&DEOH5RXWLQJ5HTXLUHPHQWV
Wire the MedLAN network ONLY in a “daisy-chain” method.
NEVER use “stub” wiring.
The maximum total length of the MedLAN network cable is 3,000
feet. Up to 30 weld controls may be connected to a single
MedLAN network.
A maximum of one WSS (Welding Support System) and one
Network Power Pack (DEP-100S), OR two Network Power Packs
(DEP-100S), OR one computer running MedView or
MedCommander may be connected to a single MedLAN
network.
The MedLAN cable wire ends terminate into MTSB (Phoenix)
connectors. Terminate the cable with this procedure:
1.
Strip the cable jacket to 1 inch.
2.
Remove foil shield strands outside the foil.
3.
Cover the shield with heat shrink.
4.
Cover the cable insulation end with one inch of heat-shrink
tubing.
5.
Strip 1/4 inch of insulation off each wire.
6.
Install wires in each connector as shown in Figure 11 below.
7.
Torque the terminal screws to 5 in.-lb.
Figure 11:
DUb]Y^QdY^WdXU
=UT<1>3QR\U
You must provide strain relief at the MedLAN cable entry and
exit points at the weld control enclosure. If the enclosure requires
drilling entry/exit holes, exercise cautions to assure that no metal
shavings are left inside the control.
3
2
1
CLEAR
SHIELD
BLUE
Heat
shrink
(Step #4)
Содержание MedWeld 3005
Страница 2: ...YY UTGU T UbQd_bµc 7eYTU V_b D 1 bY ...
Страница 84: ...3_ e YSQdY_ c Q T 9 UTGU T UbQd_bµc 7eYTU D 1 bY ...
Страница 110: ...GU T CSXUTe Uc UTGU T UbQd_bµc 7eYTU D 1 bY ...
Страница 126: ...1TfQ SUT C_VdgQbU 6UQdebUc UTGU T UbQd_bµc 7eYTU D 1 bY ...
Страница 144: ...CUde QbQ UdUbc UTGU T UbQd_bµc 7eYTU D 1 bY ...