![YOKOGAWA AQ6370D Скачать руководство пользователя страница 159](http://html2.mh-extra.com/html/yokogawa/aq6370d/aq6370d_user-manual_3665131159.webp)
6-39
IM AQ6370D-01EN
W
aveform Display
1
2
3
4
5
6
7
8
9
10
11
App
Index
MARKER DISPLAY
This key is used to set whether to display the difference relative to the moving marker
(OFFSET) or the difference relative to the next marker (SPACING) in the marker display.
(Default: OFFSET)
If the active trace span is 0 nm, the wavelength difference relative to the moving marker
is 0.000 nm.
If a fixed marker is placed at the -210 dBm wavelength value, "????????" is displayed as
the level difference from that fixed marker.
When a fixed marker is positioned and the moving marker is set to a wavelength value of
-210 dBm, the level difference is set to 23.22 dB, regardless of the fixed marker’s level.
OFFSET
Sets the difference value display as the difference between each marker and the moving
marker (OFFSET).
Data area when MARKER DISPLAY is set to OFFSET
SPACING
Sets the difference value display as the difference between each marker and the next
marker.
Data area when MARKER DISPLAY is set to SPACING
6.9 Marker Display