No. Message
Cause of Warning
124 Illegal directory name
Directory cannot be created because the name is not valid.
125 Illegal file name
A file cannot be saved due to an incorrect file name.
126 Directory already exists
Directory cannot be created because a directory of the same name
already exists.
128 File is write protected
The file cannot be re-written or deleted because its property has
been set to READ ONLY.
129 USB Storage full
The file cannot be saved, because the hard disk or the USB storage
device is full.
130 Directory full
No files can be created because the directory area is full.
131 No data
There was no data to be saved, although an attempt was made to
save the file.
132 File is not a trace file
The file cannot be read because it is not a waveform file.
133 File is not a program file
The file cannot be read because it is not a program file.
134 File is not a data file
The file cannot be read because it is not a data file.
135 File is not a settings file
The file cannot be read because it is not a settings file.
137 File is not a template file
File cannot be loaded since it is not a template file or there is a
problem with the format of the template data.
138 Cannot copy
The file copy cannot be executed because the “copy from” file and
the “copy to” file have the same name.
140 No paste possible
The paste operation cannot be executed during the editing of
programs because the number of blank rows is not sufficient.
141 No merge possible
It is not possible to execute a merge during the editing of programs
because the merged result would exceed the maximum number of
rows.
142 WL calibration failed
Calibration cannot be executed because the level of a light source
is not sufficient during the wavelength calibration or because
wavelength deviations exceed the calibration range.
143 Optical Alignment failed
Alignment adjustment cannot be executed because the level of a
light source is not sufficient during the alignment adjustment.
144 Go/No go judgment stopped
Go/No go judgment stopped because template data reading or
AUTO sweep was carried out with Go/No go judgment function set
to ON.
145 No template data
Go/No go judgment or template data display was attempted to be
made with no template data provided.
147 ResBW calibration failed.(Power)
Rms resolution calibration cannot be executed because the light
source level is too low.
148 ResBW calibration failed.(Spectrum)
Rms resolution calibration cannot be executed because the light
source spectrum width is too wide.
149 ResBW calibration failed.(Stability)
Rms resolution calibration cannot be executed because the light
source level is unstable.
150 File is not a logging file
The file cannot be loaded because it is not a logging file.
151 Disk space is not enough for logging
Logging cannot be started because there is not enough space for
saving waveform files with the logging function.
152 Logging was skipped for Auto zeroing
Unable to log at the specified interval because auto offset was in
progress at the time logging would have occurred (displayed only
the first time after logging is started).
153 Sweep time exceeds the set interval
Unable to log at the specified interval because the sweep time is
longer than the interval (displayed only the first time after logging is
started).
154 This file is not compatible
The file could not be loaded because the model does not support it.
170 Illegal character
An illegal character was entered for a network name.
171 Illegal address
An illegal address was set to an IP address.
172 This version is not compatible
Attempted to load an incompatible update file.
173 Update file read error
An incorrect number of files or file corruption was found when
attempting to load the update file.
174 Invalid password
The password entered for the key lock is not correct.
10.10 Warning Display Function