![YOKOGAWA AQ6370D Скачать руководство пользователя страница 213](http://html2.mh-extra.com/html/yokogawa/aq6370d/aq6370d_user-manual_3665131213.webp)
7-38
IM AQ6370D-01EN
Explanation
Grid Table
A grid table lists wavelengths (frequencies) that are referenced by part of the analysis
function when it is executed. There is a standard Grid table and a custom Grid table.
Standard Grid Table
This Grid table is created with pre-defined wavelength (frequency) ranges.
It can be created in the following manner by setting the reference wavelength (frequency)
and frequency spacing.
200 GHz SPACING to 12.5 GHz SPACING
Creates a grid table with 200 GHz to 12.5 GHz grid spacing.
REFERENCE WAVELENGTH
Sets the grid table reference wavelength. This can be set in the range of 1000.0000 to
1700.0000 nm.
Custom Grid Table
Users can edit this Grid table freely.
It is created automatically by setting the start/stop wavelength (frequency), reference
wavelength (frequency), and frequency spacing.
Users can add or delete an arbitrary channel to/from the created Grid table or edit
wavelength (frequency) values for each channel there.
START WL
Sets the start wavelength.
STOP WL
Sets the stop wavelength.
SPACING
Sets the frequency spacing.
REFERENCE WAVELENGTH
Sets the grid table reference wavelength. This can be set in the range of 1000.0000 to
1700.0000 nm.
VALUE EDIT
You can edit channel points.
Note
For details on the Grid table, see appendix 1, "WDM Wavelength Grid Table."
7.10 Editing the Grid Table