5-20
IM AQ6370D-01EN
5.6 Sweep Width Settings
Procedure
The following are three ways of setting the sweep width.
• Press the
SPAN WL
or
SPAN FREQ
soft key.
• Press the
START WL/STOP WL
or
START FREQ/STOP FREQ
soft key.
• Use the one-action key (soft key).
The following explains these procedures.
Setting by Pressing the SPAN WL or SPAN FREQ Soft Key
1.
Press
SPAN
. The soft key menu for sweep width settings appears along with the
sweep width setting screen.
2.
Press
SPAN WL
for wavelength measurement or
SPAN FREQ
for frequency
measurement.
3.
Enter a sweep width using the rotary knob, arrow keys, or numeric key pad.
4.
Press
nm/ENTER
.
Note
• It is not necessary to press nm/ENTER when using the rotary knob or arrow keys.
• The set value is applied to the measurement conditions area.
• When a setting is changed,
appears in the measurement conditions area.
• If a value outside the setting range is entered, the nearest permitted value is set.
For instructions on switching the wavelength and frequency displays, see section 5.2.