector.
Flow Control
:
Select between None and Xon/Xoff. Set the value to Xon/Xoff when the receiver is the AADI
Real-Time Collector.