6.3.2
Completion of Analog sensor configuration
•
Perform sensor
User maintenance
settings:
Refer chapter 4.3 for a short description. User Maintenance
holds properties necessary to operate the sensor in accordance with the way the user has chosen to apply
the actual sensor: open
Device Configuration
tab, check
Include User Maintenance
and press
Edit..
in the
User Maintenance heading. Refer Table 6-2 for a description of settings.
•
Perform sensor
System Configuration
if necessary to enable sensor output parameters:
Refer chapter
4.2.3 for a short description: open
Device Configuration
tab, and press
Edit..
in the System Configuration
heading.
–
Statistics can be calculated for each channel/parameter. Type
No Of Samples
for the statistics. Use
minimum 2 samples per recording interval, and maximum 5 samples per 10 seconds recording interval
(e.g max 30 samples per 1 minute recording interval). The sensor warm up time may reduce the
maximum samples per recording interval. If time series is enabled (configuration settings) maximum 63
first values will be available for readout.
–
Check desired output:
Average
,
MinMax
and
TimeSeries
. The latter will include statistics into the
recorded dataset.
•
Add the sensor to the correct recording group, refer chapter 4.1.1.
Table 6-2 User Maintenance settings for Analog sensor.
Property
Comment
Mandatory:
Node description
Type a description. Default is product name and serial number
Power Settings:
Continuous Power
Check if the sensor needs to be constantly powered
Warmup Time (ms)
Set value
Power Channel
Select power channel (if more than one channel)
Calculations:
Unit
Set the Unit for the scaled/linearized value
Range Min
Set the Range Min for the scaled/linearized value
Range Max
Set the Range Max for the scaled/linearized value
Coefficients Set1
Type
polynomial coefficients for Set 1. Refer explanation in the
beginning of the chapter
Set2 Enabled
Check if a second polynomial is to be used
Coefficients Set1
Type
polynomial coefficients for Set 2. Refer explanation in the
beginning of the chapter
Set2 Threshold
Type
the Set2 Threshold value for the point above which the second
polynomial shall be used