Executing Go/No Go Judgment
1.
After creating or loading template data, press
ADVANCE
followed by the
TEMPLATE
soft key.
2.
Press the
TYPE
soft key. The judgment condition setting menu is displayed.
3.
Press the
UPPER
,
LOWER
, or
UPPER &LOWER
soft key once.
4.
Press the
RETURN
soft key. The screen returns to the previous stage.
5.
Press the
GO/NO GO
soft key to select ON. The judgment results are displayed
on screen as PASS or FAIL.
Loading Template Data
Load template data on the instrument.
1.
Press
FILE
, followed by the
ITEM SELECT
soft key. The soft key menu for
selecting the data type appears.
2.
Press the
TEMPLATE
soft key.
3.
Press the
READ
soft key.
4.
Press the
FILE ->@@@@
soft key. The load target line selection menu is
displayed (where @@@@ is the currently set contents).
5.
Press the
UPPER LINE
,
LOWER LINE
, or
TARGET LINE
soft key once. The
screen returns to the previous stage.
6.
Move the cursor to the template data file to load from the file list, then press the
EXECUTE
soft key.
Template Data Types
The following types of template data are available.
• CSV (comma delimited) files created on an external PC
• Waveform files of the instrument (.CSV or .BIN files)
Note
• After loading data, WL SHIFT and LVL SHIFT in the template are set to zero.
• For information about data formats when creating template data on an external PC, see the
Explanation subsection.
7.12 Go/No-Go Judgment (Template)