![YOKOGAWA AQ6370D Скачать руководство пользователя страница 192](http://html2.mh-extra.com/html/yokogawa/aq6370d/aq6370d_user-manual_3665131192.webp)
7-17
IM AQ6370D-01EN
Analysis
1
2
3
4
5
6
7
8
9
10
11
App
Index
Parameter Settings for Analysis Results Display
These parameters are set with respect to the display format for displaying analysis
results on the screen.
DISPLAY TYPE
This parameter is used to select the analysis results display format.
DISPLAY TYPE
Description and Procedure
ABSOLUTE
(Absolute value display)
1. Set CH RELATION as OFFSET or SPACING.
OFFSET: Displays the relative value, channel to the reference channel.
SPACING: Displays the wavelength difference and level difference
compared to the following adjacent channel.
2. If OFFSET is selected, the reference channel is set through “REF CH”.
• If the highest channel is used as a reference
Set to HIGHEST.
• If any desired channel is used as a reference
Set the reference channel number in “***CH”.
RELATIVE
(Relative value display
relative to grid)
There are no setting fields for this display type.
(See appendix 1, “WDM Wavelength GRID Table” for information on
changing the grid table. )
DRIFT(MEAS)
(Drift display using past
measurement wavelength
as a reference)
The procedure for this display type varies depending on the reference.
• Press the MAX/MIN RESET key if you want to use the waveform
data of the
current active trace as a reference.
• To change the measurement conditions and set the initially
measured waveform as the reference
There are no parameter setting items since the data that was
measured first becomes the reference.
DRIFT(GRID)
(Drift display using grid
wavelength as a reference)
The procedure for this display type varies depending on the reference.
• Press the MAX/MIN RESET key if you want to use the waveform
data of the current active trace as a reference.
• To change the measurement conditions and set the initially measured
waveform as the reference.
There are no parameter setting items since the data that was
measured first becomes the reference.
(See appendix 1, “WDM Wavelength GRID Table” for information on
changing the grid wavelength.)
7.7 WDM Transmission Signal Analysis