background image

NOVUS AUTOMATION 

1/11 

 

 

 

TagTemp-S 

TEMPERATURE DATA LOGGER - INSTRUCTION MANUAL V1.0x 

1

 

INTRODUCTION 

TagTemp-S is an electronic temperature recorder. The temperature 

measured by the sensor is recorded in equipment internal memory. 

This data is transferred through a smartphone equipped with an 

embedded NFC interface or by an NFC interface connected to a 

computer through an USB port (both not included). 
LogChart-NFC application for ANDROID™ and the Nx Software for 

Windows® are the tools used to configure the operating mode, 

transfer and display data. The acquisition start and end parameters, 

logging interval, setpoint alarms and other parameters can be easily 

defined using these applications. 
The operating manual and Nx Software can be downloaded from 

the manufacturer's website. The ANDROID™ LogChart-NFC app 

can be downloaded from Google Play Store

2

 

MECHANICAL INSTALLATION 

TagTemp-S can be installed using a double-sided adhesive tape on 

the back of the device, taking care not to stick it over the 

identification tag. Another way is to use the blister hole to attach the 

transmitter. 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

3

 

TECHNICAL FEATURES 

Operating 

temperature 

-30 °C to 60 °C 

Storage 

temperature 

From -30 °C to 65 °C. 

Measuring range 

Temperature: -30.0 °C to +60.0 °C. 

Accuracy of 

measurement 

±1 °C max. over the full measuring range. 
Note:  You can correct the measurement 

error in the OFFSET parameter using the 

configuration software. 

Measurement 

resolution 

Temperature: 0.1 °C / 0.1 °F 

Memory Capacity 

4020 registers. 

Interval between 

Measurements 

1 minute min. 
4 hours max. (240 minutes).  

Power Supply 

3.0 V lithium battery (CR2032), built-in, 

non-replaceable

Battery life 

(estimated) 

Above 2 years at a 5-minute acquisition 

interval. 

Dimensions 

37 x 57 x 8 mm 

RF Interface 

ISO 15693 (NFC-V) compatible. 

Data transfer time 

between register 

and smartphone 

Approx.  20 seconds (depends on the 

smartphone processing speed). 

LogChart-NFC 

Application 

ANDROID™ version: 4.0 or above. 
Languages: Portuguese or English. 

NFC interface for 

the computer 

USB  to NFC interface (optional), 

compatible with  Nx Software  for 

Windows®

Case 

Polyvinyl Chloride (PVC) 

Degree of protection  All electronics are protected by the 

housing. IP65 compatible. 

Certification 

CE 

Summary of Contents for 8813800000

Page 1: ...ick it over the identification tag Another way is to use the blister hole to attach the transmitter 3 TECHNICAL FEATURES Operating temperature 30 C to 60 C Storage temperature From 30 C to 65 C Measuring range Temperature 30 0 C to 60 0 C Accuracy of measurement 1 C max over the full measuring range Note You can correct the measurement error in the OFFSET parameter using the configuration software...

Page 2: ...on located on the front of the product can be used to start or stop data logging according to the settings To give the command to start or stop logs press this button until the LED turns on and then release it before it goes out If the LED goes off before releasing the button the command will not be validated and the button will be disabled for 30 seconds This is a protective measure to prevent ac...

Page 3: ... about TagTemp S Log interval Indicates the value defined of the interval between each acquisition Loggings Number of records made by TagTemp S Logging state Indicates whether the log is running stopped or waiting for loggings to start Start Delay Memory usage Indicates the memory usage mode full or circular Battery Indicates the status of the battery charge Good Perfect to use Medium Significant ...

Page 4: ...ption you can start logging data by pressing the logger button Stop loggings by button When you enable this option you can stop logging by pressing the logger button Enable circular memory When this option is enabled when the memory is complete the recording will continue overwriting the oldest records Use Fahrenheit degrees when this option is enabled the temperature unit of Celsius default is ch...

Page 5: ...below the specified lower limit for the range First occurrence The time of the first event the measured value was below the minimum limit Last occurrence The time from the last event the measured value was below the minimum limit Total logs within occurrences Number of acquisitions below the minimum range limit Average temperature within occurrences Average value of temperatures below the minimum ...

Page 6: ...the last 2000 To select a graph display range the Chart range option can be used 5 2 4 1 Data Chart Options Screen When you select the icon in the upper right of the screen the chart options will appear Fig 14 Data Chart Options screen 1 By pressing the icon you can share the obtained data using the applications already installed in the smartphone or through the option of publishing in the NOVUS C...

Page 7: ... is read from the device the subsequent changes will not be considered Export to Nx Software NXD Shares a file with the collected data in a format compatible with Nx Software nxd so that the data can be read and displayed by this application Initial screen selection Allows you to select the startup screen after reading the data from the device NOVUS Cloud Authentication Hire an account in the NOVU...

Page 8: ... The Configuration File option is used to read an existing configuration file The reading of the device is done by choosing the option Read the Device If the machine is logging data the following message will be displayed Fig 24 Reading Device Screen Stop logging This option stops data logging and is read allowing the device to be configured Show Only By choosing this option you can only view the ...

Page 9: ...est records will be overwritten with the new values Fig 28 Finalization Screen The Finalization screen is where the configuration settings are sent and allows you to perform some other useful functions Send Configuration Selecting this option will send the settings made for the connected device Save to File This option allows you to save the configuration settings made to a file so it can be reuse...

Page 10: ...the transfer The transfer time will be proportional to the number of acquisitions recorded Viewing Collected Data After the transfer of the records the data collected are presented graphically as shown in the images below 6 4 1 Sharing and processing data After completing the collection a screen will be displayed with the collected data and these are presented in two forms in graph form and table ...

Page 11: ...hart Properties 8 Reports and Data Export 9 Formulas 10 Sending data to NOVUS Cloud 11 Exit 6 5 Exporting and creating reports with the data collected Using the NX Software you can export the collected data to various file formats or create reports from the available templates These two features are accessed by clicking the Reports button in the function menu Export Selecting this option from the ...