![YOKOGAWA AQ6370D Скачать руководство пользователя страница 113](http://html2.mh-extra.com/html/yokogawa/aq6370d/aq6370d_user-manual_3665131113.webp)
5-44
IM AQ6370D-01EN
5.16 External Trigger Measurement
Procedure
Setting the Trigger Input Mode
1.
Press
SYSTEM
. The soft key menu regarding the system settings appears.
2.
Press the
MORE 1/4
soft key.
3.
Press the
TRIG INPUT MODE
soft key. The trigger input mode's setting menu is
displayed.
4.
Press the
SMPL TRIG MODE
soft key to select the sample trigger mode. Press
the
SWEEP TRIG MODE
soft key to select the sweep trigger mode. Press the
SMPL ENABLE MODE
soft key to select the sample enable mode.
If you select
SWEEP TRIG MODE
or
SMPL ENABLE MODE
, the setup procedure is
finished.
Note
• In sample trigger mode, set SWEEP SPEED to 1x and TLS SYNC to OFF.
• In sweep trigger mode, set SWEEP SPEED to 1x and TLS SYNC to OFF. Also, TRIGGER
SETTING and EXT TRIGGER MODE cannot be set.