Using DSS Signal Analyzer
260
a
Select
Range
from 1 to 20 and switch each to
On
to display the selected range in the result table or
Off
to hide it from the
table.
b
Select
Start Frequency
to specify the start frequency for the selected range using the on-screen keyboard.
c
Select
Stop Frequency
to specify the stop frequency for the selected range using the on-screen keyboard.
d
Select
Start Limit
and
Stop Limit
to specify the upper limit for Pass/Fail indication using the on-screen keyboard. You
can set the threshold line by setting the start/stop limit.
e
Select
Attenuation
and specify a value in the multiple of five using the on-screen keyboard.
f
Select
RBW
to specify a RBW value using the on-screen keyboard.
g
Select
VBW
to specify a VBW value using the on-screen keyboard.
5
Tap
Menu
>
BW/AVG
>
Average
to set the number of measurements to be averaged using the on-screen keyboard.
The input value range is from 1 to 100.
Setting limit
1
Tap
Menu
>
Limit
.
2
Tap to switch the
Test Limits
to
On
or
Off
to enable or disable Pass/Fail indication.
3
Optional.
Tap the
Save
icon on the side bar and select the save option form the choices available for your measurement mode.
Spurious emissions measurement with DSS Analyzer
NOTE:
You can only set the frequency range and attenuation by tapping the Configuration icon. If you select the first icon
next to the Range table above, it only shows the selected range and if you select the second icon next to the Range
table, it keeps moving from the first selected range to the final selected range.
Power vs Time
Power vs Time (Frame)
The Power vs. Time (Frame) measures the modulation envelope in the time domain, showing the power of each time slot in a DSS
signal.
Содержание CellAdvisor 5G
Страница 1: ...1 CellAdvisorTM 5G User s Guide ...
Страница 2: ......
Страница 18: ......
Страница 22: ......
Страница 110: ......
Страница 146: ......
Страница 182: ...Using LTE LTE A TDD Analyzer 170 CCDF measurement with LTE LTE A TDD Analyzer ...
Страница 264: ......
Страница 312: ......
Страница 332: ......
Страница 338: ......
Страница 346: ...Using RFoCPRI Analyzer 334 Rx Settings with RFoCPRI Analyzer ...
Страница 352: ......
Страница 356: ...Appendix 344 ...
Страница 357: ...345 ...
Страница 358: ......