Using Spectrum Analyzer
58
Multi-adjacent channel power measurement with spectrum analyzer
Spurious emissions
The Spurious Emissions measurement is to identify or determine the power level of in-band or out-of-band spurious emissions
within the necessary channel bandwidth and modulated signal. The CellAdvisor 5G indicates either Pass or Fail based on the
specified limit of the signal.
Setting measure setup
After configuring spectrum measurement in "
Configuring Spectrum Measurements
", you can set the measure setup to continue your
measurement. The frequency setting is not used in the Spurious Emissions mode. The measurement settings can be saved and
recalled as a file by selecting File Type as Setup and load the file onto the instrument using the
Save/Load
icons on the side bar.
1
Tap the
Setup
(
) icon on the side bar.
2
Tap to switch the
Measure Type
to
Full
or
Examine
.
NOTE:
The Examine mode displays only the selected range while the Full mode lets the instrument automatically changes
the selected range from one another.
3
Tap
Configuration
to set up the range table and parameters as below.
a
Select
Range
from 1 to 20 and switch each to
On
to display the selected range in the result table or
Off
to hide it from the
table.
b
Select
Start Frequency
to specify the start frequency for the selected range using the on-screen keyboard.
c
Select
Stop Frequency
to specify the stop frequency for the selected range using the on-screen keyboard.
d
Select
Start Limit
and
Stop Limit
to specify the upper limit for Pass/Fail indication using the on-screen keyboard.
You can set the threshold line by setting the start/stop limit.
e
Select
Attenuation
and specify a value in the multiple of five using the on-screen keyboard.
f
Select
RBW
to specify a RBW value using the on-screen keyboard.
g
Select
VBW
to specify a VBW value using the on-screen keyboard.
Содержание CellAdvisor 5G
Страница 1: ...1 CellAdvisorTM 5G User s Guide ...
Страница 2: ......
Страница 18: ......
Страница 22: ......
Страница 110: ......
Страница 146: ......
Страница 182: ...Using LTE LTE A TDD Analyzer 170 CCDF measurement with LTE LTE A TDD Analyzer ...
Страница 264: ......
Страница 312: ......
Страница 332: ......
Страница 338: ......
Страница 346: ...Using RFoCPRI Analyzer 334 Rx Settings with RFoCPRI Analyzer ...
Страница 352: ......
Страница 356: ...Appendix 344 ...
Страница 357: ...345 ...
Страница 358: ......