Measurement Sensitivity and Vertical Axis Effective Range
When the measurement sensitivity is set to NORMAL HOLD, the internal amplifier has
a fixed gain. Five different gains are set automatically according to the reference (REF)
level setting. However, the effective range of measurement data is limited to the following
range, using the reference (REF) level (dBm) as a reference.
REF-20 dBm < (effective range) < REF+10 dBm
If the level scale is set to 10 dB/DIV, the display would exceed the effective range, so the
areas at 10 dB from the screen stop and 20 dB from the bottom are inaccurate.
When measurement sensitivity is set to NORMAL HOLD, we recommend setting the
level scale to 5 dB/DIV or less.
Under the measurement sensitivity settings NORMAL AUTO, MID, and HIGH 1 to 3,
an automatic gain is used, permitting measurements over a wide level range through a
single sweep. Select the appropriate sensitivity level based on the light reception level
required for the particular measurement application.
Stray Light from the Monochromator
The instrument has a newly designed, high-performance monochromator. Depending
on the measurement conditions, stray light at a level 30 to 50 dB below the original
spectrum, as well as other stray light specific to the monochromator may occur in
wavelength areas 100 to 200 nm from the peak wavelength. If this stray light is likely to
have a severe impact on measurements, the sensitivity can be set to HIGH 1 to 3 and
CHOP MODE can be set to CHOP or SWITCH in order to reduce the effects of the stray
light.
Waveform with CHOP MODE OFF
CHOP MODE set to SWITCH
CHOP MODE Function List
CHOP
MODE
Function
Merit
Defect
OFF
Measurement according to the
stray light.
Stray light suppression ratio:
40 dB or more
Measurement time is rapid
even for high sensitivity
measurements.
Since the stray light is
also measured, the low
level component is not
accurate if the light source
is of high power.
SWITCH
Two sweeps per measurement
are performed. The first sweep
measures the stray light
component, and high dynamic
measurement is possible
through subtraction.
Stray light suppression ratio:
60 dB or more
Short-duration
measurement can be taken
with stray light removed.
If the measurement
duration is long, it is
influenced by temporal
changes in the measured
light.
3.9 Important Points During Measurement