5-23
IM AQ6370D-01EN
Measurement
1
2
3
4
5
6
7
8
9
10
11
App
Index
Explanation
Wavelength Sweep Width
The available setting range is 0, and 0.1 to 1100.0 nm. If you press the COARSE key you
can use the rotary knob or arrow keys to change the numerical value in 1-2-5 steps. The
value changes in 1 nm steps if you do not press the COARSE key.
Start Wavelength
The setting range is 50,000 to 1700.000 nm. The value changes in 1 nm steps if you
press the COARSE key. If you do not press it, the value changes in 0.1 nm steps.
Stop Wavelength
The setting range is 600.000 to 2250.000 nm. The value changes in 1 nm steps if you
press the COARSE key. If you do not press it, the value changes in 0.1 nm steps.
Note
• Setting the sweep width changes the start wavelength and stop wavelength. The center
wavelength/frequency does not change.
• Changing the center wavelength changes the start wavelength and stop wavelength. The
sweep width does not change. When the start or stop wavelength is set, one wavelength
becomes fixed, and this changes the value of the sweep width. In addition, the center
wavelength value is also changed at the same time.
One-Action Keys
This is the general name for a key that uses data from the active trace waveform (the
currently displayed waveform) to set measurement conditions.
Setting conditions requires that a waveform is displayed for the active trace.
Δλ→SPAN
Sets the sweep width as six times the spectrum width (threshold 20 dB ) of the active
trace measurement waveform with the RMS method.
MKR L1-L2 →SPAN
Sets spacing between line markers 1 and 2 for sweep width.
The setting range is 0.1 to 1100 nm (in 0.1 nm steps).
Frequency Sweep Width
The allowed settings are 0 and the range of 0.01 to 330.000 THz.
If you press the COARSE key you can use the rotary knob or arrow keys to change the
numerical value in 1-2-5 steps. The value changes in 0.1 THz steps if you do not press
the COARSE key.
Start Frequency
The setting range is 11.5000 to 500,000 THz.
The value changes in 0.1 THz steps if you press the COARSE key. If you do not press it,
the value changes in 0.01 THz steps.
Stop Frequency
The setting range is 176.5000 to 665.0000 THz.
The value changes in 0.1 THz steps if you press the COARSE key. If you do not press it,
the value changes in 0.01 THz steps.
5.6 Sweep Width Settings