Correcting the Resolution
1.
Press
SETUP
. The soft key menu for sweep condition settings appears.
2.
Press the
MORE 1/2
soft key.
3.
Press the
RESOLN CORRECT OFF ON
soft key. Each time you press the soft
key the setting toggles between ON and OFF.
Note
• Since the wavelength resolution function is set according to the monochromator slit width,
the setting resolution and actual resolution may not match. For this instrument, if the
resolution is set to 0.1 nm, the actual resolution will be 0.09 nm for a 1450 nm wavelength,
and 0.07 nm for a 1550 nm wavelength.
• If the resolution correction function is turned ON, measured data is processed by the
software so that it matches the set resolution.
In the following cases, the resolution correction function has no effect even when turned ON.
• When the set resolution is 0.02 nm
• When the measured wavelength is 1400 nm or less
• Turn OFF the resolution correction function when performing measurement for which the
level accuracy is guaranteed.
When the resolution correction function is turned ON, set the correction mode (in R02.01
or later versions).
1.
Press
SYSTEM
. The soft key menu regarding the system appears.
2.
Press the
MORE
soft key repeatedly until the MORE 4/4 menu is displayed.
3.
Press the
RES CORRECT MODE1/2
soft key. Each time you press the soft key
the setting toggles between MODE1 and MODE2.
5.7 Wavelength (Frequency) Resolution Settings