3-13
•
Signal Level Units. Select the units to be used for checking
signal power levels (dBmV, dBuV, and dBm).
•
Test Point Compensation. Insert values in this menu to have
the SAM 4040D automatically include compensation levels in
readings.
•
Frequency Tuning Step Size. Adjust the Tuning Step Size
using the up/down arrow keys or the numeric entry keys (0.01 to
100.00 MHz in 10 kHz steps).
•
Fundamental Hum Frequency. Select the fundamental HUM
frequency to be measured (60 Hz, 50 Hz, 1Hz, or, Auto). With
Auto selected, the unit will automatically switch to 50 Hz on PAL
type plans and 60 Hz for NTSC plans. 1Hz is for use in the case
of a 1Hz Power Supply.
•
Scan Rate. Two scan rates are available in the Scan mode,
Normal
and
Fast
. The
Fast
scan feature allows for rapid scan
displays, but at less accuracy.
Normal
scan rate is slower but
more accurate.
•
Scan Audio Carriers. Omitting the audio carriers during
Scan
produces a faster scan.
•
C/N Calibration. To increase the accuracy of the carrier-to-
noise measurements, the measurement algorithm incorporates a
Noise floor calibration. To perform the calibration, select C/N
calibration and press the
↵
ENTER key. To get a valid
measurement, the program prompts you to be sure that no cable
is connected to RF IN connector. Once the measurement is
verified, press the
OK
soft key to perform the calibration. The
screen will display the system noise level. When the calibration
is complete, the screen displays the noise floor, expressed in
dBuV.