5-2
IM AQ6370D-01EN
5.2 Horizontal/Vertical Axis Settings
Procedure
Setting Up the Horizontal Axis
The units of the horizontal axis can be set to wavelength or frequency.
1.
Press
SETUP
. The soft key menu for sweep condition settings appears.
2.
Press the
MORE 1/2
soft key.
3.
Press the
HORIZON SCAL nm/THz
soft key. The units for the horizontal axis
switch from THz to nm, or viceversa.
Note
Pressing the HORIZON SCALE nm/THz soft key repeatedly toggles between nm and THz.
Setting the Measured Wavelength to Air Wavelength or Vacuum Wavelength
1.
Press
SETUP
. The soft key menu for sweep condition settings appears.
2.
Press the
MEAS WL AIR VACUUM
soft key. The measured wavelength is
changed to either a vacuum wavelength or an air wavelength.
Note
• This function only applies to measurements taken after the setting is selected. Previously
measured waveforms are not affected.
• If vacuum wavelength is set,
is displayed at the very bottom of the screen in inverse
video.