Using Spectrum Analyzer
50
Using marker table
▪
Select Marker
: Selects an active marker whose position can be changed by tapping and switching marker from M1 to M6
(each marker has a different color). The assigned number of the active marker is displayed in the Marker Table box and the
active marker number is also displayed right next to the active marker on the trace. When tapping the Active Marker Off
(
) icon, the selected marker disappears.
▪
Normal
: This Normal marker type provides the reading of a marker position on the trace along with the marker number
between one and six.
▪
Delta
: This Delta marker type is associated with a Normal marker. A Normal marker must be set before a Delta marker is
set. When the Delta marker is set, the position set by the Delta marker becomes the reference position of the Normal
marker and the marker’s X and Y values displays the difference compared with the Delta marker.
▪
Delta Pair
: This Delta Pair marker type is associated with a Normal marker. A Normal marker must be set before a Delta
Pair marker is set. When the Delta Pair marker is set, the position set by the Delta Pair marker becomes the reference
position of the Normal marker and the marker’s X and Y values displays the difference compared with the Delta Pair
marker. The reference position will be varied in accordance with trace change.
▪
Marker All Off (
)
: Turns all the markers the screen off. When the Marker View is selected for those markers, the
instrument displays those markers back at the previous position. If a measurement mode is changed, current settings are
not restored.
▪
Start
: Sets the frequency of the active marker to the start frequency of spectrum analyzer.
▪
Center
: Sets the frequency of the active marker to the center frequency of spectrum analyzer.
▪
Stop
: Sets the frequency of the active marker to the stop frequency of spectrum analyzer.
▪
Peak Search
: Moves the active marker to the highest peak of the trace. You can also tap the hot key icon (
)
to use
this function.
▪
Min Search
: Moves the active marker to the lowest peak of the trace.
▪
Next Peak
: Moves the active marker to the second highest peak of the trace.
▪
Left Peak
: Moves the active marker to the highest peak to the left of its current position.
▪
Right Peak
: Moves the active marker to the highest peak to the right of its current position.
▪
Always Peak
: When the Always Peak is set to On, the instrument moves the active marker automatically to the highest
peak of the trace every time the trace is refreshed.
Moving a marker to a selected trace
If you have selected marker using the marker table, you can move the marker position to the other trace by following the procedure.
1
Tap
Menu
>
Trace
.
2
Tap
Select
Trace
and select the trace from
Trace 1
to
Trace 6.
3
Tap
Trace View
to On.
The selected marker moves to the selected trace.
NOTE:
If the marker is beyond the frequency range, the purple horizontal move bar appears on the right and left edge of
the chart screen. You can move the marker by moving the bar right and left.
Содержание CellAdvisor 5G
Страница 1: ...1 CellAdvisorTM 5G User s Guide ...
Страница 2: ......
Страница 18: ......
Страница 22: ......
Страница 110: ......
Страница 146: ......
Страница 182: ...Using LTE LTE A TDD Analyzer 170 CCDF measurement with LTE LTE A TDD Analyzer ...
Страница 264: ......
Страница 312: ......
Страница 332: ......
Страница 338: ......
Страница 346: ...Using RFoCPRI Analyzer 334 Rx Settings with RFoCPRI Analyzer ...
Страница 352: ......
Страница 356: ...Appendix 344 ...
Страница 357: ...345 ...
Страница 358: ......